Manual

Page 11 14439.C 03/2013
Overview of System Components - 2
VerifyNow System User Manual | USA
2.3 Sample Collection Tubes
TheVerifyNowSystemuseswholebloodcollectedinapartial-llvacuumtube.Duringthetest,thetube
is inserted onto the needle in the sample well of the test device - requiring no cap removal, specimen
preparation, or pipetting step. There are several types of sample collection tubes recommended for use
with the system (Table 2-1).
Table 2‑1 Sample Collection Tubes
Collection Tube Description Test
Greiner Bio-One Vacuette tube
(Greiner #454322)
Partialll,2mL,3.2%
sodium citrate. Blue top
Aspirin, PRUTest,
IIb/IIIa (only for use with ReoProabciximab)
Greiner Bio-One High Altitude
Vacuette Tube
(Greiner #454323)
Partialll,2mL,3.2%
sodium citrate. Blue top
Aspirin, PRUTest,
IIb/IIIa (only for use with ReoProabciximab)
Greiner Bio-One Vacuette tube
(Greiner #454244)
Partialll,3mLlithium
heparin. Green top
IIb/IIIa - for use with Integrilin (eptibatide) or ReoPro
(abciximab)
Nipro
(NP-CW0185-1363080)
1.8mL citrate tube Aspirin, PRUTest,
NOTE: The use of sample collection tubes other than those listed may adversely affect test results.
2.4 Quality Control
Accumetrics rigorously tests each VerifyNow System to ensure it meets performance standards. In
addition, the system supports several types of quality controls to verify continued performance during
use.
2.4.1 EQC Device and Storage Slot
The Electronic Quality Control (EQC) is the primary quality control mechanism for the VerifyNow
instrument. It consists of a re-usable device that is inserted by the operator into the test port and is
usedtoperformacomprehensivetestingroutinethatconrmsinstrumentperformance.
Figure 2‑6 EQC Device
When not in use, the EQC device is kept in a storage bay on the right side of the instrument.
NOTE: If the EQC device is lost or damaged, the instrument will be inoperable.