User Manual

Appendix B: Calibration Series 2200 Programmable DC Power Supplies Reference Manual
B-30 2200S-901-01 Rev. A/November 2011
8. Enter the voltmeter reading into the DC current load regulation table for checking current load
regulation at the reference test voltage for your product.
9. Increase the electronic load to the Maximum test voltage level.
Instrument
Minimum test voltage
2200 - 20 - 5
18.0 V
2200 - 30 - 5
27.0 V
2200 - 32 - 3
28.8 V
2200 - 60 - 2
54.0 V
2200 - 72 - 1
64.8 V
10. Enter the voltmeter reading into the DC current load regulation table for checking current load
regulation at the maximum voltage for your product.
11. Calculate the two values: REF Min and Max REF. Enter the values into the appropriate
columns in the DC current load regulation table.
12. Power off the DUT and test equipment.