[SOURce:]LIST:VOLTage[:LEVel]
This command sets the voltage level of a specified step in a list in
units of V or mV.
Group
Source: List
Syntax
[SOURce:]LIST:VOLTage[:LEVel] <NR1>,<voltage>
Arguments
<NR1> is an integer in the range from 1 to 80, which is a step
number in the active list.
<voltage>::=<NRf><units>
where:
<NRf> is a flexible decimal that sets the voltage level fro the step.
<units>::={V|mV}
Returns
<NR2>
Examples
LIST:VOLT 1, 3V
LIST:VOLT? 1 might return 3.0000, which would be the voltage
level for step number 1.