Datasheet

TEK.COM 5
Series 2400 SourceMeter
®
SMU Instruments
Datasheet
Specications
Voltage Accuracy (Local or Remote Sense)
Model Range
Programming
Resolution
Source
1
Accuracy (1 Year)
23°C ±5°C
±(% rdg. + volts)
Default
Measurement
Resolution
Measurement
2, 3, 4
Accuracy (1 Year)
23°C ±5°C
±(% rdg. + volts)
Output Slew
Rate (±30%)
Source/Sink
Limit
2400,
2401
200.000 mV 5 µV 0.02% + 600 µV 1 µV 0.012% + 300 µV
±21 V @
±1.05A
±210 V @
±105mA*
2.00000 V 50 µV 0.02% + 600 µV 10 µV 0.012% + 300 µV
20.0000 V 500 µV 0.02% + 2.4 mV 100 µV 0.015% + 1.5 mV 0.08 V/µs
200.000 V* 5 mV 0.02% + 24 mV 1 mV 0.015% + 10 mV 0.5 V/µs
2410
200.000 mV 5 µV 0.02% + 600 µV 1 µV 0.012% + 300 µV
±21 V @
±1.05A
±1100 V @
±21mA
2.00000 V 50 µV 0.02% + 600 µV 10 µV 0.012% + 300 µV
20.0000 V 500 µV 0.02% + 2.4 mV 100 µV 0.015% + 1 mV 0.15 V/µ s
1000.00 V 50 mV 0.02% + 100 mV 10 mV 0.015% + 50 mV 0.5 V/µs
2420
200.000 mV 5 µV 0.02% + 600 µV 1 µV 0.012% + 300 µV
±21 V @
±3.15A
±63 V @
±1.05A
2.00000 V 50 µV 0.02% + 600 µV 10 µV 0.012% + 300 µV
20.0000 V 500 µV 0.02% + 2.4 mV 100 µV 0.015% + 1 mV 0.08 V/µs
60.0000 V 1.5 mV 0.02% + 7.2 mV 1 mV 0.015% + 3 mV 0.14 V/µ s
2440
200.000 mV 5 µV 0.02% + 600 µV 1 µV 0.012% + 300 µV
±10.5 V @
±5.25A
±42 V @
±1.05A
2.00000 V 50 µV 0.02% + 600 µV 10 µV 0.012% + 300 µV
10.0000 V 500 µV 0.02% + 1.2 mV 100 µV 0.015% + 750 µV 0.08 V/µs
40.0000 V 5 mV 0.02% + 4.8 mV 1 mV 0.015% + 3 mV 0.25 V/µs
*Not available on 2401.
NOTES
1. 2400, 2401, 2410 Only: Specifications valid for continuous output currents below 105 mA. For operation above 105 mA continuous for >1 minute, derate accuracy 10%/35 mA above 105mA.
2. Speed = Normal (1 PLC). For 0.1 PLC, add 0.005% of range to offset specifications, except 200 mV, 1 A, 10 A ranges, add 0.05%. For 0.01 PLC, add 0.05% of range to offset
specifications, except 200 mV, 1 A, 10 A ranges, add 0.5%.
3. Accuracies apply to 2- or 4-wire mode when properly zeroed.
4. In pulse mode, limited to 0.1 PLC measurement.
Temperature Coefficient (0°–18°C and 28°–50°C)
±(0.15 × accuracy specification)/°C.
Voltage Regulation Line: 0.01% of range. Load: 0.01% of range + 100 µV.
Over Voltage Protection User selectable values, 5% tolerance. Factory default = none.
Current Limit Bipolar current limit (compliance) set with single value. Min. 0.1% of range.
Overshoot <0.1% typical (full scale step, resistive load, 10 mA range).
Additional Source Specifications (All Models)
Transient Response Time 30 µs minimum for the output to recover to its spec. following a step change in load.
Command Processing Time Maximum time required for the output to begin to change following the receipt of
:SOURce:VOLTage|CURRent <nrf> command. Autorange On: 10 ms. Autorange Off: 7 ms.
Output Settling Time Time required to reach 0.1% of final value after command is processed. 100 µs typical. Resistive load.
10µA to 100 mA range.
DC Floating Voltage Output can be floated up to ±250 VDC (2440 ±40 VDC) from chassis ground.
Remote Sense Up to 1 V drop per load lead.
Compliance Accuracy Add 0.3% of range and ±0.02% of reading to base specification.
Over Temperature Protection Internally sensed temperature overload puts unit in standby mode.
Range Change Overshoot Overshoot into a fully resistive 100 kΩ load, 10 Hz to 1 MHz BW, adjacent ranges: 100 mV typical, except
20 V/200 V (20 V/60 V on 2420), range boundary, and 2440.
Minimum Compliance Value 0.1% of range.