User’s Manual 270_manual_ENG_C_170704.indd 1 2017. 7. 4.
Please note the important information below before reading this manual. This product can also be used for medical purposes as the InBody270 measures body composition. DANGER Warning 참고 주의 참조 주의 Failure to comply with safety warnings and regulations can cause serious injury or death. DANGER Caution 참고 주의 참조 주의 Failure to comply with safety cautions and regulations can cause injury or property damage. InBody Co., Ltd. [HEAD OFFICE] InBody Bldg.
User’s Manual Contents I. InBody270 Installation A. Product Components 4 B. Operating Environment 5 C. Installation Instructions 5 D. Initial Setup 8 E. Connecting Printer, Thermal Printer, Stadiometer, Blood Pressure Monitor, Barcode Reader and SD400 12 F. Maintenance 15 II. InBody Test A. Precautionary Steps 16 B. Test Instructions 17 C. Test Posture 19 III. Transportation and Storage A. Cautions during Transportation 20 B. Repacking Instructions 20 C.
I. InBody270 Installation A. Product Components The InBody270 consists of the following components. Please make sure all of the following components are present. * Please inspect each component of the InBody270 for defects prior to installation. ❶ InBody270 ❷ Adapter (DC 12V, 3.4A) 1 EA ❸ Power cord 1 EA Input Buttons ❹ User’s Manual 1 EA Upper part Hand electrodes Foot electrodes Hinge Footplate ❶ ❶ ❸ ❹ 제품보증 서 제품보증 서 제품보증 서 제품보증 서 ❷ 4 270_manual_ENG_C_170704.indd 4 2017. 7. 4.
B. Operating Environment Please make sure that the environment is adequate for the InBody270 installation. This equipment is designed for indoor use. If installing outdoors, the following requirements must be fulfilled. Temperature range 10 ~ 40°C (50 ~ 104°F) Relative humidity 30 ~ 75% RH Atmospheric pressure 70 ~ 106kPa C. Installation Instructions 1. Open the packing box of the InBody270 and remove the packing pads. Then take the InBody270 out of the box.
3. Level the InBody270 by rotating the leveling screws under the footplate to the left and right so that the air bubble is centered. * Leveling the equipment is necessary for accurate measurement of weight. There are a total of 4 leveling screws. Level Indicator Leveled Not leveled Leveling screws under the footplate DANGER Caution 참고 주의 참조 주의 • Avoid injuring your hands when rotating the leveling screws under the footplate. 4.
• If this equipment is modified, appropriate inspection and testing must be conducted to ensure continued safe use of equipment. • Do not directly contact the InBody270 with any other electronic device when the InBody270 is on. This may result in an electric shock. DANGER Caution 참고 주의 참조 주의 • If the InBody270 is not plugged into a grounded outlet, it may cause damage through electric surges or product malfunction. This may affect the test results.
D. Initial Setup 1. The InBody270 automatically starts booting when it is turned on. While booting, it performs a self weight calibration. * While booting (about 5 minutes), make sure there is nothing on top of the footplate. Please do not stand on the footplate, or place objects on the footplate. 2. Press the [Administrator Menu] button on the screen, which appears when no one is on the footplate. Administrator Menu 3. Input the password (default password: 0000) to access the Administrator Menu. 4.
1) Setup: Configure settings and manage data according to the test environment. 01. Date/Time/Units/Country/Language/Password/Volume : Change or modify the InBody’s basic settings. 02. Self Mode/Professional Mode • Self Mode: The examinee takes the InBody Test by entering only his/her height. Throughout the test, instructions and the InBody Information will be shown on screen. • Professional Mode: An examiner is present and guiding the examinee through the InBody Test. 03. N/A 04. N/A 05.
14. Results Sheet Custom Logo : Insert a logo on upper right corner of the printed results sheet. * Please contact InBody for help with uploading or modifying a logo. 15. Printing Alignment : Adjust the alignment of where the results will be printed on the results sheets. 16. Internet Options : You can connect the InBody wirelessly to the network when using the data management software, Lookin’Body. 17. Bluetooth : Connect the InBody270 to data management software Lookin’Body via Bluetooth. 18.
2) Troubleshooting: See additional information on how to use the InBody. Refer to troubleshooting checklist when there are problems that occur during the InBody use/test. 01. Customer Service Information : See the customer service contact information saved under the Setup of the Administrator Menu ‘24. Customer Service Information’. Please contact the customer service if your problem cannot be resolved through the ‘Troubleshooting’ or if you need further inquiries regarding the InBody Test. 02.
E. Connecting Printer, Thermal Printer, Stadiometer, Blood Pressure Monitor, Barcode Reader and SD400 1. Printer In order to print InBody Results Sheets, an InBody270 compatible printer is required. * A list of printers compatible with the InBody270 can be found at http://www.inbodyservice.com. 1) First turn off the InBody270 and then the printer. * You may experience connection issues in connecting the printer to the InBody270 if the InBody is turned on.
3. Stadiometer If a stadiometer is connected to the InBody270, the height values measured by the stadiometer will be sent directly to the InBody270. * Always connect a stadiometer from InBody. 1) First turn off the InBody270 and then the stadiometer. * You may experience connection issues in connecting the stadiometer to the InBody270 if the InBody is turned on. 2) Plug the serial cable provided with the stadiometer to the 9-pin serial port on the rear panel of the InBody270.
5. Barcode Reader If a barcode reader is connected to the InBody270, the ID can be inputted automatically. * If the InBody cannot recognize the barcode reader, please contact InBody. 1) First turn off the InBody270. * You may experience connection issues in connecting the barcode reader to the InBody270 if the InBody is turned on. 2) Plug the USB cable of the barcode reader into the USB HOST port on the rear panel of the InBody270. USB HOST port 3) Turn on the InBody270.
F. Maintenance DANGER Caution 참고 주의 참조 주의 • Do not bend the handles of the hand electrodes or rotate them in the reverse direction beyond its limitation. • Do not extend the handles of the hand electrodes beyond its limitation. • Do not place any objects on the footplate. • Do not apply excessive force on the equipment. • Turn off the equipment if you are not using it for a day or longer. • Do not allow any liquid substances to contact the equipment directly.
II. InBody Test A. Precautionary Steps DANGER Warning 참조 • Individuals with 참고 주의 주의 medical implant devices such as pacemakers, or essential support devices such as patient monitoring systems, must not use this equipment. Safe, low-level currents will flow through the body during the test, which may cause malfunctioning of the device or endanger lives. • Bioelectrical Impedance Analysis (BIA) uses safe low level currents, which are not harmful to the body.
B. Test Instructions 1. Step on the footplate when the screen below is shown. * The screens vary according to the Settings of the Administrator Menu ‘02. Self Mode/Professional Mode’. Professional Mode: An examiner is present and guiding the examinee through the InBody Test. Self Mode: The examinee takes the InBody Test following the instructions that are displayed on screen. 2. Weight measurement begins. 3. Input personal information. * In Professional Mode, enter the ID.
4. Maintain proper posture to take the test. * Refer to ‘C. Test Posture’ for the proper posture. 5. The InBody Test begins. Professional Mode Self Mode 6. When the test is completed, the results will be shown on screen. Professional Mode Self Mode 18 270_manual_ENG_C_170704.indd 18 2017. 7. 4.
C. Test Posture The examinee must maintain proper posture to have accurate test results. * The test will proceed when there is good electrical contact. Your arms must not touch the sides of your body. Keep your arms straight. The thighs should not touch. Hold the hand electrode so that the 4 fingers wrap the surface of the bottom hand electrode, and place the thumb on the oval electrode. Place the heels on the rear sole electrodes. Step on the footplate barefoot. 19 270_manual_ENG_C_170704.
III. Transportation and Storage A. Cautions during Transportation When transporting, have two people keep the InBody270 parallel to the ground. B. Repacking Instructions Once the InBody270 is installed, avoid transporting the equipment. If it must be transported, repack it in the following sequence. 1. Turn off the InBody270. 2. Separate the connected adapter, cords, and cables from the equipment, and cover the InBody270 with the polyethylene foam cover. 3.
IV. Frequently Asked Questions (FAQ) Even if no problems arise from the equipment, users may still have many questions, especially regarding clinical procedures. Few common questions and answers are listed below. If your questions are not answered here, please contact InBody. * Customer contact information can be found under Settings of the Administrator Menu ‘24. Customer Service Information’. A.
Question • My touchscreen is inaccurate or not responsive. Answer • Calibrate the touchscreen under Settings of the Administrator Menu ‘23. Touchscreen Alignment’. * Press firmly to optimize touchscreen response. • If you cannot enter the Administrator Menu due to touchscreen problems, please restart the InBody. The InBody stores the last touchscreen alignment and will automatically recall the previous touchscreen settings.
B. Regarding the InBody Test Some of the more common clinical questions are answered below. If additional questions or more clarification is desired, please contact InBody.
V. Others * The InBody270 is manufactured according to the quality management procedure of InBody. InBody complies with the ISO9001 and ISO13485 which are international quality management systems. * This equipment satisfies the IEC60601-1 (EN60601-1), an international safety standard for electronic medical equipment. This equipment also satisfies the IEC60601-1-2 (EN60601-1-2), an international standard for electromagnetic conformity. A.
2. Footplate ❶ Front sole electrode: The examinee makes contact with this electrode by stepping with the front part of their foot. ❷ Rear sole electrode: The examinee makes contact with this electrode by stepping with the heel of their foot. ❸ Footplate: This is connected to the scale, which measures the examinee’s weight. ❹ Level Indicator: Indicates the current horizontal level of the InBody270. ❺ Hinge: Joins the upper part and lower part of the equipment together.
3. Rear Panel ❶ Speaker: Provides audible indication for test in progress, test complete and successfully saved setting changes. ❷ 9-pin COM serial port (Female, RS-232C): Used for connecting the InBody270 to Lookin’Body installed on a computer, a stadiometer, a blood pressure monitor and SD400. * The InBody270 can be connected to Lookin’Body installed on a computer using one of the ports ❷, ❸, or ❹. * Only compatible with a InBody Stadiometer, Blood Pressure Monitor, and SD400.
B. Safety Information Indicators 9-pin serial port (Female, RS-232C) LAN port (10T Base) USB port Safety Symbols Dangerous High Voltage Warning, Caution BF Type Equipment Adapter 4 Power On Power Off Etc.
Follow instructions for use Suivez les instructions d’utilisation WARNING Electric shock hazard – do not dismantle. Dismantling will void the warranty. AVERTISSEMENT Risque de choc électrique - ne pas démonter. Le démontage annulera la garantie. DANGER Do not use this equipment with electrical medical device such as a pacemaker. Ne pas utiliser cet équipement avec des appareils médicaux électriques comme un stimulateur cardiaque. CAUTION Do not spray any liquid substance directly onto the device.
C. Classification Body Composition Analyzer of Direct Segmental Multi-frequency Bioelectrical Impedance Analysis Method Type of protection against electric shock: Class I • Type of the applied parts: BF Type • Degree of protection against water infiltration: IPX0 • EMC Immunity: Level A • EMC Emission: Level A • Equipment is not suitable for use in the presence of flammable mixtures. • D.
Outputs (InBody Thermal Result Sheet) Total Body Water, Protein, Minerals, Weight, Muscle Mass, Body Fat Mass, Percent Body Fat, BMI, Basal Metabolic Rate, Waist-Hip Ratio, Waist Circumference, Visceral Fat Level, Segmental Lean Analysis(Right Arm, Left Arm, Trunk, Right Leg, Left Leg), Segmental Fat Analysis(Right Arm, Left Arm, Trunk, Right Leg, Left Leg), InBody Score, Fat Control, Muscle Control, Impedance (Each segment and each frequency) Optional Equipment Stadiometer from InBody, Blood pressure mo
E. EMC declaration The InBody270 is intended for use in the electromagnetic environment specified below. The customer or the user of the InBody270 should assure that it is used in such an environment.
Power frequency (50/60 Hz) magnetic field 3 A/m 3 A/m Power frequency magnetic fields should be at levels characteristic of a typical location in a typical commercial or hospital environment. IEC 61000-4-8 NOTE UT is the a.c. mains voltage prior to application of the test level.
270_manual_ENG_C_170704.indd 33 2017. 7. 4.
270_manual_ENG_C_170704.indd 34 2017. 7. 4.
270_manual_ENG_C_170704.indd 35 2017. 7. 4.
www.inbody.com ©2017 InBody Co., Ltd. All rights reserved. BM-ENG-F9-C-170704 270_manual_ENG_C_170704.indd 36 2017. 7. 4.