Manual

Cleaning and Maintenance - 9
Page 64 14439.C 03/2013VerifyNow System User Manual | USA
3. Open the cover and insert the cleaning device into the test device port until it clicks (Figure 9-2). The
adhesive strip removes dust and debris on the internal pneumatic port cleaning strip.
NOTE: The instrument does not need to be in any
particular mode to use the cleaning device.
Figure 9‑2 Insert Cleaning Device
4. Leave the cleaning deviceinthetestportforveseconds,butneverlongerthantenseconds.
CAUTION: Leaving the cleaning device in the test port for an extended
period of time can damage the instrument from adhesive adhering to the
internal pneumatic port.
5. Remove the cleaning device completely from the test device port (Figure 9-3). Discard the device.
The cleaning device is for single use only.
Figure 9‑3 Remove Cleaning Device
6. Repeat using another cleaning device if, upon inspection, there is visible dust and debris.
9.2 Clean Exterior Surfaces
Occasionally, a liquid may spill near the instrument or small amounts of dust may build up on the
exterior of the instrument. Periodic cleaning of the exterior surface may be performed to remove
accumulation on the instrument.
Materials
• Lint-free cloth (clean room wipes)
• Non-abrasive cleaner (such as Cidex
®
, pHisoHex
®
, 7X
®
, Super Edisonite
®
,10%householdbleach
solution in water, dish washing detergent, or isopropyl alcohol)