User Manual

The ssDNA settings menu allows the user to adjust
measurement specific parameters such as the dilution
factor and resolution. Other method settings can be altered
if required, but the concentration units are limited to µg/ml
and mg/ml only.
The ssDNA method settings are locked and any changes
will be lost after exiting the mode. See Section 11 for
full instructions on altering the ssDNA menu options and
settings.
14.4 RNA
The RNA operating menu enables the concentration of
RNA to be determined according to the equation:-
Concentration (μg/ml) = Abs@260nm x 40.0.
The RNA settings menu allows the user to adjust
measurement specific parameters such as the dilution
factor and resolution. Other method settings can be altered
if required, but the concentration units are limited to µg/ml
and mg/ml only.
The RNA method settings are locked and any changes
will be lost after exiting the mode. See Section 11 for
full instructions on altering the RNA method options and
settings.
14.5 OLIGONUCLEOTIDES
The oligonucleotide operating menu enables the
concentration of a typical oligonucleotide to be determined
according to the equation:-
Concentration (μg/ml) = Abs@260nm x 30.0.
The oligonucleotide settings menu allows the user to adjust
measurement specific parameters such as the dilution
factor and resolution. Other method settings can be altered
if required, but the concentration units are limited to µg/ml
and mg/ml only.
The oligonucleotide method settings are locked and any
changes will be lost after exiting the mode. See Section 11
for full instructions on altering the oligonucleotide method
options and settings.
Operating Menu
Operating Menu