Manual

IKA
-WERKE C 2000 basic / control Ver. 04 08.07
3DJH
The message "Bomb securely closed?" will appear. Ensure that the decomposition
vessel is properly closed and confirm with OK.
The ignition wire in the decomposition vessel completes the electrical circuit. The
calorimeter is ready to start. The message %RPE changes to a display of the func-
tion key assignment 6WDUW If the 6WDUW function key does not appear, please check
the ignition wire of the decomposition vessel.
d
Push the 6WDUW button.
Each time 1000 ignitions have been performed using a given decomposition vessel,
the following message will appear:
1000 ignitions performed with Bomb x
Inspect decomposition vessel
or contact IKA service
This indicates that the decomposition vessel has reached a maintenance point and
that a safety check must be carried out. Confirm this message by pressing 7$%
and 2. in sequence.
This message does not release the user from the responsibility of also continuously
checking the decomposition vessel for wear and carrying out safety inspections as
required.
In ([SORVLYH mode, please observe in addition the following:
As many as 20 decomposition vessels per device can be used. As many measure-
ments can be prepared as there are decomposition vessels available.
1 cap screw
2 filling head
3 ignition contact
4 spring element
VXVSHQGLQJWKH
GHFRPSRVLWLRQ
YHVVHOLQWRWKH
ILOOHUKHDGRIWKH
PHDVXULQJFHOO
FRYHU