User Manual

Section 4: Command reference Series 2200 Programmable DC Power Supplies Reference Manual
4-34 2200S-901-01 Rev. A/November 2011
STATus:QUEStionable:NTRansition
This command sets the negative transition filter of the questionable
event register. The filter contents cause the corresponding bit in the
questionable event register to become 1 when the bit value of the
questionable condition register transitions from 1 to 0.
Group
Status
Syntax
STATus:QUEStionable:NTRansition <mask>
STATus:QUEStionable:NTRansition?
Related
Commands
STATus:QUEStionable:ENABle (on page 4-32)
STATus:QUEStionable[:EVENt]? (query only) (on page 4-33)
STATus:QUEStionable:CONDition? (query only) (on page 4-31)
STATus:QUEStionable:PTRansition (on page 4-35)
Arguments
<mask>::=<NR1>
where:
<NR1> is a number ranging from 0 through 255. The binary bits of
the Questionable NTR register are set according to this value.
Returns
<mask>
Examples
STATUS:QUESTIONABLE:NTRANSITION 8
STATUS:QUESTIONABLE:NTRANSITION? might return 8, which
would indicate that a negative transition on the Remote Inhibit bit of
the questionable condition register would be registered as an event.