Patient Testing - 6
Page 36 14439.C 03/2013VerifyNow System User Manual | USA
NOTE: If you make a mistake while entering a text character, press the Next key to return to the Patient ID Entry
Screen. Use the ( ) key to move the cursor to the end of the Patient ID entered, then press the ( ) key to erase
character(s) until the mistaken character is deleted. Press the Text
key to go back to the Text Entry Screen and select
the correct character using the directions from above.
Figure 6‑5 Insert Device
6. Open the foil pouch and remove the test devicejustbeforeuse.Holdthetestdevicebythenger
grip.
NOTE: Each test devicehasangergrip. Avoid handling other surfaces of the test device.
7. Remove the needle’s protective sheath by pulling directly up on the sheath. Do not twist the sheath,
as this may remove the needle (Figure 6-6).
Figure 6‑6 Prepare Needle
8. Open the cover.Usingthengergrip, insert the test device into the test device port until it clicks
(Figure 6-7).
Figure 6‑7 Remove Device