User Manual

Series 2200 Programmable DC Power Supplies Reference Manual Section 4: Command reference
2200S-901-01 Rev. A/November 2011 4-7
MEASure:VOLTage[:DC]? (query only)
This command initiates and executes a new voltage measurement,
and returns the measured output voltage of the power supply.
Group
Measurement
Syntax
MEASure:VOLTage[:DC]?
Related
Commands
FETCh:VOLTage[:DC]? (query only) (on page 4-5)
Returns
<NR2> is the measured output voltage in volts.
Examples
MEAS:VOLT? might return 5.0011 which would be the voltage
measured across the power supply output in volts.
*OPC
This command configures the instrument to generate an operation
complete message by setting bit 0 of the Standard Event Status
Register (SESR) when all pending commands that generate an
OPC message are complete.
The query command places the ASCII character "1" into the output
queue when all such OPC commands are complete.
Group
Synchronization
Syntax
*OPC
*OPC?
Examples
*OPC? might return 1 to indicate that all pending OPC operations
are finished.