Manual

Instrument Settings - 8
Page 60 14439.C 03/2013VerifyNow System User Manual | USA
NOTE: When the established interval has elapsed, the user is locked out from running a patient test until
the EQC test has been successfully completed.
The instrument’s default setting is daily. Use the following procedure to adjust the setting based on your
institution’s preference:
1. From the main menu, enter Operator ID and Password, if required. Use the keypad to enter
the Operator ID number at the prompt. Press the Tab key to advance to the Password prompt to
enter the password number.
2. Press the Maintenance key.
3. Press the Parameter key. Refer to the Appendix for information on navigating within the
Maintenance section.
26-Feb-2012
14:57:00
Version: X00191
Use PtId> Yes
PtId Lth:3
S/N: 895
Time out: 2 3 5
EQC Freq> Daily
Use Printer? Yes
Network 19200
Figure 8‑7 EQC Frequency
Save Setting
EQC Frequency
Back
4. Press the Tab key 6 times to advance the cursor to the EQC Frequency setting. At the EQC Freq
prompt, press either the ( ) or ( ) key to change the setting to 6 hours, 8 hours, 12 hours, daily,
weekly, or monthly.
5. Press the Save key to keep the new setting and return to the Maintenance menu. Press the
Back key twice to return to the Main Menu without saving the changes.
8.8 Enable Printer
The VerifyNow printer is an optional accessory to the instrument. The printer enables an operator to
print test results, QC results, and instrument usage statistics. To enable the VerifyNow Instrument
to output this information to the printer, the printer feature must be set to “Yes.” Use the following
proceduretoconguretheinstrumenttosupporttheprinteraccessory:
1. From the main menu, enter Operator ID and Password, if required. Use the keypad to enter
the Operator ID number at the prompt. Press the Tab key to advance to the Password prompt to
enter the password number.
2. Press the Maintenance key.
3. Press the Parameter key. Refer to the Appendix for information on navigating within the
Maintenance section.