User Manual

Series 2200 Programmable DC Power Supplies Reference Manual Section 4: Command reference
2200S-901-01 Rev. A/November 2011 4-9
*RCL (no query form)
Restores the state of the power supply from a copy of its settings
stored in the setup memory. The settings are stored using the *SAV
command. If the specified setup memory is deleted, this command
causes an error.
Group
Save and Recall
Syntax
*RCL <NR1>
Related
Commands
*SAV (no query form) (on page 4-10)
Arguments
<NR1> is an integer value in the range from 0 to 40 and specifies
the location of setup memory.
Examples
*RCL 3
sets the power supply to settings stored in memory location 3.
*RST (no query form)
This command resets the power supply to default settings, but
does not purge any stored settings.
Sending the *RST command does the following:
Returns the power supply settings to the defaults described in
Default setup (on page 2-34).
Clears the pending operation flag and associated operations
The *RST command does not change the following items:
State of the USB or GPIB interface
Calibration data that affects device specifications
Current GPIB power supply address
Stored settings
Output queue
Service Request Enable Register settings
Standard Event Status Enable Register settings
Power-On Status Clear flag setting front-panel LOCK state
Group
Status
Syntax
*RST