HP 8517B S-Parameter Test Set Operating and Service Manual Printed in USA 12 August 1998 HP part number 08517-90054
Notice The information contained in this document is subject to change without notice. Hewlett-Packard makes no warranty of any kind with regard to this material, including, but not limited to, the implied warranties of merchantability and fitness for a particular purpose. Hewlett-Packard shall not be liable for errors contained herein or for incidental or consequential damages in connection with the furnishing, performance, or use of this material.
What You’ll Find in This Manual… HP 8517B S-Parameter Test Set Chapters 1 and 2 • Test set description and installation information Chapters 3 and 4 • Principles of operation and specifications Chapters 5, 6, and 7 • Troubleshooting, replacement procedures, and replaceable parts information HP 8517B S-Parameter Test Set Manual iii
Warranty Certification Hewlett-Packard Company certifies that this product met its published specifications at the time of shipment from the factory. Hewlett-Packard further certifies that its calibration measurements are traceable to the United States National Institute of Standards and Technology (NIST, formerly NBS), to the extent allowed by the Institute’s calibration facility, and to the calibration facilities of other International Standards Organization members.
CONSEQUENTIAL DAMAGES, WHETHER BASED ON CONTRACT, TORT, OR ANY OTHER LEGAL THEORY. Assistance Product maintenance agreements and other customer assistance agreements are available for Hewlett-Packard products. For assistance, call your local Hewlett-Packard Sales and Service Office (refer to “Service and Support”).
Service and Support Any adjustment, maintenance, or repair of this product must be performed by qualified personnel. Contact your customer engineer through your local HP Service Center. You can find a list of HP Service Centers on the web at http://www.hp.com/go/tmdir.
Safety and Regulatory Information Review this product and related documentation to familiarize yourself with safety markings and instructions before you operate the instrument. This product has been designed and tested in accordance with international standards. WARNING The WARNING notice denotes a hazard. It calls attention to a procedure, practice, or the like, that, if not correctly performed or adhered to, could result in personal injury.
Safety Earth Ground This is a Safety Class I product (provided with a protective earthing terminal). An uninterruptible safety earth ground must be provided from the main power source to the product input wiring terminals, power cord, or supplied power cord set. Whenever it is likely that the protection has been impaired, the product must be made inoperative and secured against any unintended operation.
Typeface Conventions Not all of the following conventions may appear within this manual, however, refer to this listing whenever you encounter on of the special font characters. • Used to emphasize important information: Use this software only with the HP 8517B Test Set. • Used for the title of a publication: Refer to the HP 8517B S-Parameter Test Set Manual • Used to indicate a variable: Type LOAD BIN filename.
x HP 8517B S-Parameter Test Set Manual
Contents Notice . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ii Restricted Rights Legend . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ii What You’ll Find in This Manual… . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . iii HP 8517B S-Parameter Test Set . . . . . . . . . . . . . . . . . . . . . . . . . . iii Warranty . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
2. Installation Initial Inspection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Operating Environment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Storing the Test Set . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Accessories Supplied . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Figure 2-1. Accessories Supplied with the HP 8517B Test Set . . . . Table 2-1. Accessories in Figure 2-1 . .
Making Operational Checks . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-13 Performance Verification . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-14 Using Anti-Rotation Clamps . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-15 Attach the first clamp . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-15 Figure 3-5. Using a Torque Wrench Correctly . . . . . . . . . . . . . . . . 3-16 Positioning the thumb-screw . . . . .
Unratioed Power Test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Figure 5-9. Diagram of IF Signal Path . . . . . . . . . . . . . . . . . . . . . Isolating Signal Path Problems . . . . . . . . . . . . . . . . . . . . . . . . . . . . Testing Signal Paths 1-4 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Testing Paths 5 and 6 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Figure 5-10. RF Path 1 and Path 2, HP 8517B Standard Test Set . Figure 5-11.
Bias Tees Replacement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-11 A7 and A9 Bias Tees . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-11 Test Port Coupler Replacement . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-11 A6 and A8 Test Port Couplers . . . . . . . . . . . . . . . . . . . . . . . . . . 6-11 Test Port Attenuator Replacement . . . . . . . . . . . . . . . . . . . . . . . . . . 6-11 A16 or A17 Port Attenuators . . . . . . . . . . . . . . .
Contents-6 HP 8517B S-Parameter Test Set Manual
1 Getting Started This is the operating and service manual for the HP 8517B S-parameter test set. It is to be used in conjunction with the HP 8510C Network Analyzer Operating and Service Manual. Together, these manuals provide information needed to configure the system and make measurements. How to Use This Manual Step 1.
Getting Started Test Set Description Test Set Description The HP 8517B test set configured with an HP 8510B/C network analyzer and an HP 8360 series source, creates a system capable of making S-parameter measurements from 45 MHz to 50 GHz. The system is particularly suited for making two-port device measurements. You can measure all four S-parameters without physically reversing the DUT (device under test).
Getting Started Test Set Description Figure 1-2 Option 004, HP 8517B Test Set Block Diagram Figure 1-3 Option 007, HP 8517B Test Set Block Diagram HP 8517B S-Parameter Test Set Manual 1-3
Getting Started HP 8517B Test Set Options HP 8517B Test Set Options Options for the HP 8517B test set are described in Table 1-1, below. Table 1-1 HP 8517B Options and Descriptions Options HP 8517B Test Set Option Description Option 001 Adds IF switching capability for up to four test set connections to the HP 8510C Network Analyzer. The test set used is selected from the network analyzer. Refer to “Controlling Multiple Test Sets” in Chapter 3, “Operation” of this manual for more information.
Getting Started Verifying Test Set Operation Verifying Test Set Operation The test set is designed to operate with an HP 8510C network analyzer. You can verify its operation by: Table 1-2 Test Set Operation Verification Checking System Operation Perform a system calibration as described in the HP 8510C Operating and Programming Manual. A successful calibration indicates that the system, and therefore the test set, are operating properly.
Getting Started Maintaining Measurement Accuracy Maintaining Measurement Accuracy Precision measurements rely on a precision calibration of the network analyzer. As a general rule, the shorter the time lapse between a calibration and a device measurement, the more precise the measurement results will be (within the limitations of your system). Hewlett-Packard recommends calibrating your system every few hours, or at least re-verifying your system’s calibration this frequently.
Getting Started Instruments Supported by This Manual Instruments Supported by This Manual A serial number label is attached to the rear panel of the test set. There are two parts to the serial number: a prefix, and a suffix. See Figure 1-4, and read the following descriptions: • The prefix includes the first four digits of the serial number, plus the letter. The contents of this manual apply directly to test sets with the same serial number prefix as the one(s) on the title page.
Getting Started Instrument Firmware Compatibility Instrument Firmware Compatibility To use the HP 8517B test set at frequencies up to 50 GHz, use an HP 83650B or HP 83651B source. If your network analyzer or source do not meet the required frequency ranges for your measurement needs, you need to upgrade your system. Please contact an HP Sales and Service representative for information.
Getting Started Service and Support Options Available Service and Support Options Available A variety of service and support products are available for coverage of repair, calibration, and verification issues. Contact your HP sales and service engineer for details. Their office addresses and phone numbers are listed in the front of this document. The purchase of an HP 8517B includes a one year on-site service warranty. In the event of failure, Hewlett-Packard provides service for the system.
Getting Started Accessories Accessories The accessories supplied with the test set, including part numbers, are listed in “Accessories Supplied” in this section and in the “Replaceable Parts List” of this manual. Accessories Available Table 1-5 Table 1-5 describes accessories available for the HP 8517B test set. For additional HP 8510C system accessories information, refer to the HP 8510C manual set.
Getting Started Accessories Table 1-5 Accessory Type Adapters Test Fixture Kit HP 8517B Test Set Accessories Available (Continued) Accessory Model Number Contents or Description HP 85135E 7 mm Flexible Test Port Return Cable Used when measuring one end of a 7 mm device connected directly to an HP 85130E adapter at Port 1. The test port return cable is connected between the device and Port 2.
Getting Started Recommended Test Equipment Recommended Test Equipment Additional equipment and accessories required for use with the test set are listed in Table 1-6. The table lists which items are required to verify the performance of the test set and which are required to operate it.
Getting Started Operating and Safety Precautions Operating and Safety Precautions CAUTION ESD Sensitive Assemblies Test set assemblies are very sensitive to damage from electrostatic discharge. They may or may not continue to function if subjected to electrostatic discharge. Their reliability, however, will be impaired. Handle the instrument devices at static-safe work stations, only.
Getting Started Operating and Safety Precautions 1-14 HP 8517B S-Parameter Test Set Manual
2 Installation This chapter contains installation information. Topics include initial inspection, environmental considerations, test set location when using it with the HP 8510C network analyzer, and making connections to the test set. Refer to “Installation” in the HP 8510C On-Site Service Manual for complete system connection and turn-on instructions. Refer to “Packaging the Test Set” for information about shipping the instrument.
Installation Initial Inspection Initial Inspection Inspect the shipping container and cushioning materials for damage. If there is damage, keep the container until you have checked the contents for completeness. If shipping materials are damaged, complete the performance tests outlined in the HP 8510C On-Site Service Manual. If the test set fails the performance tests, or is damaged or defective, keep the shipping materials.
Installation Initial Inspection Figure 2-1 Table 2-1 Accessories Supplied with the HP 8517B Test Set Accessories in Figure 2-1 (1) Power Cord U.S.A. only (6) Female 2.4 mm to Female 2.
Installation Installing the Test Set Into a System Rack Installing the Test Set Into a System Rack The HP 85043C system rack is the recommended model to use. To rack-mount the test set into a system configured with the HP 8510C, refer to the “Installation” chapter of the HP 8510C On-Site Service Manual and to the HP 85043C System Rack Manual.
Installation Configuring the Test Set in a System Configuring the Test Set in a System Figure 2-3 illustrates the cable connections for installing an HP 8517B in a system. Refer to the information that follows to install the test set.
Installation Configuring the Test Set in a System Connecting the System Cables Table 2-2 Connect System Power and Control Cables Connection Instructions and Connector Descriptions Check the test-set line module (see item (1) in Chapter 3, Figure 3-2 of this manual). It must be set for the correct voltage in your region. Refer to Chapter 5, “Troubleshooting” of this manual for information about changing the voltage selection or replacing the line fuse.
Installation Packaging the Test Set Packaging the Test Set To repackage the test set, use its original factory packaging. Refer to the test set using its complete model and serial numbers in any correspondence to an HP Sales and Service Office. Containers and materials identical to those used in the original shipment by the factory are available through Hewlett-Packard Sales and Service Offices. However, commercially available, comparable packaging materials may be used, also. 1.
Installation Packaging the Test Set 2-8 HP 8517B S-Parameter Test Set Manual
3 Operating the HP 8517B Test Set This chapter contains illustrations and descriptions of the HP 8517B frontand rear-panel features. Information about operating multiple test sets (Option 001) is also provided. Front-Panel Features Figure 3-1 Table 3-1 Front-Panel Features of the HP 8517B Test Set Front-Panel Description (1) Line Switch This switch turns the test set on or off.
Table 3-1 Front-Panel Description (Continued) (6) a1 LED This LED indicates that the test set is internally switched to the S11 or S21 mode and source power is switched to Port 1. (7) a2 LED This LED indicates that the test set is internally switched to the S22 or S12 mode and source power is switched to Port 2. (8) Port 2 This test port transmits RF energy from the source to the DUT and receives reflected or transmitted RF energy from the DUT.
Table 3-2 Rear-Panel Descriptions (Continued) (6) J12 8510 System Bus Connector Used for HP-IB communications with the HP 85101 display/processor. (7) Port 2 Bias A female BNC connector used to supply bias through the center conductor of Port 2 to active devices under test. (8) Port 1 Bias A female BNC connector used to supply bias through the center conductor of Port 1 to active devices under test.
Figure 3-3 Table 3-3 RF and IF Switching with Two Test Sets RF and IF Switch Settings in Figure 3-3, Above New ADDRESS of Test Set 1, 2 Test Set Number Selected Coaxial Switch Port Selected 20 Number 1 Port 1 21 Number 2 Port 2 1. Not all system connections are illustrated. 2. In dual source configurations, the second source may be multiplexed in a similar manner. If only one dual source test set is used, the second source may be connected directly to the appropriate test set.
Multiple Test-Set Connections For dual test-set configurations, set each rear panel address switch on each test set to the address shown in Figure 3-3. Refer to Figure 3-4 on page 3-7 for the configuration of two or more test sets. • Use the supplied test set interconnect cable to attach test set number 1, J11 to the network analyzer. • Use the supplied test set interconnect cable to attach test set number 2, J11 to test set number 1, J10.
Table 3-4 Test Set IF Switching Selecting a Test Set When Multiple Units are Configured A test set is selected via the built-in capability of the analyzer to generate an addressed command to each test set.
Figure 3-4 Table 3-5 RF and IF Switching with Four Test Sets Coaxial Switch Settings for Figure 3-4, Above New ADDRESS of Test Set1, 2, 3 Test Set Selected1, 2, 3 Coaxial Port Selected with Switch #1 Coaxial Port Selected with Switch #2 20 1 Port 1 Port 1 21 2 Port 1 Port 2 22 3 Port 2 Port 1 23 4 Port 2 Port 2 1. Not all system connections are shown. 2. In dual source configurations, the second source can be multiplexed in a similar manner. 3.
Measurement Calibration After selecting the active test set, complete the system calibration procedure as usual. When you select a different test set, make sure that you recall the cal set that applies to that test set. NOTE Since the cal-set limited instrument state does not include the number of the active test set, a cal set which does not apply to the current test set can be turned on without displaying the HP 8510C caution messages.
Measuring High-Power Devices with Option 004 The HP 8517B, Option 004 test set enables devices with output power up to 30 dBm (1 watt) to be measured at Port 2, without adding external attenuators. Reverse dynamic range (S12) degrades by approximately 3 dB. When measuring devices having greater than +17 dBm output power, you must use the Port 2 step attenuator to decrease the power to ≤+17 dBm. The step attenuator has to be set during calibration for the value needed during device measurements.
Changing Signal Path States After System Calibration Changing an internal attenuator or any external equipment after completing the calibration means the measurement results cannot be specified. You must use your own estimation of the potential error contribution due to the change. For example, when the port attenuation is changed with correction ON, the message CAUTION: CORRECTION MAY BE INVALID displays.
If you view the response of a short circuit, notice that: • • Changing Port 1 attenuator has negligible effect on S11 marker reading Changing Port 2 attenuator has negligible effect on S22 marker reading This is due to the way the attenuators are paired. Both the reference and the test signal are changed by approximately equal amounts. When viewing S21 or S12, however, changing Port 1 or Port 2 attenuators offsets the marker reading by the difference value between Port 1 attenuator and Port 2 attenuator.
5. View the S21 measurement. If the thru is connected, the transmission coefficient should be 1 ∠0° . 6. Press PARAMETER [S12], then press the following DISPLAY: menu keys: {SELECT DEFAULTS} {DEFAULT to MEMORY: 2} 7. Press PARAMETER [S12] to view the S12 measurement. If the thru is connected, the transmission coefficient should be 1 ∠0° . Normalization is turned OFF for S11 and S22 measurements. Then the appropriate memory is selected and normalization is turned on to view S21 and S12.
Making Operational Checks To check the operation of multiple test-set configurations, do the following: 1. Connect a device with a known response to test set number 1. Press the following HP 8510C front-panel keys: INSTRUMENT STATE [LOCAL] AUXILIARY MENUS [SYSTEM] {HP-IB ADDRESSES} {TEST SET} 2. When the message TEST SET HP-IB ADDRESS appears, enter [20], the address of test set number 1, then press [x1]. The measurement trace from test-set number 1 should display. 3.
Performance Verification Standard system performance verification procedures verify the operation of an Option 001 test set used as test set number 1. To verify the performance of a different test set in the configuration, enter its HP-IB address (as explained in “Making Operational Checks” on page 3-13) to select it, then continue with this procedure. Refer to the HP 8510C On-Site Service Manual for its performance verification procedure.
Using Anti-Rotation Clamps During the performance verification procedure, use anti-rotation clamps to secure RF connections at the test ports of each test set. While installed, each clamp stabilizes the connector on RF cable to the large nut on the test set’s RF test port. If an adapter is used, the clamp stabilizes the adapter to the front panel RF port connector. Without these clamps, the test port connections can loosen when a device under test is moved.
CAUTION Important! The test set RF connector becomes loosened easily. Hold the RF cable securely throughout the remainder of this procedure. Do not allow the cable to rotate. Figure 3-5 Using a Torque Wrench Correctly Positioning the thumb-screw See Figure 3-6. Position the clamp so the thumb-screw is positioned at the top of the clamp. • Turn the clamp to visually align the clamp flats with the flats on the test port connector nut.
NOTE The flats may actually be in any orientation, with respect to the front panel. Figure 3-6 Visually Aligning Clamp and Nut Flats Positioning the connector See Figure 3-7. Maneuver the clamp over the RF connector and onto the test port connector. 1. Hold the test cable with one hand. Use the other hand to press the clamp gently and steadily, as you wiggle it into position straight over the RF connector and onto the test port connector nut.
Operating the HP 8517B Test Set Using Anti-Rotation Clamps Aligning the thumb-screw See Figure 3-8. Ensure that the thumb-screw is aligned with the counter-sink hole on the clamp’s body. 1. Push the clamp toward the test set front panel. 2. Finger-tighten the thumb-screw. The cable cannot be damaged if the thumb-screw is tightened too tightly.
HP 8517B S-Parameter Test Set Manual -1
Specifications 4 Specifications Specifications describe the warranted performance of the instrument. The electrical specifications of the test set when used with an HP 8510 network analyzer, are defined in the HP 8510C On-Site Service Manual. Mechanical Specifications Table 4-1 HP 8510/HP 8517B Mechanical Specifications Test Port Description Specification Connector Type: Precision, rugged male test-port connector 2.4 mm1 Center Pin Recession2 +0.00254 mm to +0.0178 mm +0.0001 inches to +0.
Specifications Table 4-2 HP 8510/HP 8517B Supplemental Characteristics (Continued) Rear Panel RF Connectors Connector type precision 2.4 mm female Recommended connection torque 90 N-cm (8 in-lb), maximum for precision-2.
Specifications HP 8517B S-Parameter Test Set Manual 4-4
5 Troubleshooting the Test Set The troubleshooting strategy for the HP 8517B S-Parameter test set is a systematic sequence of procedures. The information may be used after system-level troubleshooting has identified the test set as the problem instrument. Use the troubleshooting flowchart in Figure 5-2 to help isolate the faulty assembly. Notice procedure numbers within some of the flow diagram blocks. These are keyed to troubleshooting procedures.
Troubleshooting the Test Set Theory of Operation Theory of Operation RF paths for Port 1 and Port 2 are similar. This section discusses in detail the source power signal, as it is routed via the RF IN connector, through the test set, and to Port 1 for S11 and S21 measurements. Figure 5-1 Standard HP 8517B RF Block Diagram The RF Source Power • RF source power enters the test set through the rear panel. The RF power is applied to the switch/splitter (A18).
Troubleshooting the Test Set Theory of Operation The Reference Signal Path The Test Signal Path The S11 and S21 Measurements The Sampler Control Switching • The reference signal is routed from the switch/splitter through the 20-7 dB tapered pad (AT1) to attenuate any VTO feed-through signals. • The signal is then applied to A10, the a1 sampler assembly, where it is down-converted to a 20 MHz IF (intermediate frequency).
Troubleshooting the Test Set Theory of Operation • The LO Signal Control If the test set is equipped with Option 001 (multiple test set capability), its begin state is inactive. That is, rather than using the test set’s internal IF, a switch is set in the test set to accept an IF from another test set. The LO, generated by A14 VTO/driver, is phase locked to the source so the desired VTO harmonic is 20 MHz greater than the source frequency.
Troubleshooting the Test Set Troubleshooting Sequence Troubleshooting Sequence Use this troubleshooting flowchart, along with the procedures in the following section, to help isolate faulty assemblies.
Troubleshooting the Test Set Troubleshooting Sequence Equipment Needed But Not Supplied Table 5-1 Gather the equipment listed in Table 5-1 to use for troubleshooting the test set (these items are not supplied with the instrument at shipment).
Troubleshooting the Test Set Troubleshooting Procedures Troubleshooting Procedures The troubleshooting procedures in this chapter are listed below: • • • • • • • A15 Primary Regulator Board Assembly Table 5-2 “A15 Primary Regulator Board Assembly” on page 5-7 “A27 Post-Regulator Board Assembly” on page 5-7 “Self-Test Indicators” on page 5-9 “HP 85102 IF Amplifier Test” on page 5-12 “Unratioed Power Test” on page 5-14 “HP 8517B, Option 007 Unratioed Power Test” on page 5-20 “Finding the Faulty Assembly”
Troubleshooting the Test Set Troubleshooting Procedures Figure 5-3 Power Supply Fuses and Test Points HP-IB Address Switch The HP-IB address switch is located on the rear-panel of the instrument. It is easy to access, but unnecessary to change unless the error message SYSTEM BUS ADDRESS ERROR appears on the system display. To set the HP-IB address switch, position the switches as illustrated in Figure 5-4 (the LSB is on the left-hand side; the MSB is on the right-hand side).
Troubleshooting the Test Set Troubleshooting Procedures Figure 5-4 Instrument HP-IB Switch Setting Fuse Location The fuse is located inside the line module at the rear panel of the test set. The value rating and part number of the fuse is listed in Chapter 7, “Replaceable Parts” of this manual. Review the information in Figure 5-5 to replace the line fuse or to set the voltage selector cam.
Troubleshooting the Test Set Troubleshooting Procedures Figure 5-6 The ACTIVE LED Location If the Self-Test Fails to Run Properly If the portion of memory holding the self-test program is faulty, the self-test will not run properly. If any of the following conditions occur, it indicates that the A4 HP-IB board is likely to be faulty.
Troubleshooting the Test Set Troubleshooting Procedures Figure 5-7 HP 8518B Test Set, A4 Board Assembly LEDs Table 5-4 Instrument Self-Test Indicator State Definitions Self-Test Indication LSN TLK SRQ REM Time (Following Power-On) PWON ON ON ON ON 0 to 0.
Troubleshooting the Test Set Troubleshooting Procedures Table 5-4 Instrument Self-Test Indicator State Definitions Self-Test Indication LSN TLK SRQ REM Time (Following Power-On) Pass ROM Test OFF OFF ON ON 0.5 to 0.2 seconds Start RAM Test OFF OFF OFF ON N/A Pass RAM Test OFF OFF OFF OFF after 2 seconds HP 85102 IF Amplifier Test The service adapter can substitute as a test set by returning the 20 MHz IF signal from the HP 85102 to the amplifier on the HP 85102.
Troubleshooting the Test Set Troubleshooting Procedures Service Adapter Conclusions Table 5-5 Results Observations If all channels look good The HP 85102 is working properly. The problem is most likely in the source or the test set. Refer to Procedure 4, “Unratioed Power Tests” in this chapter. If one or more channels look bad The problem is most likely in the HP 85102. Refer to the HP 8510C On-Site Service Manual for information about troubleshooting the HP 85102.
Troubleshooting the Test Set Troubleshooting Procedures Unratioed Power Test We will examine six RF signal paths in the test set by observing their associated IF signals. Figure 5-10 through Figure 5-15 show the test set traces (and simplified block diagrams) that are typical for RF paths 1 through 6. Figure 5-9 Diagram of IF Signal Path Isolating Signal Path Problems Determining the RF component in the test set that is most likely the cause of a failure can be accomplished within four steps: 1.
Troubleshooting the Test Set Troubleshooting Procedures 3. To examine each sampler’s IF signal, redefine which port and sampler the network analyzer uses for the phase lock. Use the following steps: a. Redefining to Port 2, sampler a2, press: PARAMETER [MENU] {USER 3 a2} then {REDEFINE PARAMETER} {DRIVE} DRIVE: {PORT 2} {PHASE LOCK} {a2} {REDEFINE DONE} Observe the resulting a2 IF trace on the display. b.
Troubleshooting the Test Set Troubleshooting Procedures Testing Paths 5 and 6 1. Connect a cable between Port 1 and Port 2. 2. To observe the b1 power-level trace through path 5, press: PARAMETER [MENU] {USER 4 b1} then {REDEFINE PARAMETER} {DRIVE} DRIVE: {PORT 2} {PHASE LOCK} {a2} {REDEFINE DONE} The b1 path signal is illustrated in Figure 5-12. Typically, each trace should resemble the examples in Figure 5-12, (±5 dB). 3. Record your results in Figure 5-6. 4.
Troubleshooting the Test Set Troubleshooting Procedures Figure 5-10 RF Path 1 and Path 2, HP 8517B Standard Test Set HP 8517B S-Parameter Test Set Manual 5-17
Troubleshooting the Test Set Troubleshooting Procedures Figure 5-11 RF Path 3 and Path 4, HP 8517B Standard Test Set 5-18 HP 8517B S-Parameter Test Set Manual
Troubleshooting the Test Set Troubleshooting Procedures Figure 5-12 RF Path 5 and Path 6, HP 8517B Standard Test Set HP 8517B S-Parameter Test Set Manual 5-19
Troubleshooting the Test Set Troubleshooting Procedures HP 8517B, Option 007 Unratioed Power Test Figure 5-13 through Figure 5-15 illustrate the signal path and trace display for the unratioed power tests on HP 8517B, Option 007 model test sets.
Troubleshooting the Test Set Troubleshooting Procedures Figure 5-14 RF Signal for Path 1 through Path 4, HP 8517B Option 007 HP 8517B S-Parameter Test Set Manual 5-21
Troubleshooting the Test Set Troubleshooting Procedures Figure 5-15 RF Path 5 and Path 6, HP 8517B Option 007 5-22 HP 8517B S-Parameter Test Set Manual
Troubleshooting the Test Set Troubleshooting Procedures Finding the Faulty Assembly Compare the results recorded in Table 5-6 with the matrix in Table 5-7. There are 32 possible combinations, twelve of which point to probable failures. These are shown in Table 5-7. Table 5-6 a1 Path1 Test Results for Path 1 through Path 6 a2 Path2 Sample Diagnostics b1 Path 3 b2 Path 4 b1 thru Path5 b2 thru Path6 Suppose you have determined that paths 3 and 5 are bad. Paths 1, 2, 4, and 6 are good.
Troubleshooting the Test Set Troubleshooting Procedures 5-24 HP 8517B S-Parameter Test Set Manual
6 HP 8517B Test Set Replacement Procedures The replacement procedures in this chapter are listed below: CAUTION 1 “A18 Switch/Splitter Assembly” on page 6-5 2 “A14, A10 to A13 Frequency Converter Assembly” on page 6-5 3 “A15 Regulator Board Assembly” on page 6-6 4 “C1 to C4 Filter Capacitors” on page 6-6 5 “2.
HP 8517B Test Set Replacement Procedures Figure 6-1 ESD-Safe Workstation Configuration Performance Tests Performance test information and procedures are located in Chapter 8 of the HP 8510C On-Site Service Manual. Note that the HP 8510C performance test software (shipped with the test set) is required to test the test set. The results generated during the HP 8517B performance-test are based on its operation as part of the HP 8510C system.
HP 8517B Test Set Replacement Procedures Equipment Needed But Not Supplied Table 6-1 lists the equipment needed to replace the assemblies documented in the following procedures. See Figure 6-2 to locate the test set assemblies. Table 6-1 Replacement-Procedure Tools Needed Tools Used for Procedure: HP Part Number 2 point Pozidriv screwdriver all components 8710-0900 1 point Pozidriv screwdriver all components 8710-0899 5/16 inch torque wrench, 90 N-cm (8 in-lb) all components and 3.
HP 8517B Test Set Replacement Procedures Figure 6-2 HP 8517B Major Assembly Location Diagram 6-4 HP 8517B S-Parameter Test Set Manual
HP 8517B Test Set Replacement Procedures Assembly Replacement Procedures Assembly Replacement Procedures Initial Precautions Before beginning any of the procedures in this chapter, do the following: 1. Turn off the HP 8517B test set. 2. Disconnect the test set’s power cord. 3. Remove the top- and side-covers, trim strips, and handles. 4. Do not exceed the maximum torque necessary to secure a connector. Refer to Table 6-1 for torque specifications. 5.
HP 8517B Test Set Replacement Procedures Assembly Replacement Procedures 6. Unplug the four harnessed (multi-colored) wire and socket assemblies. 7. Lift the frequency converter out of the test set. 8. Remove the frequency converter bracket by removing the two Pozidriv screws. Regulator Board Replacement A15 Regulator Board Assembly 1. See Figure 6-2 for component locations. 2. Unplug the transformer socket from the regulator board. 3.
HP 8517B Test Set Replacement Procedures Assembly Replacement Procedures Test Port Connector Replacement 2.4 mm Test Port Connectors 1. Incline the rear of the test set approximately four inches and support it with a stable, solid object (such as a thick book). 2. Using a 9/16-inch open-end wrench, unscrew the test port connector nut. 3. Pull out the center conductor assembly and replace it with a new assembly. Do not touch either end of the new assembly.
HP 8517B Test Set Replacement Procedures Assembly Replacement Procedures Fan Assembly Replacement B1 Fan Assembly WARNING Electrocution can result if ac power remains connected to the test set during this procedure. Disconnect the ac power cord before starting. Turn off the test set and disconnect the power cord from the ac-power mains. To replace the B1 fan assembly, remove T1 transformer to expose the ends of the fan wires which must be de-soldered from the line module, FL1.
HP 8517B Test Set Replacement Procedures Assembly Replacement Procedures WARNING Electrocution can result if ac power remains connected to the test set during this procedure. Disconnect the ac power cord before starting. 1. See Figure 6-2 for component locations. 2. Position the test set onto its right-hand side. 3. Using a 5/16-inch open-end wrench, remove the hard-line cables connected between the coupler A7 and the bulkhead connector J2. 4.
HP 8517B Test Set Replacement Procedures Assembly Replacement Procedures O-Ring Replacement Anti-Rotation Clamp O-Ring Replacement 1. Pry the O-ring out of the clamp with fine tweezers or a similar tool. 2. To insert the new O-ring (HP part number 0900-0007), engage one side of the ring in the slot of the phenolic clamp donut (shown in Figure 6-5). 3. Use your fingers to seat the O-ring into the groove within the clamp.
HP 8517B Test Set Replacement Procedures Assembly Replacement Procedures Bias Tees Replacement A7 and A9 Bias Tees 1. See Figure 6-2 for component locations. 2. Remove the two semi-rigid cables attached to the bias tee to be replaced. 3. De-solder the wires attached to the bias tee. 4. Remove the two screws that secure the bias tee to the bias-tee bracket. 5. Remove the bias tee. NOTE This procedure does not apply to Option 002 or Option 002 “plus” 007.
HP 8517B Test Set Replacement Procedures Assembly Replacement Procedures Input Amplifier Replacement A22 Input Amplifier Assembly NOTE This procedure is for Option 007, Option 002 “plus” 007, and Option 004 “plus” 007, only. 1. See the figure in Chapter 7, Table 7-17 for the location of the A22 amplifier. 2. Without disconnecting AT1 and AT2 fixed attenuators, remove the other five semi-rigid cables attached to the A18 switch/splitter. 3.
HP 8517B Test Set Replacement Procedures Assembly Replacement Procedures Post Regulator Board Replacement A27 Post Regulator Board Assembly NOTE This procedure is for Option 007, Option 002 “plus” 007, and Option 004 “plus” 007, only. 1. See the figure in Chapter 7, Table 7-17 for the location of the A22 amplifier bracket. The A27 board assembly is located beneath the A22 amplifier bracket. 2.
HP 8517B Test Set Replacement Procedures Assembly Replacement Procedures 6-14 HP 8517B S-Parameter Test Set Manual
7 HP 8517B Replaceable Parts Introduction This section provides information about ordering replacement parts. Parts that can be replaced include major assemblies and chassis hardware, but not individual pieces of major assemblies. Table 7-1 lists the major assembly reference designations and abbreviations used in the parts lists. Rebuilt-Exchange Assemblies Rebuilt-exchange assemblies (R-E assemblies) cost less than new assemblies. They are available through the rebuilt-exchange program.
HP 8517B Replaceable Parts Ordering Information To order parts listed in the replaceable parts table, record the following information, then address the order to the nearest Hewlett-Packard Sales and Service Office: the Hewlett-Packard part number the quantity required To order parts not listed in the replaceable parts lists, record the following information: the instrument model number the complete instrument serial number the description and function of the part, and the number of parts required Address
HP 8517B Replaceable Parts Table 7-1 Reference Designation Abbreviations (Continued) DC Directional Coupler DL Delay Line DS E Table 7-2 MP Miscellaneous Mechanical Part VR Breakdown Diode (Zener), Voltage Regulator P Electrical Connector (Movable Portion), Plug W Cable, Wire, Jumper Annunciator, Lamp, Light Emitting Diode (LED), Signaling Device (visible) Q Silicon Controlled Rectifier (SCR), Transistor, Triode Thyristor X Socket Miscellaneous Electrical Part R Resistor Y Crystal
HP 8517B Replaceable Parts Table 7-3 Standard Abbreviations A A Across Flats, Acrylic, Air (Dry Method), Ampere ADJ Adjust, Adjustment ANSI American National Standards Institute (formerly USASI-ASA) ASSY Assembly AWG American Wire Gage B BCD Binary Coded Decimal BD Board, Bundle BE-CU Beryllium Copper BNC Type of Connector BRG Bearing, Boring BRS Brass BSC Basic BTN Button C C Capacitance, Capacitor, Cold, Center Tapped, Cermet, Compression CCP Carbon Composition Plastic CD
HP 8517B Replaceable Parts Table 7-3 Standard Abbreviations (Continued) DCDR Decoder DEG Degree D-HOLE D-Shaped hole DIA Diameter DIP Dual In-Line Package DIP-SLDR DIP Solder DO Package Type Designation DP Deep, Depth, Diametric Pitch, Dip D-MODE Depletion Mode DP3T Double Pole Triple Throw DPDT Double Pole Double Throw DWL Dowel E E-R E-Ring EXT Extended, Extension, External, Extinguish F F Fahrenheit, Farad, Female, Fixed, Flange, Frequency, Film (Resistor) FC Carbon Film
HP 8517B Replaceable Parts Table 7-3 Standard Abbreviations (Continued) HDW Hardware HEX Hexadecimal, Hexagon, Hexagonal HLCL Helical HP Hewlett-Packard Company, High Pass I IIC Collector Current, Integrated Circuit ID Identification, Inside Diameter IF Intermediate Frequency, Forward Current IN Inch INCL Including INT Integral, Intensity, Internal J J-FET Junction Field Effect Transistor JFET Junction Field Effect Transistor K K Kelvin, Key, Kilo, Potassium KNRLD Knurled KVDC
HP 8517B Replaceable Parts Table 7-3 Standard Abbreviations (Continued) MINTR Miniature MIT Miter MLD Mold, Molded MM Magnetized Material, Millimeter MOM Momentary MTG Mounting MTLC Metallic MW Milliwatt N N Nano, None N_CHAN N-Channel NH Nano Henry NM Nanometer, nonmetallic NO Normally Open, Number NOM Nominal NPN Negative Positive Negative (Transistor) NS Nanosecond, Non-Shorting, Nose NUM Numeric NYL Nylon (Polyamide) O OA Over-All OD Outside Diameter OP AMP Op
HP 8517B Replaceable Parts Table 7-3 Standard Abbreviations (Continued) PF Pico Farad, Power Factor PKG Package PLSTC Plastic PNL Panel PNP Positive Negative Positive (Transistor) POLYC Polycarbonate POLYE Polyester POT Potentiometer POZI Pozidriv Recess PREC Precision PRP Purple, Purpose PSTN Piston PT Part, Point, Pulse Time PW Pulse Width Q Q Figure of Merit R R Range, Red, Resistance, Resistor, Right, Ring REF Reference RES Resistance, Resistor RF Radio Frequency
HP 8517B Replaceable Parts Table 7-3 Standard Abbreviations (Continued) SMB Subminiature, B-Type (Slip-on Connector) SMC Subminiature, C-Type (Threaded Connector) SPCG Spacing SPDT Single Pole Double Throw SPST Single Pole SIngle Throw SQ Square SST Stainless Steel STL Steel SUBMIN Subminiature SZ Size T T Teeth, temperature, Thickness, Time, Timed, Tooth, Typical TA Ambient Temperature, Tantalum TC Temperature Coefficient THD Thread, Threaded THK Thick TO Package Type Desi
HP 8517B Replaceable Parts Table 7-3 VAR Variable VDC Volts—Direct Current Standard Abbreviations (Continued) W W Watt, Wattage, White, Wide, Width W/SW With, Switch WW Wire Wound X X By (as used with dimensions), Reactance Y YIG Yttrium-Iron-Garnet Z ZNR Table 7-4 Zener Diode Multipliers, Abbreviation and Description Abbreviation Prefix Multiple T tera 1012 G giga 109 M mega 106 k kilo 103 da deka 102 d deci 10-1 c centi 10-2 m milli 10-3 µ micro 10-6 n nan
HP 8517B Replaceable Parts Figure 7-1 Power Cable and Plug Part Numbers HP 8517B S-Parameter Test Set Manual 7-11
HP 8517B Replaceable Parts Table 7-5 Fuse Name Instrument Fuses Location Amps for 100 to 120 Vac (part number) Amps for 220 to 250 Vac (part number) ac Line F1 Rear Panel 2.0 2110-0002 1.0 2110-0001 +5 Vdc F1 Regulator Board 1.5 2110-0043 1.5 2110-0043 - 5 Vdc F2 Regulator Board 1.0 2110-0001 1.0 2110-0001 +15 Vdc F3 Regulator Board 1.5 2110-0043 1.5 2110-0043 - 15 Vdc F4 Regulator Board 1.0 2110-0001 1.0 2110-0001 Bias Bias Fuses Front Panel 0.5 2110-0012 0.
HP 8517B Replaceable Parts Use this fast, efficient economical method to keep your HewlettPackard instrument in service. *Hewlett-Packard pays postage on boxes mailed in the U.S.A.
HP 8517B Replaceable Parts Table 7-6 Major Assembly Reference Designator and Parts Information Ref Des HP Part Number QTY Description A1 08513-60005 1 Front Panel Interface Board Assembly A6, A8 5086-7518 5086-6518 2 50 GHz Coupler, (NEW) 50 GHz Coupler, (R-E) A7, A9 5086-7484 5086-6484 2 Bias Tee (NEW) Bias Tee (R-E) A16, A17 33325-60004 33325-69004 2 60 dB Step Attenuator (NEW) 60 dB Step Attenuator (R-E) A18 5086-7570 5086-6570 1 Switch/Splitter (NEW) Switch/Splitter (R-E) AT1,
HP 8517B Replaceable Parts Figure 7-3 Major Assembly Locations in the Standard HP 8517B Test Set HP 8517B S-Parameter Test Set Manual 7-15
HP 8517B Replaceable Parts Table 7-7 Semi-Rigid Cable Assemblies in the Standard HP 8517B Test Set *Components not visible Ref Des HP Part Number Quantity Description W10 08517-20039 1 Cable Assembly AT1 to A10J2 W31 08517-20042 1 Cable Assembly A8J3 to A11J2 W32 08517-20041 1 Cable Assembly A6J3 to A12J2 W33 08517-20038 1 Cable Assembly A8J2 to A9J2 W34 08517-20038 1 Cable Assembly A6J2 to A7J2 W35 08517-20028 1 Cable Assembly A17J2 to A18J7 W36 08517-20031 1 Cable Assembl
HP 8517B Replaceable Parts Table 7-8 Flexible Cable Assemblies in the Standard HP 8517B Test Set Ref Des HP Part Number Quantity W1 08513-60121 1 Cable Assembly A12J3 to A1J11 W2 08513-60122 1 Cable Assembly A13J3 to A4J11 W3 08513-60123 1 Cable Assembly A10J3 to A2J11 W4 08513-60124 1 Cable Assembly A11J3 to A3J11 W13 08513-60133 1 Cable Assembly A3J1 to A14J1 W14 08513-60134 1 Cable Assembly A3J2 to A7J11 W16 08513-60136 1 Cable Assembly A3J5 to A5J11 W18 08513-60138 1
HP 8517B Replaceable Parts Table 7-8 Flexible Cable Assemblies in the Standard HP 8517B Test Set (Continued) W20 08513-60140 1 Cable Assembly A3J4 to A14J2 W21 08513-60141 1 Cable Assembly A5J2 to A18J4 W22 08513-60142 1 Cable Assembly A5J1 to A18J3 W23 08513-60143 1 Cable Assembly A5J3 to A18J6 W28 08517-60014 1 Ribbon Cable Assembly A16 through A19 W29 85102-60226 1 Cable Assembly Line Switch W30 08513-60014 1 Cable Assembly J10 to J11 W42 08513-60036 1 Cable Assembly A4
HP 8517B Replaceable Parts Table 7-9 Miscellaneous Parts, Front Panel HP 8517B Test Set Ref Des HP Part Number Quantity DS1,DS2 1990-0858 2 LED Lamp 25 MA, Maximum E33,E34 2110-0797 2 Fuse Holder F1,F2 2110-0012 2 Fuse 0.5 A 250 V J13,J14 – 1 ea 43 5021-8747 1 Front Bezel 48 1450-0615 2 Retainer LED 65 08340-40002 2 Mount LED 75 08517-00002 1 Front Dress Panel 76 08517-00003 1 Front Sub-Panel 1 Description See Table 7-10 1. See Table 7-10 for parts detail.
HP 8517B Replaceable Parts Table 7-10 Test Port Connector Assembly Components Item HP Part Number Quantity Description 1 5062-7274 1 2 5001-3930 5001-3904 5001-3905 5001-3906 3 5062-7276 1 Test Port Center Conductor 4 5021-3428 2 Test Port Connector Nut 5 5021-3427 2 Test Port Washer Test Port Connector .0002 IN SHIM .0005 IN SHIM .001 IN SHIM .
HP 8517B Replaceable Parts Table 7-11 Miscellaneous Parts, Rear Panel HP 8517B Test Set Ref Des HP Part Number Quantity E10 5062-7243 1 2.4 mm RF Connector Assembly FL1 9135-0217 1 Line Module-Filtered J10 1251-2197 1 Connector R&P 24F (Part of W30) J11 1251-2204 1 Connector R&P 24M (Part of W30) 1 0380-0643 2 Standoff-Hex 0.255-IN-LG 6-32 THD 21 0400-0010 4 Grommet-Round 0.188-IN-ID .312-IN-GRV-OD 31 0380-0926 4 Threaded Insert-STDF 6-32 0.
HP 8517B Replaceable Parts Table 7-11 Miscellaneous Parts, Rear Panel HP 8517B Test Set 49 5021-0906 14 Teflon™ Cable Sleeve 50 5001-3907 2 Line Module Retainer Clips 51 2190-0016 2 Washer-LK INTL T 3/8 IN 0.377-IN-ID 52 2190-0104 1 Washer-LK INTL T 7/16 IN 0.439-IN-ID 593 2950-0001 2 Nut-HEX-DBL-CHAM 3/8-32-THD 0.094-IN-THK 60 2950-0132 1 Nut-HEX-DBL-CHAM 7/16-28-THD 0.
HP 8517B Replaceable Parts Figure 7-4 Detailed Views of Hardware Stack-Ups HP 8517B S-Parameter Test Set Manual 7-23
HP 8517B Replaceable Parts Table 7-12 Miscellaneous Parts, Top Internal View Item HP Part Number Quantity Description 1 0360-0031 1 Terminal Crimp 2 2360-0115 5 Machine Screw 3 08512-20005 1 RFI Gasket 4 08513-00001 1 Top Deck 5 08513-00002 1 Transformer Bracket 6 08513-00015 1 Capacitor Plate 7 2200-0109 2 Machine Screw 8 08514-00007 1 Mounting Plate 9 See Table 7-13 for all part numbers and locations – C1 through C4 and R1 through R4 10 0513-00037 1 Enclosure
HP 8517B Replaceable Parts Table 7-12 11 Miscellaneous Parts, Top Internal View 2200-0107 Table 7-13 3 Machine Screws Miscellaneous Parts, Motherboard bottom.pcx Ref Des C1 - C4 HP Part Number Quantity 0180-2671 4 Description Capacitor-FXD .
HP 8517B Replaceable Parts Table 7-13 C5 - C6 0160-4834 J1 1251-574 Ref Des Miscellaneous Parts, Motherboard (Continued) Capacitor-FXD .047 µF ±10% 100VDC CER 2 Connector 20-PIN M Post Type (A19J1) DOES NOT INCLUDE A19MP1 and A19MP2 HP Part Number Quantity J2 1251-6868 4 Connector 5-PIN M Post Type J3 1251-7939 1 Connector - Post Type 0.100-PIN-SPCG 14-CONT (A19J3 DOES NOT INCLUDE A19MP3) J4, J5, and J6 1251-6868 J7 1251-3825 1 J8 and J9 1200-0508 1 ea.
HP 8517B Replaceable Parts Table 7-14 Parts Unique to HP 8517B Option 001 Test Set Ref Des HP Part Number Quantity A2 08513-60004 1 08513-69004 Description Board Assembly, IF Multiplexer (NEW) Board Assembly, IF Multiplexer (R-E) Cable Assemblies Unique to Option 001 W5 08513-60125 1 Cable Assembly, A2J2 to J11A1 W6 08513-60126 1 Cable Assembly, A2J8 to J11A4 W7 08513-60127 1 Cable Assembly, A2J5 to J11A2 W8 08513-60128 1 Cable Assembly, A2J11 to J11A3 W9 08513-60129 1 Cable As
HP 8517B Replaceable Parts Table 7-14 Parts Unique to HP 8517B Option 001 Test Set W10 08513-60130 1 Cable Assembly, A2J9 to J10A4 Ref Des HP Part Number Quantity W11 08513-60131 1 Cable Assembly, A2J6 to J10A2 W12 08513-60132 1 Cable Assembly, A2J12 to J10A3 W15 08513-60135 1 Cable Assembly, A3J3 to J10A7 W17 08513-60137 W19 08513-60139 1 Cable Assembly, A3J8 to J10A6 W24 08513-60144 1 Cable Assembly, A12J3 to A2J1 W25 08513-60145 1 Cable Assembly, A13J3 to A2J7 W26 085
HP 8517B Replaceable Parts Table 7-15 Parts Unique to HP 8417B Option 002 Test Set Ref Des HP Part Number Quantity Description W59 08517-20743 1 Cable Assembly, A12J3 to Port 1 J13 W60 08517-20737 1 Cable Assembly, A13J3 to Port 2J14 HP 8517B S-Parameter Test Set Manual 7-29
HP 8517B Replaceable Parts Table 7-16 Parts Unique to HP 8517B Option 004 Test Set Ref Des HP Part Number Quantity Description W55 08517-20051 1 Cable Assembly, AT7 to A9J1 W57 08517-20052 1 Cable Assembly, A8J3 to A17J2 W58 08517-20050 1 Cable Assembly, A17J1 to A11J2 AT7 08517-60020 1 3 dB Attenuator 7-30 HP 8517B S-Parameter Test Set Manual
HP 8517B Replaceable Parts Table 7-17 Parts Unique to HP 8517B Option 007 Test Set Ref Des HP Part Number Quantity Description AT8 and AT9 08516-60021 2 3 dB Attenuators A22 5086-7650 1 Input Amplifier A23, A24, A25, and A26 5086-7649 4 Buffer Amplifiers A27 08517-60045 1 Post Regulator Board Assembly (Beneath A22 Input Amplifier) HP 8517B S-Parameter Test Set Manual 7-31
HP 8517B Replaceable Parts Table 7-17 Ref Des HP Part Number Parts Unique to HP 8517B Option 007 Test Set (Continued) Quantity Description Cable Assemblies Unique to Option 007 W45 08517-20023 1 Front Panel J7 to A22 Input Amplifier W46 08517-20024 1 A22 Input Amplifier to A18J5 W47 08517-2003 1 A6 Coupler to A23 W48 08517-20061 1 AT1 to A24 W49 08517-20036 1 A8 Coupler to A25 W50 08517-20062 1 AT2 to A26 W51 08517-20034 1 A23 to AT3 W52 08517-20029 1 A24 to AT4 W53 085
HP 8517B Replaceable Parts Table 7-18 Parts Unique to Option 002 “plus” Option 007 There are no unique parts to Option 002 “plus” Option 007. All other parts are included in Option 002 or Option 007.
HP 8517B Replaceable Parts Table 7-19 Parts Unique to Option 004 “plus” Option 007 Ref Des HP Part Number Quantity W56 08517-20049 1 Description A17J1 to A25 All other parts are included in Option 004 or Option 007.
HP 8517B Replaceable Parts Table 7-20 Instrument Chassis Replaceable Parts Item Number HP Part Number Quantity Description 1 08513-00040 1 Top Cover 2 08513-00041 1 Side Cover, Perforated 3 5021-8403 1 Front Frame 4 5062-3747 1 Bottom Cover 5 5062-3799 1 Handle Assembly 6 5021-8496 1 Trim, Front Handle HP 8517B S-Parameter Test Set Manual 7-35
HP 8517B Replaceable Parts Table 7-20 Instrument Chassis Replaceable Parts (Continued) 7 5041-8821 4 Bottom Foot 8 5062-3757 1 Side Cover Item Number HP Part Number Quantity 9 5021-5837 1 18-inch Corner Strut 10 5021-5804 1 Rear Frame 11 5041-8802 1 Trim Strip 12 5041-8821 2 Back Panel Foot 13 5021-8538 1 Lock Foot 14 5021-8540 1 Lock Foot Description 7-36 HP 8517B S-Parameter Test Set Manual
Index A E abbreviations used in parts descriptions, 7-4 accessories descriptions, 2-3 active LED location, 3-1 adapter part numbers, 1-11 addressing a test set, 3-6 altering signal path conditions, 3-10 attaching anti-rotation clamps, 3-15 environmental requirements, 2-2 equipment requirements for replacing components, 6-3 ESD information, 6-1 ESD precaution, 1-13 B bias supply connector, 3-3 bias tees replacement, 6-11 bias tees usage, 1-2 block diagram, Option 004 HP 8517B, 1-3 block diagram, Option
N S network analyzer compatibility matrix, 1-8 normalizing parameters, viewing, 3-11 operation precautions, 1-13 operation verification, 1-5 option descriptions, 1-4 options for service and support, 1-9 ordering information, 7-2 O-ring replacement, 2-6, 6-10 safety precautions, 1-13 sales and service internet address, vi sales and service office locations, vi selecting a test set, 3-5 serial prefix location and description, 1-7 service and support options, 1-9 shipping the test set, 2-7 source compatibi