6.3 CALIBRATION
Once the settings have been entered and if required a new quantitation curve has been created, the
zero calibration should be performed.
Insert the cuvette containing the blank sample into the sample chamber and close the instrument lid.
Click on the Zero button; this sets the instrument to zero absorbance and 100% transmittance.
Once a calibration is complete the
Read and Auto Log buttons are
activated.
If the settings menu is re-entered following the zero calibration it is possible to change the
wavelength. However a warning will be displayed to warn that critical settings are being changed after
the zero calibration has been performed and a quantitation curve has been created.