Quality Controls - 5
Page 24 14439.C 03/2013VerifyNow System User Manual | USA
13. When the results display, record the results or print the results, if your instrument has an
attached printer (Figure 5-15).
26-Feb-2012
14:57:00
WQC P2Y12
Figure 5‑15 WQC Level 1 Result
14. Open the cover. Remove the test device and tube together in one piece by grasping the test device
ngergrip and pulling straight up (Figure 5-16). Never separate the tube from the test device. Close
the cover. Discard the test device and tube in a biohazard waste in accordance with institutional
guidelines.
Figure 5‑16 Remove Device
15. At the conclusion of the test, the resulting value will be displayed on the screen. Look up the
expected values printed on the test device pouch for Control 1 (Figure 5-17).
Figure 5‑17 Interpreting WQC Level 1 Results
Control Level 1 Range
26-Feb-2012
14:57:00
WQC P2Y12
PRU:
20
16. Compare the WQC result to the Control Level 1 range, and verify the result falls within the expected
range. If the result falls within the expected range, the instrument has passed WQC Level 1. If the