Appendix B: Calibration Series 2200 Programmable DC Power Supplies Reference Manual
B-24 2200S-901-01 Rev. A/November 2011
8. Enter the voltmeter reading (where OVP triggering occurred) into the DC overvoltage detection
(on page B-6) table for checking DC Overvoltage Detection.
9. Reset OVP.
a. Press Shift and Menu (1).
b. Press the down arrow until Protect is shown and press Enter.
c. Press the down arrow until Reset Protect is shown and press Enter.
d. Press the right arrow to select Yes and press Enter.
e. Press the Esc key as needed to exit the menu system.