Operation Manual

• Please do not place pads near heart, genitals or dieseased skin.
• Users with implanted electronic equipment, such as pacemakers and
intracardiac de brillators who have not gotten doctor‘s permission must
not use the device. Pregnant women should not use the device during
the rst trimester, and should always consult a doctor, midwife or physi-
otherapist prior to use.
• Simultaneous connection of a PATIENT to a h.f. surgical EQUIPMENT may
result in burns at the site of the STIMULATOR electrodes and possible
damage to the STIMULATOR.
• Operation in close proximity (e.g. 1 m) to a shortwave or microwave
therapy EQUIPMENT may produce instability in the STIMULATOR output.
• Application of electrodes near the thorax may increase the risk of cardiac
brillation.
• Please do not drop, repair, or rebuild it privately.
• Please do not use electrode pads other than those supplied by the ma-
nufacturer, otherwise it may bring biocompatible hazard and may result
in errors.
• Please do not share the electrode pads with infective persons to avoid
cross-infection.
Setup and operating procedures
1. Battery loading
a Open battery cover at the back of the device.
b Load two „AAA“ size batteries. Please pay attention to polarity.
c Close the battery cover.
- Rechargeable batteries are not suitable for this device.
- Remove the batteries if the monitor will not be used for a month
or more to avoid relevant damage of battery leakage.
- Please do not mix use new and old batteries or different type of
batteries.
- Please do not throw battery into re.
Disposal of appliances and batteries
The device and the batteries, must be disposed of according to local regu-
lations at the end of their usage.