User Manual

Series 2200 Programmable DC Power Supplies Reference Manual Section 2: General operation
2200S-901-01 Rev. A/November 2011 2-11
Primary
menu
Secondary menu
First option
Additional
option(s)
Allows you to ...
Immediate
Set rear panel port mode to
immediate trigger.
>RI Mode
Off
Turn remote inhibit (RI) mode off.
Latching
Set remote inhibit mode to latching.
Latching mode turns the output
channel off when the instrument
detects a high-to-low transition (5 V to
0 V).
Live
Set remote inhibit mode to live. Live
mode the output channel off when the
RI input is low (0 V) and on when the
RI input is high (5 V)
>DFI Source
Off
Turn the discrete fault input (DFI)
source mode off.
Ques
Set the DFI source to question
(Ques).
Oper
Set the DFI source to operation
(Oper).
ESB
Set the DFI source to error status
byte (ESB).
RQS
Set the DFI source to request service
(RQS.
>Communication
USBTMC
Set the communications mode to
USBTMC.
GPIB
Address=22
Set the communications mode to
GPIB. Set or edit the GPIB primary
address. See To change GPIB
address settings (on page 3-2) for
more information.
RS232
9600,8,N,1
Set the communication mode to
RS232. Set or edit the RS232
address. USBTCM and GPIB are the
recommended communication
modes.