Section 4: Command reference Series 2200 Programmable DC Power Supplies Reference Manual
4-10 2200S-901-01 Rev. A/November 2011
*SAV (no query form)
Saves the state of the power supply into a specified
nonvolatile memory location. Any settings that had
been stored previously at the location are overwritten.
You can later use the *RCL command to restore the
power supply to this saved state.
Group
Status
Syntax
*SAV <NR1>
Related
Commands
*RCL (no query form) (on page 4-9)
Returns
<NR1> is an integer value in the range from 1 to 40.
Examples
*SAV 2
saves the settings in memory location 2.
[SOURce:]CURRent[:LEVel]
This command sets the current value of the power
supply in units of A or mA.
Group
Source
Syntax
[SOURce:]CURRent[:LEVel]
{<current>|MIN|MAX|DEF}
[SOURce:]CURRent[:LEVel]?
Arguments
<current>::=<NRf><units>
where:
<NRf> is a flexible decimal between the minimum
current and maximum nameplate current for the power
supply. <units>::={mA|A}
MIN sets the current to the minimum level (0 A).
MAX sets the current to the maximum level.
DEF sets the current to the default level (0.1 A).
Returns
<NR2> is the current setting in amperes.
Examples
CURR 3A
CURR 30mA