Series 2200 Programmable DC Power Supplies Reference Manual Section 4: Command reference
2200S-901-01 Rev. A/November 2011 4-13
[SOURce:]FUNCtion:MODE
This command configures the power supply to respond to discrete
commands or to operate in list mode. Use this command with the
FIXed parameter to exit list mode and prepare the instrument to
respond to discrete settings changes.
Group
Source
Syntax
[SOURce:]FUNCtion:MODE {FIXed|LIST}
[SOURce:]FUNCtion:MODE?
Arguments
FIXed configures the power supply to respond to discrete
commands.
LIST configures the power supply to operate in list mode.
Returns
FIX|LIST
Examples
[SOURCE:]FUNCTION:MODE LIST
[SOURce:]LIST:COUNt
This command configures the number of times the active list will
execute before stopping.
Group
Source
Syntax
[SOURce:]LIST:COUNt {<NR1>|ONCE|REPeat}
[SOURce:]LIST:COUNt?
Related
Commands
[SOURce:]LIST:MODE (on page 4-15)
[SOURce:]LIST:COUNt (on page 4-13)
Arguments
<NR1> is an integer between 2 and 65535. It determines the
number of times the active list will execute.
ONCe: configures the lists to execute once and stop.
REPeat: configures the lists to loop continuously.
Returns
<NR1> is the number of times the active list is set to execute.
Examples
LIST:COUNT 6 would set the active list to execute 6 times before
stopping.