Datasheet

DC POWER SUPPLIES
A Greater Measure of Condence
www.keithley.com
1.888.KEITHLEY (U.S. only)
A Tektronix Company
Series 2280S
Precision Measurement, Low Noise,
Programmable DC Power Supplies
Specifications
23°C ±5°C with 1-hour instrument warm-up.
DC OUTPUT RATINGS
2280S-32-6 2280S-60-3
Voltage
0 to 32 V 0 to 60 V
Current
0 to 6 A 0 to 3.2 A
Maximum Power
192 W 192 W
VOLTAGE
1
SOURCE SETTING
2280S-32-6 2280S-60-3
Accuracy
± (0.02% + 3 mV) ± (0.02% + 6 mV)
Resolution
1 mV 1 mV
MEASUREMENT
2
(0.5V over-range)
2280S-32-6 2280S-60-3
Accuracy
± (0.02% + 2 mV) ± (0.02% + 4 mV)
Resolution
0.1 mV 0.1 mV
ADDITIONAL OFFSET AT FASTER MEASUREMENT SETTINGS
2280S-32-6 2280S-60-3
5½ (0.1 PLC)
0.21 mV 0.40 mV
4½ (0.01 PLC)
1.44 mV 2.7 mV
3½ (0.002 PLC)
7.60 mV 14.25 mV
REGULATION
2280S-32-6 2280S-60-3
Load
±(0.01% + 2 mV) ±(0.01% + 2 mV)
Line
±(0.01% + 1 mV) ±(0.01% + 1 mV)
OUTPUT RIPPLE AND NOISE
2280S-32-6 2280S-60-3
Bandwidth 20Hz–20MHz
< 1mV
RMS
< 2mV
RMS
< 5mV p-p < 7mV p-p
LOAD TRANSIENT RECOVERY TIME: Resistive load change 50% load to 100% load or 100% load to 50% load: <50µs to within
15mV of V-set.
SLEW RATE: Rising voltage and falling voltage: 10V/s to 100V/s. Up to 1000V/s under limited conditions
3
. 100V/s (default).
MAXIMUM SOURCE VOLTAGE DROP PER LEAD: To maintain specified voltage accuracy: 1V.
MAXIMUM SENSE HI AND SENSE LO LEAD RESISTANCE: To maintain specified voltage accuracy: 2.
CURRENT
CURRENT LIMIT SETTING
2280S-32-6 2280S-60-3
Full-scale Amps
6 A 3.2 A
Accuracy
±(0.05% + 5 mA) ±(0.05% + 5 mA)
Resolution
0.1 mA 0.1 mA
MEASUREMENT
4
(120% over-range except 10A)
Range Resolution
2280S-32-6 2280S-60-3
10 mA 10 nA ±(0.05% + 10 µA) ±(0.05% + 10 µA)
100 mA 100 nA ±(0.05% + 10 µA) ±(0.05% + 10 µA)
1 A 1 µA ±(0.05% + 250 µA) ±(0.05% + 250 µA)
10 A 10 µA ±(0.05% + 250 µA) ±(0.05% + 250 µA)
vide fault status outputs or control an external
relay or status lamp. Series 2280S supplies can
be controlled via their built-in GPIB, USB, or
LAN interfaces. The USB interface is test and
measurement system (TMC) compliant. The LXI
Core compliant LAN interface supports con-
trolling and monitoring a Series 2280S supply
remotely, so test engineers can always access
the power supply and view measurements, even
if theyre located on a different continent than
their test systems.
To facilitate production test software develop-
ment, native National Instruments LabVIEW™
drivers, IVI-C and IVI-COM drivers are available
on the Keithley website, www.keithley.com.
Series 2280 specifications
Series 2280S specifications