Quality Controls - 5
VerifyNow System User Manual | USA
4. Remove the EQC device from the storage bay on the right side of the instrument. Open the cover
andusingthengergrip, insert the EQC device into the test device port until it clicks (Figure 5-3).
The instrument will produce two audible beeps. Close the cover to the test port.
NOTE: Do not open the cover until the test has completed.
Figure 5‑3 Insert Device
The instrument will automatically proceed with the EQC test. A countdown screen will display while
electronic quality control is in progress. When the electronic quality control is complete, and the
instrument will prompt the user to completely remove the EQC device from the test port (Figure
5‑4).
EQC
26-Feb-2012
14:57:00
Figure 5‑4 Prompt
5. Open the cover. Remove the EQC device and return it to the storage bay (Figure 5-5). Close the
cover.
Figure 5‑5 Remove Device
6. After the EQC device is completely removed from the port, the instrument will beep and a calculation
screenwilldisplaybrieypriortodisplayingthenalresult.Waitfortheelectronicqualitycontrol
results to display.