If the same sample needs to be measured more than once click on the Auto Log button. The
instrument will perform the required number of readings with the specified time interval between each
measurement. The results will be displayed on the screen. The batch name entered in the settings
screen will have 01, 02 etc after the name, to signify the order in which the results were generated.
auto logging can be stopped before all the measurements have been completed by clicking on the
Stop button. To recommence auto logging click the Auto Log button again.
If the settings menu is re-entered following the sample measurement it is not possible to change the
wavelength. Alternatively a new concentration tab can be opened and the required settings can be
entered.
4.2.2 Calibration and sample measurement using a standard
Once the settings have been
entered the zero calibration
should be performed.
Insert the cuvette containing
the blank sample into the
sample chamber and close
the instrument lid. Click on
the Zero button; this sets the
instrument to zero
absorbance.
The calibration status will
show the zero calibration has
been performed and the unit
is ready to measure the
Standard. The standard
button is now active. Insert
the cuvette containing the
standard sample into sample
chamber and close the
instrument lid.