User Manual

MCP-CPF Process/ISMATEC SA/30.10.07/CB/GP
37
Analog interface
Pin 1, GND (ground)
Connected to the protective earth
conductor. Reference potential for
all other inputs.
Pin 2, remote
For changing between manual
control and analog interface. For
activating the analog interface,
pin 2 must be connected with pin
1 (GND).
Pin 3, start
In normal operation (pin 2 open)
pin 3 operates the foot-switch
(see basic settings page 17)
In remote operation (pin 2 to
GND) the pump starts when
connected to pin 1 (GND)
Pin 4, direction
In the open position the pump
turns clockwise; when connected
to pin 1 (GND) it turns counter-
clockwise.
Pin 5, speed IN
External speed control
(0 – 5 V, 0 – 10 V, 0 – 20 mA,
4 – 20 mA) Input impedance and
input range can be selected via a
DIP switch inside the pump (see
page 39)
Pin 6, input 1
Pin 8, input 2
Digital inputs (TTL-level). Active
when connected to pin 1. Are
only retrievable via RS232 inter-
face; e.g. with a software like the
ProgEdit.
Pin 7, +26V
DC
About +26 V
DC
are available (max.
current 1A).
Interface analogi�ue
Pin 1, GND (masse)
Connecté à la mise à terre. Poten-
tiel de référence pour toutes les
autres entrées.
Pin 2, remote
Pour commuter du service manuel
à l’interface analogique.
Pour activer l'interface analo-
gique, le pin 2 doit être connecté
au pin 1 (GND).
Pin 3, start
en exploitation normale (pin 2
ouvert), le pin 3 sert d’interrup-
teur au pied (réglages de base
p.17)
– en exploitation à distance (pin
2 sur GND), la pompe se met en
route dès qu’elle est connectée au
pin 1 (GND)
Pin 4, direction
Si ouvert, le sens de rotation de
la pompe est celui des aiguilles
d’une montre; si relié avec le pin
1 (GND), la pompe tourne dans le
sens anti-horaire.
Pin 5, speed IN
Réglage externe du nombre de
tours (0 – 5 V, 0 – 10 V, 020 mA,
4 – 20 mA ) Impédance d‘entrée
et réglage de zone au moyen de
l’interrupteur DIP à l’intérieur de
l’appareil (v.p. 39)
Pin 6, input 1
Pin 8, input 2
Entrées numériques (niveau TTL).
Actifs si reliés au pin 1. Ne peu-
vent être interrogés que par
l’interface RS232, p.e. avec le
logiciel ProgEdit.
Pin 7, +26V
DC
Environ +26 V
DC
sont à dispositi-
on (courant maximal 1A).
Digitale Eingänge (TTL-Pegel)
Digital inputs (TTL-level)
Entrées numéri�ues (niveau TTL)
Pin 2, remote
Pin 3, start
Pin 4, direction
Pin 6, input 1
Pin 8, input 2
Pin 13, speed intern
Analoge Eingänge
Analog inputs
Entrées analogi�ues
Pin 5, speed IN
0 – 5 V
DC
/ 0 – 10 V
DC
0 – 20 mA / 4 – 20 mA
Universal Ausgänge (PWM)
Universal outputs (PWM)
Sorties universelles (PWM)
Pin 14, output 1
Pin 15, output 2
Analog-Ausgang
Analog output
Sortie analogi�ue
Pin 9, speed OUT
0-10 V
DC
/ 0-7.2 kHz
Analogschnittstelle
Pin 1, GND (Masse)
Mit dem Schutzleiter verbundene
Masse. Bezugspotential für alle
anderen Eingänge.
Pin 2, remote
r Umschaltung zwischen ma-
nueller Bedienung und der Analog-
schnittstelle. Zur Aktivierung der
Analog-Schnittstelle muss Pin 2 mit
Pin 1 (GND) verbunden werden.
Pin 3, start
Im Normalbetrieb (Pin 2 offen)
dient Pin 3 als Fußschalter
(Grundeinstellungen Seite 17)
Im Remote-Betrieb (Pin 2 auf
GND) startet die Pumpe bei Ver-
bindung mit Pin 1 (GND)
Pin 4, direction
Wenn offen, dreht die Pumpe
im Uhrzeigersinn; wenn mit Pin
1 (GND) verbunden, dreht sie im
Gegenuhrzeigersinn
Pin 5, speed IN
Externe Drehzahlsteuerung
(0 – 5 V, 0 – 10 V, 0 – 20 mA,
4 – 20 mA) Eingangsimpedanz
und Wahlmöglichkeiten mittels
DIP switch im Geräteinnern (siehe
Seite 39)
Pin 6, input 1
Pin 8, input 2
Digitale Eingänge (TTL-Pegel).
Aktiv, wenn mit Pin 1 verbunden.
Können nur über RS232-Schnitt-
stelle bedient werden; z.B. mit
einer Software wie ProgEdit.
Pin 7, +26V
DC
Es stehen ca. +26 V
DC
zur Verfü-
gung (max. Strom 1A).
8 7 6 5 4 3 2 1
15 14 13 12 11 10 9
speed IN direction
input 1 start
+26VDC remote
input 2 GND
output 2 speed OUT
output 1 +5VDC
speed intern