software Manual

58
To continue click Yes, to cancel click No or Cancel.
If the wavelength is changed then the zero calibration must be performed again before a sample can
be measured. Alternatively a new time scan & kinetics tab can be opened and the required settings
can be entered.
7.3 SAMPLE MEASUREMENT
Insert the cuvette containing the
sample into the sample chamber and
close the instrument lid.
Click on the Scan button; the
instrument initiates the scan and the
live spectrum scan is displayed on
the screen. The results are also
plotted in a table. To stop the scan
before completion click on the Stop
button. To restart the scan click on
the Repeat button the scan is
restarted from the beginning.
If the same sample needs to be measured more than once click on the Auto Log button. The
instrument will perform the required number of readings with the specified time interval between each
measurement. Unlike with photometrics, concentration and quantitation the auto logged results will
not be displayed on the same screen. For each repeat measurement required a new tab is
automatically opened displaying the new kinetics run and results. Auto logging can be stopped before
all the measurements have been completed by clicking on the Stop button. To recommence Auto
logging click on the Auto Log button again.
If the settings menu is re-entered following sample measurement it is not possible to change the
wavelength, run time, time interval or start on level. A new time scan kinetics tab can be opened and
the new settings can be entered.