User Manual

Section 4: Command reference Series 2200 Programmable DC Power Supplies Reference Manual
4-18 2200S-901-01 Rev. A/November 2011
[SOURce:]LIST:WIDth
This command sets the duration of a specified step in a list.
Group
Source
Syntax
[SOURce:]LIST:WIDth <NR1>,{<duration>|MIN|MAX}
[SOURce:]LIST:WIDth? <NR1>
Related
Commands
[SOURce:]LIST:VOLTage[:LEVel] (on page 4-17)
[SOURce:]LIST:CURRent[:LEVel] (on page 4-14)
Arguments
<NR1> is an integer in the range from 1 to 80, which is a step
number in the active list. <duration>::=<NRf><units>
where:
<NRf> is the time duration of the step <units>::={s|ms}
Returns
<NR2>
Examples
LIST:WIDTH 12, 100ms
LIST:WIDTH? 12 might return 0.1, which would be the duration
of step 12.