Manual

Page 15 14439.C 03/2013
Reagent Storage and Handling - 4
VerifyNow System User Manual | USA
4 Reagent Storage and Handling
This section describes special considerations for receiving, storing and handling of VerifyNow reagents
and devices.
4.1 Storage and Stability
Store test devices (in their unopened pouches) and quality control materials until use at temperatures
described in Table 4-1. Do not freeze.
Table 4‑1 Recommended Storage Conditions
Test Storage Temperature
VerifyNow Aspirin Test device 2-25ºC (36-77ºF)
VerifyNow PRUTest device 2-25ºC (36-77ºF)
VerifyNow IIb/IIIa Test device 2-8ºC – extended storage
18-25ºC (64-77ºF) - may be stored for up to 8 weeks,
but not beyond the expiration date. When removed
from refrigeration, write the discard date on the
pouches and/or kit box
VerifyNow WQC pellets and diluents 15-30ºC (59-86ºF)
Test devices and quality control materials are suitable for use until the expiration date printed on the
label. Do not use test devices or WQC materials beyond the expiration date. Refer to individual test
instructions for use, kit box and pouch for designated expiration date and recommended storage
conditions.