User Manual

Series 2200 Programmable DC Power Supplies Reference Manual Section 4: Command reference
2200S-901-01 Rev. A/November 2011 4-5
FETCh:VOLTage[:DC]? (query only)
This command returns the last measured output voltage stored in
the communications buffer of the power supply. A new
measurement is not initiated by this command.
Group
Measurement
Syntax
FETCh:VOLTage[:DC]?
Related
Commands
MEASure:VOLTage[:DC]? (query only) (on page 4-7)
Returns
<NR2> is the measured output voltage in volts.
Examples
FETC:VOLT? might return 5.0011, which would be the measured
voltage across the power supply outputs in volts.
FETCh[:SCALar]:POWer? (query only)
This command returns the last measured output current stored in
the communications buffer of the power supply. A new
measurement is not initiated by this command. The power
calculation in the instrument is performed approximately every 100
ms. Insure that the voltage and current are stable longer than this
for good results.
Group
Measurement
Syntax
FETCh[:SCALar]:POWer?
Returns
<NR2> is the measured output power in watts.
Examples
FETCh:POW? might return 6.01667, which would be the power
measured at the output of t he power supply in watts.