User Manual

Series 2200 Programmable DC Power Supplies Reference Manual Section 2: General operation
2200S-901-01 Rev. A/November 2011 2-29
Operation examples
You can save tests in 40 memory locations to save up to 40 setups or use the Series 2200 List mode
to define custom test sequences of up to 80 steps. This makes it easy to perform tests such as
analyzing how your circuit or device under test performs at each voltage level within a range of
voltages.
This topic provides examples of these operations. Each example describes how to complete the
operation using both of the following methods:
Front-panel keystrokes
Programming commands
Saving setups example
This example describes how to set up the power supply to output 3.33 V, 1.6 A and save the setup in
location 29 of the 40 setup memory locations available.
Keystrokes
Perform the following steps using the keys on the instrument's front panel to save the test setup.
To configure the power supply:
1. Press V-Set.
2. Use the arrow keys, navigation wheel , or the keypad to input 3.33 V. Press Enter.
3. Press I-Set.
4. Use the arrow keys, navigation wheel , or the keypad to input 1.6 A. Press Enter.
To save the setup:
1. Press Save.
2. Use the arrow keys, navigation wheel , or the keypad to select memory location 29. Press
Enter.
To recall the setup:
Use the following keystrokes to recall the setup saved in memory location 29.
1. Press Recall.
2. Use the arrow keys, navigation wheel , or the keypad to select memory location 29. Press
Enter. The condition previously set displays.
Programming commands
To configure and save the setup:
*RST
SOURCE:FUNC:MODE FIX
SOURCE:VOLT 3.33V
SOURCE:CURR 1.6A
*SAV 29
To recall the setup:
Program the following command:
*RCL 29