DC current accuracy
Instrument
Test voltage
DUT current
Min
Measured
current
Max
0% Test Current
2200-20-5
10 V
0.0 A
-0.002 A
__________
0.002 A
2200-30-5
15 V
0.0 A
-0.0026 A
__________
0.0026 A
2200-32-3
16 V
0.0 A
-0.002 A
__________
0.002 A
2200-60-2
30 V
0.0 A
-0.0015 A
__________
0.0015 A
2200-70-1
36 V
0.0 A
-0.001 A
__________
0.001 A
25% Test Current
2200-20-5
10 V
1.25 A
1.247375 A
__________
1.252625 A
2200-30-5
15 V
1.25 A
1.246875A
__________
1.253125A
2200-32-3
16 V
0.75 A
0.747625 A
__________
0.752375 A
2200-60-2
30 V
0.625 A
0.6231875 A
__________
0.6268125 A
2200-70-1
36 V
0.3 A
0.29885 A
__________
0.30115 A
50% Test Current
2200-20-5
10 V
2.5 A
2.49675 A
__________
2.50325 A
2200-30-5
15 V
2.5 A
2.49625 A
__________
2.50375 A
2200-32-3
16 V
1.5 A
1.49725 A
__________
1.50275 A
2200-60-2
30 V
1.25 A
1.247875 A
__________
1.252125 A
2200-70-1
36V
0.6 A
0.5987 A
__________
0.6013 A
75% Test Current
2200-20-5
10 V
3.75 A
3.746125 A
__________
3.753875 A
2200-30-5
15 V
3.75 A
3.745625 A
__________
3.754375 A
2200-32-3
16 V
2.25 A
2.246875 A
__________
2.253125 A
2200-60-2
30 V
1.875 A
1.8725625 A
__________
1.8774375 A
2200-70-1
36 V
0.9 A
0.898553 A
__________
0.90145 A
100% Test Current
2200-20-5
10 V
5 A
4.9955 A
__________
5.0045 A
2200-30-5
15 V
5 A
4.995 A
__________
5.005 A
2200-32-3
16 V
5 A
2.9965 A
__________
3.0035 A
2200-60-2
30 V
2.5 A
2.49725 A
__________
2.50275 A
2200-70-1
36 V
1.2 A
1.1984 A
__________
1.2016 A