Manual

Retrieving Test Results - 7
Page 50 14439.C 03/2013VerifyNow System User Manual | USA
7.4 View Event Log
The instrument stores the last 50 error/attention events and the last 11 alarms in its memory. The
instrument Event log displays the errors and alarms that have occurred during use. They are arranged
chronologically. Use the following procedure to view the Event log (Figure 7-7):
1. From the main menu, enter Operator ID and Password, if required. Use the keypad to enter
the Operator ID number at the prompt. Press the Tab key to advance to the Password prompt to
enter the password number.
2. Press the Maintenance key.
3. Press the Next key two times. Refer to the Appendix for information on navigating within the
Maintenance section.
4. Press the Event Log key to retrieve the appropriate usage information.
See Section 7.4.1 for a description of the event log.
7.4.1 Alarm and Error/Attention Log
The cursor is positioned on the most recent event (e.g. errors, alarms, alerts). Navigate up and down an
individual list using the ( , ) keys or between lists (Alarms, Errors) using the Tab key.
26-Feb-2012
14:57:00
CODE OpId MMDDHHMM
021IL 6789 06061230
24W 7779807042359
36A 9876508110900
090 333 20090915
ALL
Alarms
Errors
Figure 7‑7 Alarm and Error/Attention Log
Error/Attention List
Alarms List
Print Result
Print All Results
Back
Feature Description
Alarms List
Lists the most recent of alarms logged during operation. This includes the alarm
code,operatoridentication (if enabled), and the date and time stamp in month, day,
hour (24 hr clock format), and minute, in “MMDDHHMM” format.
Error/Attention
List
Lists the most recent errors or alerts (attentions) logged during operation. This
includes the error/alertcode,operatoridentication (if enabled), and the date and
time stamp in month, day, hour (24 hr clock format), and minute, in “MMDDHHMM”
format.
In addition to viewing earlier results, icons keys enable results to be printed (optional) or another test
started.