Manual

Page 47 14439.C 03/2013
Retrieving Test Results - 7
VerifyNow System User Manual | USA
Feature Description
EQC Result
Displays the result as either a Pass (P) or Fail (F). A Pass result
indicatestheinstrumentwasoperatingwithinitsspecicationlimits.
If there was an earlier problem detected during the EQC process,
then a FAIL result would display.
Operator ID
Displaystheoperatoridentication for the user who performed the
test(whencongured).
Date and Time
Stamp
Displays the date and time by month, day, hour (24 hr clock
format), and minute using the following notation, “MMDD HHMM.”
Parameter
Displays the result for the following measured parameters:
O1: High optical signal level
O2: Low optical signal level
V1: Maximum vacuum level
V2: Maximum leak rate
P1: Maximum pressure level
F1: Mixerfrequencyagcount
F2: Mixer frequency transitions high-to-low
Mask: Identiesfailingcomponentelement
Flagged
Parameter
The failed parameter(s) will be indicated by an arrow ( ) to
the right of the measured value(s). Refer to the Appendix for a
description of parameters.
In addition to viewing earlier results, icons keys enable results to be printed (optional).
Icon Description
Print Result
Scroll to the selected record, and select this icon to print the
selected EQC results.
Print All
Results
Scroll backwards through the results using the ( )key, and
select this icon to print all EQC results in the log that are
forward of the cursor position.
Back Select this icon to return to the previous screen.