User Manual

Series 2200 Programmable DC Power Supplies Reference Manual Section 4: Command reference
2200S-901-01 Rev. A/November 2011 4-31
STATus:OPERation[:EVENt]? (query only)
This command returns the contents of the operation event register
(OEVR). After executing this command the operation event register
is reset. Details about the status registers are available in Appendix
C, Registers (on page C-3).
Group
Status
Syntax
STATus:OPERation[:EVENt]?
Related
Commands
STATus:OPERation:CONDition? (query only) (on page 4-29)
STATus:OPERation[:EVENt]? (query only) (on page 4-31)
Arguments
<NR1> is a decimal integer representation of the contents of the
Operation Event Register (OEVR), ranging from 0 to 255.
Examples
STATUS:OPERATION:EVENT? might return 10, which indicates
that the power supply is waiting for trigger and is in a constant
current mode.
STATus:QUEStionable:CONDition? (query only)
This command returns the contents of the questionable condition
register (QCR).Details about the status registers are available in
Appendix C, Registers (on page C-3).
Group
Status
Syntax
STATus:QUEStionable:CONDition?
Related
Commands
STATus:QUEStionable:ENABle (on page 4-32)
STATus:QUEStionable[:EVENt]? (query only) (on page 4-33)
Returns
<NR1> is a decimal integer representation of the contents of the
Questionable Condition Register (OCR), ranging from 0 to 255.
Examples
STATUS:QUESTIONABLE:CONDITION? might return 1, which
would indicate an over voltage condition.