www.keithley.com Series 2200 Programmable DC Power Supplies Reference Manual 2200S-901-01 Rev.
Series 2200 Programmable DC Power Supplies Reference Manual © 2011, Keithley Instruments, Inc. Cleveland, Ohio, U.S.A. All rights reserved. Any unauthorized reproduction, photocopy, or use the information herein, in whole or in part, without the prior written approval of Keithley Instruments, Inc. is strictly prohibited. All Keithley Instruments product names are trademarks or registered trademarks of Keithley Instruments, Inc.
Safety Precautions The following safety precautions should be observed before using this product and any associated instrumentation. Although some instruments and accessories would normally be used with nonhazardous voltages, there are situations where hazardous conditions may be present. This product is intended for use by qualified personnel who recognize shock hazards and are familiar with the safety precautions required to avoid possible injury.
Do not exceed the maximum signal levels of the instruments and accessories, as defined in the specifications and operating information, and as shown on the instrument or test fixture panels, or switching card. When fuses are used in a product, replace with the same type and rating for continued protection against fire hazard. Chassis connections must only be used as shield connections for measuring circuits, NOT as safety earth ground connections.
Table of Contents Introduction ............................................................................................................... 1-1 Welcome .............................................................................................................................. 1-1 Extended warranty ............................................................................................................... 1-1 Contact information ........................................................................
Table of Contents Series 2200 Programmable DC Power Supplies Reference Manual Command and query structure .................................................................................................. 3-4 Command entry......................................................................................................................... 3-5 Command groups ................................................................................................................. 3-9 Command reference .........
Series 2200 Programmable DC Power Supplies Reference Manual Table of Contents [SOURce:]OUTPut[:STATe] ............................................................................................... 4-22 [SOURce:]OUTPut:TIMer:DELay ....................................................................................... 4-23 [SOURce:]OUTPut:TIMer[:STATe] .................................................................................... 4-24 [SOURce:]VOLTage[:LEVel] .........................................
Table of Contents Series 2200 Programmable DC Power Supplies Reference Manual Next Steps ................................................................................................................. 6-1 Additional Series 2200 information ...................................................................................... 6-1 Maintenance ..............................................................................................................A-1 Cleaning .........................................
Series 2200 Programmable DC Power Supplies Reference Manual Table of Contents Index ........................................................................................................................... I-1 2200S-901-01 Rev.
Section 1 Introduction In this section: Welcome .................................................................................. 1-1 Extended warranty ................................................................... 1-1 Contact information .................................................................. 1-1 CD-ROM contents .................................................................... 1-2 Key features .............................................................................
Section 1: Introduction Series 2200 Programmable DC Power Supplies Reference Manual CD-ROM contents A CD-ROM is shipped with each Series 2200 order. The Series 2200 Product Information CD-ROM contains: Reference Manual: Includes advanced operation topics and maintenance information. Programmers looking for a command reference, and users looking for an in-depth description of the way the instrument works (including troubleshooting and optimization), should refer to the Reference Manual.
Series 2200 Programmable DC Power Supplies Reference Manual Section 1: Introduction Optional accessories Optional accessory Keithley part number IEEE-488.2 Interface Board for the PCI Bus USB Cable Type A to B 1m (3.2 ft) USB Cable Type A to B 3m (10 ft) Fixed Rack Mount Kit Double Shielded Premium IEEE-488 Interface Cables, 0.5m (1.6 ft) Double Shielded Premium IEEE-488 Interface Cables, 1m (3.2 ft) Double Shielded Premium IEEE-488 Interface Cables, 2m (6.
Section 1: Introduction Series 2200 Programmable DC Power Supplies Reference Manual Electrical ratings for the power connection Model Line selector switch setting Frequency Fuse rating Max power 2200-20-5 110 V / 220 V 50 / 60 Hz 350 VA 2200-30-5 110 V / 220 V 50 / 60 Hz 2200-32-3 110 V / 220 V 50 / 60 Hz 2200-60-2 110 V / 220 V 50 / 60 Hz 2200-72-1 110 V / 220 V 50 / 60 Hz For 110 V: 5 A TH, 250 V For 220 V: 2.5 A TH, 250 V For 110 V: 5 A TH, 250 V For 220 V: 2.
Section 2 General operation In this section: Operating requirements ........................................................... 2-1 Installing the system ................................................................. 2-3 Operating basics ...................................................................... 2-5 Operation examples ............................................................... 2-29 Default setup ..........................................................................
Section 2: General operation Series 2200 Programmable DC Power Supplies Reference Manual Figure 1: Place bottom of instrument on flat stable surface 2. Before operating, ensure that the ambient temperature is between +0 °C and +40 °C (+32 °F to +104 °F). To ensure proper cooling, keep the front, sides, and rear of the instrument clear of obstructions Use properly rated load wires.
Series 2200 Programmable DC Power Supplies Reference Manual Section 2: General operation Installing the system This section contains information on how to install your Model 2200 power supply. Unpack the instrument and check that you received all the items listed as standard accessories. Check that you also received any other accessories that you ordered with your instrument.
Section 2: General operation Series 2200 Programmable DC Power Supplies Reference Manual Voltage output check To verify basic voltage functions without a load, follow these steps. 1. Remove all leads from the output connectors. 2. Turn on the power supply. 3. Press Shift and Menu (1). >Default Set shows on the display. 4. Press Enter to bring up the default settings options. No and Yes appear on the display. 5. Press the right arrow key to select Yes. Press Enter to enable the default settings. 6.
Series 2200 Programmable DC Power Supplies Reference Manual Section 2: General operation Operating basics Front panel at a glance The front panel of the Keithley Instruments Series 2200 contains the following items: Power switch Display Arrow keys The navigation wheel Output connectors Keys Number keys You can use the keys, display, and the navigation wheel configuration menu items and set values.
Section 2: General operation Series 2200 Programmable DC Power Supplies Reference Manual (2) Information During operation, the display provides information about the output (top row) and settings (bottom row) for volts (left column) and current (right column). During configuration, the display shows the configuration menus (bottom row). (3) Arrow keys Up, down, right, and left arrow keys (▲ and ▼) and Enter key.
Series 2200 Programmable DC Power Supplies Reference Manual Section 2: General operation Key descriptions The keys provide front panel control and configuration. See Front panel at a glance (on page 2-5) for an illustration of each key location. The following tables describe how to use each key. Key descriptions Keys Description V-set Sets the voltage limit. See Setting the voltage limit (on page 2-17) for more information. Sets the current limit.
Section 2: General operation Series 2200 Programmable DC Power Supplies Reference Manual Keypad special operations See the following table for the Series 2200 keypad special operations. Operation Key press Display the instrument firmware version. Display the internal temperature of the instrument. Shift + 8 Shift + . (decimal point) + 1 Perform a factory reset. All setups and lists are erased. Unlock the keypad. Shift + .
Series 2200 Programmable DC Power Supplies Reference Manual Section 2: General operation Configuration menus description Use the configuration menus to configure (setup) the instrument. Sometimes the configuration menus are called the settings menus. See the following table for information on the primary menu selections, secondary menu selections, options, and a brief description of each selection. The following table indicates default settings in bold.
Section 2: General operation Primary menu Secondary menu Series 2200 Programmable DC Power Supplies Reference Manual First option Additional option(s) Allows you to ... the instrument will advance to each step only after it receives a trigger signal. Repeat allows you to set or edit number of steps. List Steps allows you to indicate the number of steps in the list.
Series 2200 Programmable DC Power Supplies Reference Manual Primary menu Secondary menu First option Section 2: General operation Additional option(s) Immediate >RI Mode Off Latching Live >DFI Source Off Ques Oper ESB RQS >Communication 2200S-901-01 Rev. A/November 2011 USBTMC GPIB Address=22 RS232 9600,8,N,1 Allows you to ... Set rear panel port mode to immediate trigger. Turn remote inhibit (RI) mode off. Set remote inhibit mode to latching.
Section 2: General operation Series 2200 Programmable DC Power Supplies Reference Manual Upper display messages The instrument shows messages on the upper row of the display to indicate output operating and error conditions. See Configuring the instrument for your application (on page 2-16) for general information about these messages. See the following tables for the messages and a description of what the messages indicate. Operating conditions Messages Description OFF Power supply output is off.
Series 2200 Programmable DC Power Supplies Reference Manual Section 2: General operation Rear panel at a glance The rear panel of the Series 2200 contains the following items: Cooling vents Factory test port USB device port GPIB IEEE-488 port 110 V/220 V power connector 12-pin connector; includes remote sense Shorting clip. Use this for proper regulation when not using remote sense. This connector comes installed. It shorts the + drive to the + sense and the - drive to the - sense.
Section 2: General operation Series 2200 Programmable DC Power Supplies Reference Manual (2) Factory test port Do not use the factory test port. Unauthorized use of the factory test port could damage this product. (3) USB device port The USB 2.0 high-speed device port supports remote control and data transfer to a PC. (4) General purpose interface bus (GPIB) connector The GPIB supports remote control and data transfer to a PC.
Series 2200 Programmable DC Power Supplies Reference Manual Section 2: General operation 110 V/220 V selector switch The 110 V/220 V selector switch is located on the bottom of the instrument as shown in the following figure. Figure 5: Selector switch Front panel operation Within a few seconds after turning on, the power supply displays the actual voltage and current output value on the display’s top line and the voltage and current settings on the bottom line.
Section 2: General operation Series 2200 Programmable DC Power Supplies Reference Manual Configuring the instrument for your application The configuration menus (see Configuration menus description (on page 2-9)) includes settings, like OVP and Max Volt, that determine the maximum voltage output of the power supply. The menu system also includes settings, like Save Last and Output Recall, that determine how the instrument initializes itself when it is turned on.
Series 2200 Programmable DC Power Supplies Reference Manual Section 2: General operation Initializing to the default setup Use the default setup to get the power supply into a default initial state: 1. Remove all leads from the output connectors. 2. Turn on the power supply. 3. Press Shift and Menu (1) to access the configuration menus. >Default Set shows on the display. 4. Press Enter to bring up the default settings menu. No and Yes show on the display. 5.
Section 2: General operation Series 2200 Programmable DC Power Supplies Reference Manual Save and recall setups You can store up to 40 different setups in setup memory locations (1 to 40). Each setup includes the set voltage limit, the set current limit, and the protection menu settings. When shipped from the factory, setup memories 1 through 40 are empty. Save and recall setups as follows: To save a setup: 1.
Series 2200 Programmable DC Power Supplies Reference Manual Section 2: General operation Setting the overvoltage protection The OVP Set control turns overvoltage protection (OVP) on or off. If overvoltage protection is turned on and the instrument senses a voltage level above the threshold set for over-voltage control, the instrument will turn off the output and clamp the voltage on the output to below one volt. To set the overvoltage protection: 1.
Section 2: General operation Series 2200 Programmable DC Power Supplies Reference Manual Recall the power supply ON/OFF output state at power on The Output Recall control determines the On or Off output state after the power supply is powered on. If you select On, the power supply will restore the state of the output to that which was in use before the power was last turned off.
Series 2200 Programmable DC Power Supplies Reference Manual Section 2: General operation Recall the power supply operating parameters at power on The Save Last control determines whether the power supply saves its most recent settings, such as voltage and current, and restores these settings at power on. If you set this control to Off then the power supply returns to the default settings at power on.
Section 2: General operation Series 2200 Programmable DC Power Supplies Reference Manual Lock the navigation wheel The Knob Lock control can be used to lock the navigation wheel settings or to select menu items. so it cannot be used to change To lock the navigation wheel : 1. Press Shift and Menu (1) to access the configuration menus. 2. Use the arrow keys, numeric keypad, or navigation wheel 3. Press Enter. to select >User. 4. Use the arrow keys, numeric keypad, or navigation wheel 5. Press Enter.
Series 2200 Programmable DC Power Supplies Reference Manual Section 2: General operation Using remote sense Use remote sensing to regulate the output voltage at the device under test. This feature lets you compensate for the voltage drop in the leads between the power supply front-end terminals and the device-under-test. To set the remote sensing mode: 1. Remove any jumpers or shorting clip on the rear panel terminal strip connectors between DRIVE + and SENSE + and between DRIVE – and SENSE –. 2.
Section 2: General operation Series 2200 Programmable DC Power Supplies Reference Manual Define a list of voltage and current steps The Edit List feature lets you create up to seven sequences of steps, each with a voltage level, current level, and time duration. To define and save a sequence, do the following: 1. Press Shift and Menu (1) to access the configuration menus. The instrument displays: >Default Set 2.
Series 2200 Programmable DC Power Supplies Reference Manual Section 2: General operation 7. Press Enter. The instrument displays: Repeat 1 8. Use the keypad or navigation wheel to select the number of times to repeat the list. In this example, 2 is used. This means that the instruction will run through the list of steps two times before stopping. The instrument displays: Repeat 2 9. Press Enter. The instrument displays: List Steps 4 10.
Section 2: General operation Series 2200 Programmable DC Power Supplies Reference Manual Run a list of voltage and current steps To run a defined list of voltage and current steps, do the following: 1. Set the output voltage to what you want it to be set at before running the list and press the Output On/Off key to turn the output on. 2. Press Shift and List (2). 3. Use the keypad or navigation wheel to select the list to run. 4. Press Enter.
Series 2200 Programmable DC Power Supplies Reference Manual Section 2: General operation Figure 9: IEEE-488 connector To allow many GPIB connections to other instruments, stack the cable connectors. Two screws are located on each connector to ensure that connections remain secure. The figure below shows a typical connection scheme for a multi-unit test system. Figure 10: Series 2200 multiple parallel connections To avoid possible mechanical damage, stack no more than three connectors on any one unit.
Section 2: General operation Series 2200 Programmable DC Power Supplies Reference Manual To connect the Series 2200 to the IEEE-488 bus, line up the cable connector with the connector located on the rear panel. Install and tighten the screws securely, making sure not to overtighten them (see Rear panel at a glance (on page 2-13) for the location of the connections). Connect any additional connectors from other instruments as required for your application.
Series 2200 Programmable DC Power Supplies Reference Manual Section 2: General operation Operation examples You can save tests in 40 memory locations to save up to 40 setups or use the Series 2200 List mode to define custom test sequences of up to 80 steps. This makes it easy to perform tests such as analyzing how your circuit or device under test performs at each voltage level within a range of voltages. This topic provides examples of these operations.
Section 2: General operation Series 2200 Programmable DC Power Supplies Reference Manual Timer example This example sets up the power supply to output 6 V, 1 A for 7 seconds before the output automatically turns off. Keystrokes Perform the following procedure using the keys on the instrument's front panel to save the test setup. To configure the duration of the timer and enable the timer feature: 1. Press Shift and Menu (1) to access the configuration menus. 2.
Series 2200 Programmable DC Power Supplies Reference Manual Section 2: General operation List mode example This example configures a 5 step list to be executed 2 times in continuous mode. The list is saved in location 3. The instrument runs the list upon entering the list mode operation. Step number Voltage level (V) 1 2 3 4 5 2 5 1 7 3 Current level (A) Duration (sec) 5 5 1 3 1 0.001 0.001 1.000 0.001 0.
Section 2: General operation Series 2200 Programmable DC Power Supplies Reference Manual To configure the second step in the list: 1. Input S 002= 5.000 V. Press Enter. The display shows S 002= 0.0000A. 2. Input S 002= 5.000 A as the current limit. Press Enter. The display shows S 002= 1.0000A. 3. Input S 002= 0.001 S as the duration of the step. The display shows S 002= Enter. The display shows S 003= 0.000V. 0.001S. Press This completes the configuration of the second step.
Series 2200 Programmable DC Power Supplies Reference Manual Section 2: General operation To stop running the list: 1. Press Output On/Off to turn off the output 2. Press ESC to exit the list mode.
Section 2: General operation Series 2200 Programmable DC Power Supplies Reference Manual Default setup The following table lists the settings that are restored when you return the power supply to default settings. Default setup Menu or system Defaults VOLT:PROT VOLT:PROT:STAT MAX OFF OUTP OFF VOLT VOLT:RANG CURR OUTP:TIM OUTP:TIM:STAT 1V MAX 0.
Series 2200 Programmable DC Power Supplies Reference Manual Section 2: General operation Terms The following terms may be used in this topic. Clear. This removes data on media/memory before reusing it in a secured area. All reusable memory is cleared to deny access to previously stored information by standard means of access. Erase. This is equivalent to clear. Media storage/data export device.
Section 2: General operation Series 2200 Programmable DC Power Supplies Reference Manual Memory devices The following tables list nonvolatile and volatile memory devices. Detailed procedures to sanitize or clear these devices are shown after the tables.
Series 2200 Programmable DC Power Supplies Reference Manual Section 2: General operation Data export devices The following table lists the data export device. There are no procedures to disable these devices.
Section 3 Remote operation In this section: Section overview ...................................................................... 3-1 Communication interfaces ........................................................ 3-1 Command syntax ..................................................................... 3-3 Command groups..................................................................... 3-9 Section overview This chapter explains how to use commands for remotely controlling the Series 2200 instruments.
Section 3: Remote operation Series 2200 Programmable DC Power Supplies Reference Manual Using the GPIB Connector for IEEE-488 (GPIB) operation. Use a shielded cable, such as the Keithley Instruments Model 7007-1 or Model 7007-2. Connect a GPIB cable from the Series 2200 instrument to your PC.
Series 2200 Programmable DC Power Supplies Reference Manual Section 3: Remote operation Command syntax You can control the power supply through the GPIB or USB interface using commands and queries. This section describes the syntax these commands and queries use and the conventions the power supply uses to process them. The commands and queries themselves are listed by group (see Command groups (on page 3-9)) and alphabetically (see Commands listed in alphabetical order (on page 4-2)).
Section 3: Remote operation Series 2200 Programmable DC Power Supplies Reference Manual Command and query structure Commands consist of set commands and query commands (usually simply called commands and queries). Commands change power supply settings or perform a specific action. Queries cause the power supply to return data and information about its status. Most commands have both a set form and a query form.
Series 2200 Programmable DC Power Supplies Reference Manual Section 3: Remote operation Commands Commands cause the power supply to perform a specific function or change one of its settings. Commands have the structure: [:][[]...] A command header is made up of one or more mnemonics arranged in a hierarchical or tree structure. The first mnemonic is the base or root of the tree and each subsequent mnemonic is a level or branch off of the previous one.
Section 3: Remote operation Series 2200 Programmable DC Power Supplies Reference Manual SCPI commands and queries The power supply uses a command language based on the Standard Commands for Programmable Instruments (SCPI) standard. The SCPI standard was created by a consortium to provide guidelines for remote programming of instruments. These guidelines provide a consistent programming environment for instrument control and data transfer.
Series 2200 Programmable DC Power Supplies Reference Manual Section 3: Remote operation Parameter types Many power supply commands require parameters. Parameters are indicated by angle brackets, such as . There are several different types of parameters, as listed in the following table. The parameter type is listed after the parameter. Some parameter types are defined specifically for the arbitrary/ function generator command set and some are defined by SCPI.
Section 3: Remote operation Series 2200 Programmable DC Power Supplies Reference Manual Chaining commands and queries You can chain several commands or queries together into a single message. To create a chained message, first create a command or query, then add a semicolon (;), and finally add more commands or queries and semicolons until you are done. If the command following a semicolon is a root node, precede it with a colon (:).
Series 2200 Programmable DC Power Supplies Reference Manual Section 3: Remote operation You can use upper case, lower case, or a mixture of both. No embedded spaces are allowed between or within nodes. :SOURCE:FREQUENCY 10MHZ is the same as :source:frequency 10mhz and :SOURCE:frequency 10MHZ No embedded spaces are allowed between or within nodes.
Section 4 Command reference In this section: Commands listed in alphabetical order .................................... 4-2 *CLS (no query form) ............................................................... 4-2 CONFigure:SOUNd[:STATe].................................................... 4-3 *ESE ........................................................................................ 4-3 *ESR? (query only) ..................................................................
Section 4: Command reference Series 2200 Programmable DC Power Supplies Reference Manual SYSTem:KEY......................................................................... 4-37 SYSTem:LOCal (no query form) ............................................ 4-38 SYSTem:POSetup ................................................................. 4-38 SYSTem:REMote (no query form) ......................................... 4-39 SYSTem:RWLock (no query form) .........................................
Series 2200 Programmable DC Power Supplies Reference Manual Section 4: Command reference CONFigure:SOUNd[:STATe] This command turns the key beep sound on or off. Group System Syntax CONFigure:SOUNd[:STATe] {0|1|ON|OFF} CONFigure:SOUNd[:STATe]? Arguments Returns Examples 0 or OFF turns the key beep sound off. 1 or ON turns the key beep sound on. 0|1 CONF:SOUN:OFF CONF:SOUN? might return 0, indicating the key beeper is turned off.
Section 4: Command reference Series 2200 Programmable DC Power Supplies Reference Manual *ESR? (query only) Returns the contents of the Standard Event Status Register (SESR). *ESR? also clears the SESR (since reading the SESR clears it) See Appendix C, Registers (on page C-3) for more information about SESR.
Series 2200 Programmable DC Power Supplies Reference Manual Section 4: Command reference FETCh:VOLTage[:DC]? (query only) This command returns the last measured output voltage stored in the communications buffer of the power supply. A new measurement is not initiated by this command. Group Measurement Syntax FETCh:VOLTage[:DC]? Related Commands Returns Examples MEASure:VOLTage[:DC]? (query only) (on page 4-7) is the measured output voltage in volts. FETC:VOLT? might return 5.
Section 4: Command reference Series 2200 Programmable DC Power Supplies Reference Manual *IDN? (query only) Returns the power supply identification code in IEEE 488.2 notation. Group Status Syntax *IDN? Returns Examples A string that includes , , , and as defined in the following table. Keithley Instruments 2200-XX-X XX-X.XX. X.XX-X.
Series 2200 Programmable DC Power Supplies Reference Manual Section 4: Command reference MEASure:VOLTage[:DC]? (query only) This command initiates and executes a new voltage measurement, and returns the measured output voltage of the power supply. Group Measurement Syntax MEASure:VOLTage[:DC]? Related Commands FETCh:VOLTage[:DC]? (query only) (on page 4-5) Returns is the measured output voltage in volts. Examples MEAS:VOLT? might return 5.
Section 4: Command reference Series 2200 Programmable DC Power Supplies Reference Manual *PSC Sets and queries the power-on status flag that controls the automatic power-on execution of SRER and ESER. When *PSC is true, the SRER and ESER are set to 0 at power-on. When *PSC is false, the current values in the SRER and ESER are preserved in nonvolatile memory when power is shut off and are restored at power-on.
Series 2200 Programmable DC Power Supplies Reference Manual Section 4: Command reference *RCL (no query form) Restores the state of the power supply from a copy of its settings stored in the setup memory. The settings are stored using the *SAV command. If the specified setup memory is deleted, this command causes an error.
Section 4: Command reference Series 2200 Programmable DC Power Supplies Reference Manual *SAV (no query form) Saves the state of the power supply into a specified nonvolatile memory location. Any settings that had been stored previously at the location are overwritten. You can later use the *RCL command to restore the power supply to this saved state.
Series 2200 Programmable DC Power Supplies Reference Manual Section 4: Command reference CURR MIN CURR? might return 2.0000, which would be the current setting in amperes. [SOURce:]DIGital:DATA This command sets the output state of the rear-panel TTL control output and queries the state of the rear-panel TTL control input. When the port mode is DIGITAL, this command is enabled.
Section 4: Command reference Series 2200 Programmable DC Power Supplies Reference Manual [SOURce:]DIGital:FUNCtion This command sets or queries the function of the TTL control lines on the rear panel of the power supply. *RST value is TRIGger.
Series 2200 Programmable DC Power Supplies Reference Manual Section 4: Command reference [SOURce:]FUNCtion:MODE This command configures the power supply to respond to discrete commands or to operate in list mode. Use this command with the FIXed parameter to exit list mode and prepare the instrument to respond to discrete settings changes. Group Source Syntax [SOURce:]FUNCtion:MODE {FIXed|LIST} [SOURce:]FUNCtion:MODE? Arguments FIXed configures the power supply to respond to discrete commands.
Section 4: Command reference Series 2200 Programmable DC Power Supplies Reference Manual [SOURce:]LIST:CURRent[:LEVel] This command sets the current for a list step in units of A or mA. Group Source Syntax [SOURce:]LIST:CURRent[:LEVel] , [SOURce:]LIST:CURRent[:LEVel]? Related Commands [SOURce:]LIST:VOLTage[:LEVel] (on page 4-17) [SOURce:]LIST:WIDth (on page 4-18) Arguments is an integer in the range from 1 to 80, , which is a step number in the active list.
Series 2200 Programmable DC Power Supplies Reference Manual Section 4: Command reference [SOURce:]LIST:MODE This command determines the response of the power supply to a trigger in list mode. Group Source Syntax [SOURce:]LIST:MODE {CONTinuous|STEP}[SOURce:]LIST:MODE? Related Commands [SOURce:]LIST:COUNt (on page 4-13), [SOURce:]LIST:STEP (on page 4-16) Arguments CONTinuous: Sets the power supply to execute the entire list in response to a trigger.
Section 4: Command reference Series 2200 Programmable DC Power Supplies Reference Manual [SOURce:]LIST:SAVe (no query form) This command saves the active list file to a storage location in nonvolatile memory. Group Source Syntax [SOURce:]LIST:SAVe Related Commands [SOURce:]LIST:RCL (no query form) (on page 4-15) Arguments is an integer in the range from 1 to 8, representing a list storage location. Examples LIST:SAV 4 saves the active list in list storage location 4.
Series 2200 Programmable DC Power Supplies Reference Manual Section 4: Command reference [SOURce:]LIST:VOLTage[:LEVel] This command sets the voltage level of a specified step in a list in units of V or mV. Group Source: List Syntax [SOURce:]LIST:VOLTage[:LEVel] , Arguments is an integer in the range from 1 to 80, which is a step number in the active list. ::= where: is a flexible decimal that sets the voltage level fro the step.
Section 4: Command reference Series 2200 Programmable DC Power Supplies Reference Manual [SOURce:]LIST:WIDth This command sets the duration of a specified step in a list. Group Source Syntax [SOURce:]LIST:WIDth ,{|MIN|MAX} [SOURce:]LIST:WIDth? Related Commands [SOURce:]LIST:VOLTage[:LEVel] (on page 4-17) [SOURce:]LIST:CURRent[:LEVel] (on page 4-14) Arguments is an integer in the range from 1 to 80, which is a step number in the active list.
Series 2200 Programmable DC Power Supplies Reference Manual Section 4: Command reference [SOURce:]OUTPut:DFI:SOURce This command associates the DFI TTL output on the rear panel with a specified bit in the status byte register (SBR). Once the bit is associated with the DFI signal, the DFI signal will reflect the state of the specified bit. The port needs to be in the DFI or RI mode before using this command. Use the [SOURce:]DIGital:FUNCtion (on page 4-12) command to set the port mode. *RST value is OFF.
Section 4: Command reference Series 2200 Programmable DC Power Supplies Reference Manual [SOURce:]OUTPut:PON[:STATe] This command configures the power supply to power up with its output turned off, or to return the output to the state it was in when it powered down. Group Source Syntax [SOURce:]OUTPut:PON[:STATe] {RST|RCL0} [SOURce:]OUTPut:PON[:STATe]? Arguments RST sets the power supply to power-up with output off.
Series 2200 Programmable DC Power Supplies Reference Manual Section 4: Command reference [SOURce:]OUTPut:RI:MODE This command sets the input mode of the RI (remote inhibit) input pin. In order for this command to be effective, the rear panel TTL input and output must be configured in RI/DFI mode using the [SOURce:]DIGital:FUNCtion (on page 4-12) command.
Section 4: Command reference Series 2200 Programmable DC Power Supplies Reference Manual [SOURce:]OUTPut[:STATe] This command turns the power supply output channel on or off. Group Source Syntax [SOURce:]OUTPut[:STATe] {0|1|ON|OFF} [SOURce:]OUTPut[:STATe]? Related Commands [SOURce:]VOLTage:PROTection[:LEVel] (on page 4-26) [SOURce:]VOLTage:RANGe (on page 4-28) Arguments 0 or OFF turns the power supply output off. 1 or ON turns the power supply output on.
Series 2200 Programmable DC Power Supplies Reference Manual Section 4: Command reference [SOURce:]OUTPut:TIMer:DELay This command sets the time duration of the output timer. When the timer is activated and a duration is set, the output of the power supply will turn off automatically if left on longer than the specified duration. In order to ensure proper operation of the output timer, the timer must be activated using the [SOURce:]OUTPut:TIMer[:STATe] (on page 4-24) command before turning the output on.
Section 4: Command reference Series 2200 Programmable DC Power Supplies Reference Manual [SOURce:]OUTPut:TIMer[:STATe] This command turns the output timer function on and off. When the timer is activated and a duration is set, the output of the power supply will turn off automatically if left on longer than the specified duration. In order to ensure proper operation of the output timer, the power supply must be turned on by using the [SOURce:]OUTPut:TIMer[:STATe] (on page 4-24) command.
Series 2200 Programmable DC Power Supplies Reference Manual Section 4: Command reference [SOURce:]VOLTage[:LEVel] This command sets the voltage value of the power supply. Group Source Syntax [SOURce:]VOLTage[:LEVel] {|MIN|MAX|DEF} [SOURce:]VOLTage[:LEVel]? Related Commands [SOURce:]CURRent[:LEVel] (on page 4-10) Arguments where: ::= is a flexible decimal specifying the voltage setting, ranging from 0 to the maximum nameplate voltage of the power supply.
Section 4: Command reference Series 2200 Programmable DC Power Supplies Reference Manual [SOURce:]VOLTage:PROTection[:LEVel] This command sets the over voltage protection (OVP) threshold level.
Series 2200 Programmable DC Power Supplies Reference Manual Section 4: Command reference [SOURce:]VOLTage:PROTection:STATe This command activates, deactivates, or checks the status of overvoltage protection (OVP). Group Source Syntax [SOURce:]VOLTage:PROTection:STATe {0|1|OFF|ON} [SOURce:]VOLTage:PROTection:STATe? Related Commands [SOURce:]VOLTage:PROTection[:LEVel] (on page 4-26) [SOURce:]VOLTage:RANGe (on page 4-28) Arguments 0 or OFF sets the over voltage protection to off.
Section 4: Command reference Series 2200 Programmable DC Power Supplies Reference Manual [SOURce:]VOLTage:RANGe This command limits the maximum voltage that can be programmed on the power supply. This command corresponds to the front-panel Max Voltage setting that can be found under the Protection submenu. This function is different from OVP, since it cannot turn the output off.
Series 2200 Programmable DC Power Supplies Reference Manual Section 4: Command reference *SRE Service request enable (SRE) sets and queries the bits in the service request enable register (SRER). See Appendix C, Registers (on page C-3) for more information about SRER. Group Status Syntax *SRE *SRE? Related Commands *CLS (no query form) (on page 4-2) *ESR? (query only) (on page 4-4) *PSC (on page 4-8) Arguments is an integer value in the range from 0 to 255.
Section 4: Command reference Examples Series 2200 Programmable DC Power Supplies Reference Manual *SRE 48 sets the bits in the SRER to 00110000 binary. *SRE? might return a value of 32, showing that the bits in the SRER have the binary value 00100000. STATus:OPERation:ENABle This command sets and queries the contents of the operation enable register (OENR).
Series 2200 Programmable DC Power Supplies Reference Manual Section 4: Command reference STATus:OPERation[:EVENt]? (query only) This command returns the contents of the operation event register (OEVR). After executing this command the operation event register is reset. Details about the status registers are available in Appendix C, Registers (on page C-3).
Section 4: Command reference Series 2200 Programmable DC Power Supplies Reference Manual STATus:QUEStionable:ENABle This command sets and queries the contents of the questionable enable register (QENR). The QENR is an eight-bit mask register that determines which bits in the Questionable Event Register (QEVR) will affect the state of the QUES bit in the Status Byte Register (SBR).
Series 2200 Programmable DC Power Supplies Reference Manual Section 4: Command reference STATus:QUEStionable[:EVENt]? (query only) This command returns the contents of the questionable event register (QEVR). After executing this command, the quest event register is reset. Details about the QEVR are available in Appendix C, Registers (on page C-3).
Section 4: Command reference Series 2200 Programmable DC Power Supplies Reference Manual STATus:QUEStionable:NTRansition This command sets the negative transition filter of the questionable event register. The filter contents cause the corresponding bit in the questionable event register to become 1 when the bit value of the questionable condition register transitions from 1 to 0.
Series 2200 Programmable DC Power Supplies Reference Manual Section 4: Command reference STATus:QUEStionable:PTRansition This command sets the positive transition filter of the questionable event register. The filter contents cause the corresponding bit in the questionable event register to become 1 when the bit value of the questionable condition register transitions from 0 to 1.
Section 4: Command reference Series 2200 Programmable DC Power Supplies Reference Manual *STB? (query only) The byte query returns the contents of the status byte register (SBR) using the Master Summary Status (MSS) bit. Refer to Appendix C, Registers (on page C-3) for more information. Group Status Syntax *SRE *SRE? Related Commands *CLS (no query form) (on page 4-2) *ESR? (query only) (on page 4-4) *PSC (on page 4-8) Arguments is an integer value in the range from 0 to 255.
Series 2200 Programmable DC Power Supplies Reference Manual Section 4: Command reference SYSTem:KEY This command can produce the same effect as pressing one of the front-panel buttons. The instrument must be in local mode in order for this command to simulate a front-panel button press. Group System Syntax SYSTem:KEY SYSTem:KEY? Arguments Returns Examples 2200S-901-01 Rev. A/November 2011 is an integer key code (see the following table).
Section 4: Command reference Series 2200 Programmable DC Power Supplies Reference Manual SYSTem:LOCal (no query form) This command sets the power supply for control from the frontpanel. Group System Syntax SYSTem:LOCal Related Commands Examples SYSTem:LOCal (no query form) (on page 4-38) SYSTem:RWLock (no query form) (on page 4-39) SYS:LOC SYSTem:POSetup This command determines how the power supply initializes when its power switch is turned on.
Series 2200 Programmable DC Power Supplies Reference Manual Section 4: Command reference SYSTem:REMote (no query form) This command sets the power supply to remote control mode. Group System Syntax SYSTem:REMote Related Commands SYSTem:LOCal (no query form) (on page 4-38) SYSTem:RWLock (no query form) (on page 4-39) Arguments None. Examples SYSTEM:REMOTE SYSTem:RWLock (no query form) If the power supply is in remote mode, this command locks out the front panel LOCAL button.
Section 4: Command reference Examples Series 2200 Programmable DC Power Supplies Reference Manual SYSTEM:VERSION? might return 1991.0, which is the SCPI version number. *TRG (no query form) This command generates a trigger event. Group Trigger Syntax *TRG Related Commands Examples TRIGger[:IMMediate] (no query form) (on page 4-40) *TRG TRIGger[:IMMediate] (no query form) This command forces an immediate trigger event.
Series 2200 Programmable DC Power Supplies Reference Manual Section 4: Command reference TRIGger:SOURce This command sets the source of trigger events. Group Trigger Syntax TRIGger:SOURce {MANual|IMMediate|EXTernal|BUS} TRIGger:SOURce? Related Commands [SOURce:]DIGital:FUNCtion (on page 4-12) *TRG (no query form) (on page 4-40) TRIGger[:IMMediate] (no query form) (on page 4-40) Arguments MANual: When this value is sent, the power supply will wait for a trigger from the front-panel.
Section 4: Command reference Series 2200 Programmable DC Power Supplies Reference Manual *WAI (no query form) This command prevents the instrument from executing further commands or queries until all pending commands are complete. Group Synchronization Syntax *WAI Examples *WAI Programming examples This topic provides the following programming examples: 4-42 2200S-901-01 Rev.
Series 2200 Programmable DC Power Supplies Reference Manual Section 4: Command reference Example 1 This example is written in the C programming language, the Virtual Instrument Software Architecture (VISA) standard can be used. It demonstrates basic communication with the power supply and error checking. The program establishes communication with the power supply, and puts it into remote mode. It then initializes the voltage and current and turns the output on.
Section 4: Command reference Series 2200 Programmable DC Power Supplies Reference Manual reading CheckError("Unable to read voltage"); ErrorStatus = viPrintf(PWS4000,"MEASURE:CURRENT?\n"); //measure the output current CheckError("Unable to write the device"); ErrorStatus = viScanf(PWS4000,"%f",¤t); //retrieve reading CheckError("Unable to read current"); printf("Measured voltage(V) & current(A): %f,%f \n", voltage, current); SendSCPI("OUTPUT 0"); //turn output off ClosePort(); while(1); //return 0;
Series 2200 Programmable DC Power Supplies Reference Manual Section 4: Command reference { clock_t goal; goal = wait + clock(); while(goal > clock()); } Example 2 This example shows a command sequence that configures a three-step list to execute once in continuous mode. It saves the list in location 1, and then places the instrument in list mode and runs the list. SYSTEM:REMOTE *RST TRIGGER:SOURCE BUS LIST:MODE CONT LIST:COUNT ONCE LIST:STEP 3 LIST:VOLT 1,2V LIST:CURRENT 1,1.
Section 4: Command reference Series 2200 Programmable DC Power Supplies Reference Manual Example 4 This example is written in the C programming language; NI VISA can be used. The program demonstrates setting up and executing a list. It shows how to use the Operation Condition Register to determine when the list has finished executing. // pws_visa.c // #undef MICROSOFT // Change this to #define if Microsoft #ifdef MICROSOFT #include “stdafx.h” #include
Series 2200 Programmable DC Power Supplies Reference Manual Section 4: Command reference SendSCPI("VOLTAGE 0V"); //set the voltage to 0V SendSCPI("OUTPUT 1"); // turn output on SendSCPI("TRIGGER:SOURCE BUS"); //set trigger source to bus /* configure the list*/ SendSCPI("LIST:MODE CONT"); //set the list mode to continuous ErrorStatus = viPrintf(PWS4000,"LIST:COUNT %d\n",List_count); //set the list count CheckError("Unable to set list count"); ErrorStatus = viPrintf(PWS4000,"LIST:STEP %d\n",List_step); //se
Section 4: Command reference Series 2200 Programmable DC Power Supplies Reference Manual strcat(commandString, "\n"); ErrorStatus = viPrintf(PWS4000, commandString); CheckError("Can't Write to Power Supply"); pdest = strchr(commandString, '?'); //Search for query command if (pdest != NULL) { ErrorStatus = viBufRead(PWS4000, (ViBuf)ReadBuffer, sizeof(ReadBuffer), VI_NULL); CheckError("Can't read from driver"); strcpy(pString, ReadBuffer); } } void ClosePort() { viClose(PWS4000); viClose(defaultRM); } void
Section 5 Troubleshooting guide In this section: What to do if the power supply does not turn on ...................... 5-1 What to do if the power supply does not turn on To solve problems you might encounter when turning on the instrument, follow these steps: 1. Verify that there is AC power to the power supply. First, check that the AC power cord is firmly plugged into the power connector on the rear panel of the power supply.
Section 6 Next Steps In this section: Additional Series 2200 information........................................... 6-1 Additional Series 2200 information This manual has prepared you to start using your new Series 2200 instrument. For additional information on the Series 2200, refer to: The Product Information CD-ROM (ships with the product): Contains software tools, drivers, and product documentation. The Keithley Instruments website (http://www.keithley.
Section 6: Next Steps 6-2 Series 2200 Programmable DC Power Supplies Reference Manual 2200S-901-01 Rev.
Appendix A Maintenance In this appendix: Cleaning ...................................................................................A-1 Cleaning Inspect the power supply as often as operating conditions require. To clean the exterior surface, perform the following steps: 1. Remove loose dust on the outside of the power supply with a lint-free cloth. Use care to avoid scratching the display. 2. Use a soft cloth dampened with water to clean the power supply.
Appendix A: Maintenance A-2 Series 2200 Programmable DC Power Supplies Reference Manual 2200S-901-01 Rev.
Appendix B Calibration In this appendix: Performance verification ...........................................................B-1 Performance verification procedures ...................................... B-11 Performance verification This topic contains performance verification procedures. Additional test equipment is required to complete the verification procedures. (See Test equipment (on page B-12). These procedures cover the 2200-20-5, 2200-30-5, 2200-32-3, 2200-60-2, and 2200-72-1 models.
Appendix B: Calibration Series 2200 Programmable DC Power Supplies Reference Manual Test record Model Serial Test Procedure performed by Passed Date Failed Self test DC voltage accuracy with remote sense DC voltage accuracy without remote sense DC voltage readback accuracy DC voltage line regulation DC voltage load regulation DC voltage detection DC current accuracy DC current readback accuracy DC current line regulation DC current load regulation Voltage noise at 7 MHz Voltage noise at 20 MHz Curr
Series 2200 Programmable DC Power Supplies Reference Manual Appendix B: Calibration DC voltage accuracy with remote sense Instrument DUT voltage Test current Min Measured Max 2200-20-5 0V 2.5 A -0.0030 V __________ 0.0030 V 2200-30-5 2200-32-3 2200-60-2 2200-70-1 25% Test Voltage 2200-20-5 2200-30-5 2200-32-3 2200-60-2 2200-70-1 50% Test Voltage 2200-20-5 2200-30-5 2200-32-3 2200-60-2 0V 0V 0V 0V 2.5 A 1.5 A 1.25 A 0.63 A -0.0030 V -0.0030 V -0.0060 V -0.
Appendix B: Calibration Series 2200 Programmable DC Power Supplies Reference Manual DC voltage accuracy without remote sense Instrument DUT voltage Test current Min Measured Max 2200-20-5 0V 2.5 A -0.0030 V __________ 0.0030 V 2200-30-5 2200-32-3 2200-60-2 2200-70-1 25% Test Voltage 2200-20-5 2200-30-5 2200-32-3 2200-60-2 2200-70-1 50% Test Voltage 2200-20-5 2200-30-5 2200-32-3 2200-60-2 0V 0V 0V 0V 2.5 A 1.5 A 1.25 A 0.63 A -0.0030 V -0.0030 V -0.0060 V -0.
Series 2200 Programmable DC Power Supplies Reference Manual Appendix B: Calibration DC voltage readback accuracy Instrument DUT voltage Test current Measured voltage Voltage readout Absolute difference Maximum difference 2200-20-5 0V 2.5 A __________ __________ __________ 0.0030 V 2200-30-5 2200-32-3 2200-60-2 2200-70-1 25% Test Voltage 2200-20-5 2200-30-5 2200-32-3 2200-60-2 2200-70-1 50% Test Voltage 2200-20-5 2200-30-5 2200-32-3 2200-60-2 0V 0V 0V 0V 2.5 A 1.5 A 1.25 A 0.
Appendix B: Calibration Series 2200 Programmable DC Power Supplies Reference Manual DC voltage line regulation Instrument Min line Max line Nom - Min Max - Nom Largest Max value 2200-20-5 __________ __________ __________ __________ __________ 0.
Series 2200 Programmable DC Power Supplies Reference Manual Appendix B: Calibration DC current accuracy Instrument 0% Test Current 2200-20-5 2200-30-5 2200-32-3 2200-60-2 2200-70-1 25% Test Current 2200-20-5 2200-30-5 2200-32-3 2200-60-2 2200-70-1 50% Test Current 2200-20-5 2200-30-5 2200-32-3 2200-60-2 2200-70-1 75% Test Current 2200-20-5 2200-30-5 2200-32-3 2200-60-2 2200-70-1 100% Test Current 2200-20-5 2200-30-5 2200-32-3 2200-60-2 2200-70-1 Test voltage DUT current Min Measured current Max 10 V
Appendix B: Calibration Series 2200 Programmable DC Power Supplies Reference Manual DC current readback accuracy Instrument Test current Measured current Current readout Absolute difference Maximum difference 0A 0A 0A 0A 0A ________ ________ ________ ________ ________ ________ ________ ________ ________ ________ ________ ________ ________ ________ ________ 0.0020 A 0.0025 A 0.0020 A 0.0015 A 0.0010 A 1.2500 A 1.2500 A 0.7500 A 0.6250 A 0.
Series 2200 Programmable DC Power Supplies Reference Manual Appendix B: Calibration DC current line regulation Instrument Test voltage Min line Max line Nom line Nom - Min Max Nom Largest Max value 2200-20-5 10 V ______ ________ ________ ________ ________ ________ ________ 0.
Appendix B: Calibration Series 2200 Programmable DC Power Supplies Reference Manual Voltage noise at 7 MHz Instrument Voltage test load R1 Measured rms Maximum rms Measured pk - Maximum pk pk pk 2200-20-5 5 ________ 1 mV ________ 3 mV 2200-30-5 10 ________ 1 mV ________ 4 mV 2200-32-3 12 ________ 1 mV ________ 4 mV 40 ________ 1 mV ________ 5 mV ________ 65 * Load R must be rated at 100 W minimum.
Series 2200 Programmable DC Power Supplies Reference Manual Appendix B: Calibration Performance verification procedures Use the procedures in this topic to verify the performance of the Series 2200 instruments. Performance verification conditions The following conditions must be met before performing these procedures: 1.
Appendix B: Calibration Series 2200 Programmable DC Power Supplies Reference Manual Test equipment Item Connectors 1. Connectors Minimum requirements 2. AC power source 3. Electronic load 4. DC voltmeter 5. High current hook up wire (High current connections are indicated with bold lines in the setup illustrations) Example Connectors for hook up wire assemblies and test resistors, item numbers 5, 6, 9, 10, and 11. (Qty.
Series 2200 Programmable DC Power Supplies Reference Manual Item Connectors Appendix B: Calibration Minimum requirements Example 2200-20-5 5 HL10006E5R000JJ 2200-30-5 10 HL10006E10R00JJ 2200-32-3 2200-60-2 12 40 HL10006E12R00JJ HL10006E40R00JJ 2200-72-1 65 HL10006E65R00JJ 11. Load Resistor for Current Noise testing 12. Oscilloscope 13. Oscilloscope probe 14.
Appendix B: Calibration Series 2200 Programmable DC Power Supplies Reference Manual 4. If any of the messages listed in the following table appear in the display, it indicates that the EEPROM is experiencing the condition associated with the message. 5. If there is no response when you power on the DUT, verify that there is AC power to the power supply, verify the power-line voltage settings, and verify that the correct power-line fuse is installed. If you need more help, contact Keithley. 6.
Series 2200 Programmable DC Power Supplies Reference Manual Appendix B: Calibration Figure 18: DC voltage accuracy with remote sense test setup 2200S-901-01 Rev.
Appendix B: Calibration Series 2200 Programmable DC Power Supplies Reference Manual 4. Power on the DUT. Ensure the warm-up criteria has been met as described in the Performance Verification Conditions. 5. Set the voltmeter as follows: a. Set to measure DC volts. b. Set to auto range. c. Verify that the Math mX+b function is disabled (shift Dcv), assuring that volts are being read. 6. Set the electronic load as follows: a. Set to Constant Current. b.
Series 2200 Programmable DC Power Supplies Reference Manual Appendix B: Calibration Check DC voltage accuracy without remote sense and voltage readback accuracy Equipment required (Item 2) AC Power Source (Item 5) High current hook up wire (bold line connections) (Item 3) Electronic Load (Item 4) DC Voltmeter (Item 6) Low current hook up wire (fine line connections) 1. Set up the equipment as shown in the following figure.
Appendix B: Calibration Series 2200 Programmable DC Power Supplies Reference Manual Figure 20: Voltage accuracy regulation and protection test setup 2. Set the voltmeter as follows: a. Set to measure DC volts. b. Set to auto range. c. Verify that the Math mX+b function is disabled (shift Dcv), assuring that volts are being read. 3. Set the electronic load as follows: a. Set to Constant Current. b.
Series 2200 Programmable DC Power Supplies Reference Manual Appendix B: Calibration Check DC voltage line regulation This check procedure uses the same test setup as the previous procedure. (see Check DC voltage accuracy without remote sense and voltage readback accuracy (on page B-17)). 1. Change the AC Power Source output to the minimum voltage specified in the following table.
Appendix B: Calibration Series 2200 Programmable DC Power Supplies Reference Manual 10. Change the AC Power Source output to match the Voltage Selector switch setting of the DUT 65as shown in the following table. DUT voltage selector switch AC Power Source voltage 110 220 115 V 230 V 11. Enter the voltmeter reading into table for checking DC Voltage Line Regulation under the Nom column for your product. (see DC voltage line regulation (on page B-6).) 12.
Series 2200 Programmable DC Power Supplies Reference Manual Appendix B: Calibration Check DC voltage load regulation This check procedure uses the same test setup as the previous procedure. (see Check DC voltage accuracy without remote sense and voltage readback accuracy (on page B-17)) Ensure the warm-up criteria has been met as described in the Performance Verification Conditions. 1. Set the voltmeter as follows: a. Set to measure DC volts. b. Set to auto range.
Appendix B: Calibration Series 2200 Programmable DC Power Supplies Reference Manual 12. Enter the largest of the three values calculated in step 11 into the Largest column in the DC voltage load regulation (on page B-6) table. B-22 2200S-901-01 Rev.
Series 2200 Programmable DC Power Supplies Reference Manual Appendix B: Calibration Check DC overvoltage detection This check procedure uses the same test setup as the previous procedure. (see Check DC voltage accuracy without remote sense and voltage readback accuracy (on page B-17)) Ensure the warm-up criteria has been met as described in the Performance Verification Conditions. 1. Set the voltmeter as follows: a. Set to measure DC volts. b. Set to auto range. c.
Appendix B: Calibration Series 2200 Programmable DC Power Supplies Reference Manual 8. Enter the voltmeter reading (where OVP triggering occurred) into the DC overvoltage detection (on page B-6) table for checking DC Overvoltage Detection. 9. Reset OVP. a. Press Shift and Menu (1). b. Press the down arrow until Protect is shown and press Enter. c. Press the down arrow until Reset Protect is shown and press Enter. d. Press the right arrow to select Yes and press Enter. e.
Series 2200 Programmable DC Power Supplies Reference Manual Appendix B: Calibration Check DC current and DC current readback accuracy Equipment required (Item 2) AC Power Source (Item 5) High current hook up wire (bold line connections) (Item 3) Electronic Load (Item 4) DC Voltmeter (Item 7) High Accuracy 0.05 Ω Resistor (Item 8) Current Shunt Resistor Cabling 1. Set up the equipment as shown in the following figure.
Appendix B: Calibration Series 2200 Programmable DC Power Supplies Reference Manual 2. Set the voltmeter as follows: a. Set to measure DC volts. b. Set to auto range. c. Set to show amps (instead of volts) by multiplying the voltmeter result by 20. Press Shift DC V (mX+B) Use the arrow key to move the cursor as far to the right as it can go until it is positioned on the ^. Press the up range button once: M=10.00000 Move the arrow key to the 1's digit and up range once: M=20.
Series 2200 Programmable DC Power Supplies Reference Manual Appendix B: Calibration Check DC current line regulation This check procedure uses the same test setup as the previous procedure (see Check DC current and DC current readback accuracy). 1. Change the AC Power Source output to the minimum voltage specified in the following table.
Appendix B: Calibration Series 2200 Programmable DC Power Supplies Reference Manual 9. Enter the voltmeter reading into the Max line for your product.in the table DC current line regulation (on page B-9). 10. Change the AC Power Source output to the voltage specified in the following table. DUT voltage selector switch AC power source voltage 110 220 115 V 230 V 11. Enter the voltmeter reading into the Nom line for your product in the table DC current line regulation (on page B-9). 12.
Series 2200 Programmable DC Power Supplies Reference Manual Appendix B: Calibration Check DC current load regulation This check procedure uses the same test setup as the previous procedure (see Check DC current line regulation (on page B-27)). 1. Set the voltmeter as follows: a. Set to measure DC volts. b. Set to auto range. c. Set to show amps (instead of volts) by multiplying the voltmeter result by 20.
Appendix B: Calibration Series 2200 Programmable DC Power Supplies Reference Manual 8. Enter the voltmeter reading into the DC current load regulation table for checking current load regulation at the reference test voltage for your product. 9. Increase the electronic load to the Maximum test voltage level. Instrument Minimum test voltage 2200 - 20 - 5 2200 - 30 - 5 18.0 V 27.0 V 2200 - 32 - 3 2200 - 60 - 2 2200 - 72 - 1 28.8 V 54.0 V 64.8 V 10.
Series 2200 Programmable DC Power Supplies Reference Manual Appendix B: Calibration Check voltage noise (7 MHz) Equipment required (Item 5) High current hook up wire (bold line connections) (Item 10) Load Resistor (Item 12) Oscilloscope (Item 13) Oscilloscope 1X probe 1. Plug the DUT into your local line power from the mains. 2. Plug the test oscilloscope into the same mains outlet as the DUT.
Appendix B: Calibration Series 2200 Programmable DC Power Supplies Reference Manual 5. Set the oscilloscope as follows: a. 1 mV/division 6. 7. 8. 9. b. 1 M input resistance c. 20 MHz bandwidth (BW) limit d. AC Coupled e. Auto-trigger f. 1 ms/div g. Set to measure Vp-p and VRMS Set the oscilloscope probe to 1X mode. Set the DUT to the 100% FS output voltage. Set the DUT to the 100% FS output current. Turn the DUT output on. Do not touch the load resistor.
Series 2200 Programmable DC Power Supplies Reference Manual Appendix B: Calibration Check voltage noise (20 MHz) Equipment required (Item 5) High current hook up wire (bold line connections) (Item 10) Load Resistor (Item 12) Oscilloscope (Item 15) Coaxial cable (BNC M-M) (Item 16) BNC F-to-Banana 1. Plug the DUT into your local line power from the mains. 2. Plug the test oscilloscope into the same mains outlet as the DUT.
Appendix B: Calibration Series 2200 Programmable DC Power Supplies Reference Manual Figure 23: 20 MHz voltage noise test setup 5. Set the oscilloscope as follows: a. 1 mV/division b. 1 M input resistance c. 20 MHz bandwidth (BW) limit d. AC Coupled e. Auto-trigger f. 1 ms/div g. Set to measure Vp-p and VRMS 6. Set the DUT to the 100% FS output voltage. 7. Set the DUT to the 100% FS output current. 8. Turn the DUT output on. Do not touch the load resistor.
Series 2200 Programmable DC Power Supplies Reference Manual Appendix B: Calibration Check current noise Equipment required (Item 5) High current hook up wire (bold line connections) (Item 11) 1 Load Resistor (Item 12) Oscilloscope (Item 14) Oscilloscope 10X probe 1. Plug the DUT into your local line power from the mains. 2. Plug the test oscilloscope into the same mains outlet as the DUT.
Appendix B: Calibration Series 2200 Programmable DC Power Supplies Reference Manual 5. Set the oscilloscope as follows: a. 1 mV/division b. 1 M input resistance c. 20 MHz bandwidth (BW) limit d. AC Coupled e. Auto-trigger f. 1 ms/div g. Set to measure Vp-p and VRMS 6. Set the DUT to the 100% FS output voltage. 7. Set the DUT to the 100% FS output current. 8. Turn the DUT output on. Do not touch the load resistor. The load resistor may become hot enough to cause burns. 9.
Series 2200 Programmable DC Power Supplies Reference Manual 2200S-901-01 Rev.
Appendix C Status model In this appendix: Status and events .................................................................... C-1 Status reporting structure ........................................................ C-1 Registers ................................................................................. C-3 Status and events This appendix provides details about the status information and events the power supply reports.
Appendix C: Status model Series 2200 Programmable DC Power Supplies Reference Manual Figure 25: Error and event handling process C-2 2200S-901-01 Rev.
Series 2200 Programmable DC Power Supplies Reference Manual Appendix C: Status model Registers The registers in the event reporting system fall into two functional groups: Status registers contain information about the status of the power supply. They include the standard event status register (SESR). Enable registers determine whether selected types of events are reported to the status registers and the event queue.
Appendix C: Status model Series 2200 Programmable DC Power Supplies Reference Manual SBR bit functions Bit Function 7 (MSB) OSB Operation Status Bit. Indicates that an operation event has occurred. 6 RQS 6 MSS Request Service. Obtained from a serial poll. Shows that the power supply requests service from the GPIB controller. Master Status Summary. Obtained from *STB? query. Summarizes the OSB, ESB, MAV, QSB, and EQS bits in the SBR. 5 ESB Event Status Bit.
Series 2200 Programmable DC Power Supplies Reference Manual Appendix C: Status model Operation condition register (OCR) The operation condition register is made up of eight bits, which note the occurrence of events as shown here. Figure 28: Operation Condition Register (OCR) OCR bit functions Bit Function 7 (MSB) ———— This bit is not used. 6 ———— This bit is not used. 5 RUN Running List. Indicates that the power supply is executing a lis 4 3 TBF CC 2 CV Trace Buffer Full.
Appendix C: Status model Series 2200 Programmable DC Power Supplies Reference Manual SESR bit functions Bit Function 7 (MSB) ———— This bit is not used. 6 ———— This bit is not used. 5 ———— This bit is not used. 4 3 PS RI Protection Shutdown. Remote Inhibit. This bit indicates the state of the Remote Inhibit input. 2 UNR Unregulated. Indicates that the power supply has gone out of regulation. 1 OT 0 (LSB) OV Over Temperature.
Series 2200 Programmable DC Power Supplies Reference Manual Appendix C: Status model Enable registers This topic describes the following four types of enable registers: Event status enable register (ESER) (on page C-7) Service request enable register (SRER) (on page C-7) Operation enable register (OENR) (on page C-8) Questionable enable register (QENR) (on page C-8) ESER, SRER, OENR, and QENR allow you to select which events are reported to the status byte register (SBR).
Appendix C: Status model Series 2200 Programmable DC Power Supplies Reference Manual Operation enable register (OENR) The OENR consists of bits defined exactly the same as bits 0 through 7 in the OEVR. You can use this register to control whether or not the operation status bit (OSB) in the SBR is set when an event occurs and the corresponding OEVR bit is set. Use the STATus:OPERation:ENABle command to set the bits in the OENR. Use the STATus:OPERation:ENABle? query to read the contents of the OENR.
Series 2200 Programmable DC Power Supplies Reference Manual Appendix C: Status model Error/event queue The Event Queue stores detailed information on up to 32 events. When 32 events stack up in the Event Queue, the 32nd event is replaced by event code 350, "Queue Overflow.
Appendix C: Status model Series 2200 Programmable DC Power Supplies Reference Manual Command errors The following table shows the command error messages generated by improper syntax. Check that the command is properly formed and that it follows the rules in the section on command Syntax.
Series 2200 Programmable DC Power Supplies Reference Manual Appendix C: Status model System errors The following table lists the system errors that can occur during power supply operation.
Appendix C: Status model Series 2200 Programmable DC Power Supplies Reference Manual Device dependent errors The following table lists the device errors that can occur during power supply operation. These errors may indicate that the power supply needs repair.
Appendix D Display character codes In this appendix: ASCII code chart .....................................................................
Appendix D: Display character codes Series 2200 Programmable DC Power Supplies Reference Manual ASCII code chart Figure 34: ASCII code chart D-2 2200S-901-01 Rev.
Index A Abbreviating commands, queries, and parameters • 3-7 Accessories • 1-2 optional • 1-3 standard • 1-2 Arrow keys • 2-6 B BNF notation • 3-3 C Calibration • B-1 CC message • 2-12 Chaining commands and queries • 3-8 Cleaning • A-1 Command and query structure • 3-4 Commands Abbreviating • 3-7 Alphabetical order • 4-2 BNF notation • 3-3 Chaining • 3-8 Entry • 3-5 Errors • C-10 Functional groups • 3-9 Message terminators • 3-6 Parameter types • 3-7 Programming examples • 4-42 Queries • 3-5 SCPI commands
Index Series 2200 Programmable DC Power Supplies Reference Manual N Navigation wheel • 2-6 Lock the navigation wheel • 2-22 Numeric keypad • 2-6 Special operations • 2-8 Unlock • 2-8 O On/Off key • 2-6, 2-7 Optional accessories • 1-3 P Parameter types • 3-7 Performance verification Messages and codes • C-9 Overview • B-1 Procedures • B-11 Power Connector • 2-14 On and off • 2-3 Switch • 2-6 Voltage selector switch • 2-15 Shift key • 2-6, 2-7 Standard accessories • 1-2 Status and events • C-1 T Trouble
Specifications are subject to change without notice. All Keithley trademarks and trade names are the property of Keithley Instruments, Inc. All other trademarks and trade names are the property of their respective companies. A G R E A T E R M E A S U R E O F C O N F I D E N C E Keithley Instruments, Inc. Corporate Headquarters • 28775 Aurora Road • Cleveland, Ohio 44139 • 440-248-0400 • Fax: 440-248-6168 • 1-888-KEITHLEY • www.keithley.