Instruction Manual

Flowmaster / ISMATEC SA / 11.07.07/CB/GP
22
Analog interface
Pin 1, +24 V
DC
, 0.5 A
Auxiliary voltage for peripherals
(max. current 0.5A)
Pin 2, status error
This pin is connected to 'status 0'
(pin 9) if the pump is switched off
or in the case of an error.
max.: AC 70 V / 1 A
DC 48 V / 1 A
Pin 3, output 0
Reference point for the 'busy'
output, potential-free.
Pin 4, busy
See page 23
Pin 5, direction
If this input is set to high (5, 12 or
24 V) relative to 'GND' (Pin 8) the
pump rotates in the counter-
clockwise direction.
Configuration see page 17
Pin 7, speed IN
For external speed control
0 – 5 V (1M
)
, 0 – 10*V (20k
)
0 20 mA (250), 4 20 mA (250)
Configuration see page 17
Pin 8, GND (ground)
Reference potential for all analog
and digital inputs and the analog
output. GND lies on earth
potential.
Pin 9, status 0
Reference potential for both
status outputs, potential-free.
Analogschnittstelle
Pin 1, +24 V
DC
, 0.5 A
Hilfsspannung für Peripheriegerä-
te (Belastbarkeit: 0.5 A).
Pin 2, status error
Bei abgeschalteter Pumpe oder
bei einem Fehler ist dieser Aus-
gang intern mit dem Anschluss
'status 0' (Pin 9) verbunden.
max.: AC 70 V / 1 A
DC 48 V / 1 A
Pin 3, output 0
Bezugspunkt für den 'busy'-Aus-
gang, potentialfrei.
Pin 4, busy
Siehe Seite 23
Pin 5, direction
Wird dieser Eingang gegenüber
'GND' (Pin 8) auf High (5, 12 oder
24 V) gelegt, dreht die Pumpe im
Gegen-Uhrzeigersinn. Konfigurati-
on siehe Seite 17
Pin 7, speed IN
Für externe Drehzahlsteuerung
0 – 5 V (1M
)
, 0 – 10*V (20k
)
0 20 mA (250), 4 20 mA (250)
Konfiguration siehe Seite 17
Pin 8, GND (Masse)
Bezugspunkt für alle analogen
und digitalen Eingänge, sowie
den analogen Ausgang. GND liegt
auf Erdpotential.
Pin 9, status 0
Bezugspunkt für die beiden Sta-
tus-Ausgänge, potentialfrei.
Digitale Eingänge (5,12 oder 24
*
V
DC
)
Digital inputs (5,12 or 24* V
DC
)
Entrées numériques (5,12 ou 24* V
CC
)
Pin 5, direction
Pin 13, start
Digitale Ausgänge (Relais)
Digital outputs (Relais)
Sorties numériques (Relais)
Pin 2 / 9, error
Pin 10 / 9, ok
Pin 4/11/12, busy
Analog-Eingang
Analog input
Entrée analogique
Pin 7, speed IN
0 – 5 V
DC
/ 0 – 10 V
DC
* / 0 – 20 mA / 4 – 20 mA
Analog-Ausgang
Analog output
Sortie analogique
Pin 17, speed OUT, 0 – 5 V
DC
/ 0 – 10 V
DC
Pin 1, +24 V
DC
0.5 A
* = Default Wert
* = default value
* = valeur par défaut
Interface analogique
Pin 1, +24 V
CC
, 0.5 A
Tension auxiliaire pour appareils
périphériques (courant maximal
0.5 A)
Pin 2, status error (erreur d'état)
Lorsque la pompe est éteinte ou
en cas d'erreur, cette sortie est
reliée au raccordement «status 0»
(pin 9) max. : CA 70 V / 1 A
CC 48 V / 1 A
Pin 3, output 0 (sortie 0)
Point de référence pour la sortie
'busy' (active), sans potentiel
Pin 4, busy
Voir page 23
Pin 5, direction (sens)
L'entrée du Pin 5 doit avoir une
différence de potentiel de 5, 12
ou 24 V par rapport à 'GND'
(PIN 8). La pompe tourne dans le
sens inverse des aiguillles d'une
montre.
Configuration voir page 17
Pin 7, speed IN (entrée vitesse)
Pour la commande externe de la
vitesse
0 – 5 V (1M
)
, 0 – 10*V (20k
)
0 20 mA (250), 4 20 mA (250)
Configuration voir page 17
Pin 8, GND (ground) (masse)
Point de référence pour toutes les
entrées analogiques et
numériques et pour la sortie ana-
logique. GND est sur le potentiel
de terre.
Pin 9, status 0 (état 0)
Point de référence pour les deux
sorties d'état, sans potentiel.