User Manual

Appendix B: Calibration Series 2200 Programmable DC Power Supplies Reference Manual
B-28 2200S-901-01 Rev. A/November 2011
9. Enter the voltmeter reading into the Max line for your product.in the table DC current line
regulation (on page B-9).
10. Change the AC Power Source output to the voltage specified in the following table.
DUT voltage selector switch
AC power source voltage
110
115 V
220
230 V
11. Enter the voltmeter reading into the Nom line for your product in the table DC current line
regulation (on page B-9).
12. Calculate the two values: Nom Min and Max Nom. Enter the values into the appropriate
columns in the DC current line regulation (on page B-9) table.
13. Select the largest of the two calculations from step 12 and enter the value into the Largest column
in the DC current line regulation (on page B-9) table.
14. Turn the DUT output off.