Manual

Page 23 14439.C 03/2013
Quality Controls - 5
VerifyNow System User Manual | USA
10. Wait for the image of the tube to display, then insert the WQC sample into the sample well of the
test device with the rubber stopper facing downward, so that the needle fully pierces the stopper.
The instrument will produce two audible beeps when the tube has been fully inserted (Figure 5-12).
26-Feb-2012
WQC P2Y12
14:57:00
Figure 5‑12 Insert Tube
11. Close the cover to the test device port (Figure 5-13).
Figure 5‑13 Close Cover
12. The instrument automatically draws the sample from the vacuum collection tube into the test device,
heats the sample, and proceeds with the analysis of the sample. During this time the screen will
ashthetestdeviceicon,indicatingthatthesampleisprocessing.Donotopenthecover until the
test has completed and a result displays. A calculator will display when the test is near completion
(Figure 5-14).
Figure 5‑14 Test Progress
WQC P2Y12
26-Feb-2012
14:57:00
KEEP COVER CLOSED.
999
WQC P2Y12
26-Feb-2012
123
14:57:00
KEEP COVER CLOSED.
WQC P2Y12
26-Feb-2012
123
14:57:00
CAUTION: The sample is pressurized during parts of the test. Never remove the sample
tube or test device during the test. To abort a test, press the Back key. Wait for the
prompt before removing the test device and tube.
P2Y12