Datasheet

Datasheet
TEK.COM6
Series 2400 SourceMeter
®
SMU Instruments
Current Accuracy (Local or Remote Sense)
Model Range
Programming
Resolution
Source
1, 3
Accuracy
(1 Year)
23°C ±5°C
±(% rdg. + amps)
Default
Measurement
Resolution
Measurement
5, 6, 7
Accuracy (1 Year)
23°C ±5°C
±(% rdg. + amps)
Source/Sink
Limit
2400, 2401
1.00000 µA 50 pA 0.035% + 600 pA 10 pA 0.029% + 300 pA
±1.05A @ ±21 V
±105 mA @
±210V
8
10.0000 µA 500 pA 0.033% + 2 nA 100 pA 0.027% + 700 pA
100.000 µA 5 nA 0.031% + 20 nA 1 nA 0.025% + 6 nA
1.00000 mA 50 nA 0.034% + 200 nA 10 nA 0.027% + 60 nA
10.0000 mA 500 nA 0.045% + 2 µA 100 nA 0.035% + 600 nA
100.000 mA 5 µA 0.066% + 20 µA 1 µA 0.055% + 6 µA
1.00000 A
2
50 µA 0.27 % + 900 µA 10 µA 0.22 % + 570 µA
2410
1.00000 µA 50 pA 0.035% + 600 pA 10 pA 0.029% + 300 pA
±1.05A @ ±21 V
±21 mA @ ±1100
V
10.0000 µA 500 pA 0.033% + 2 nA 100 pA 0.027% + 700 pA
100.000 µA 5 nA 0.031% + 20 nA 1 nA 0.025% + 6 nA
1.00000 mA 50 nA 0.034% + 200 nA 10 nA 0.027% + 60 nA
20.0000 mA 500 nA 0.045% + 4 µA 100 nA 0.035% + 1.2 µA
100.000 mA 5 µA 0.066% + 20 µA 1 µA 0.055% + 6 µA
1.00000 A
2
50 µA 0.27 % + 900 µA 10 µA 0.22 % + 570 µA
2420
10.0000 µA 500 pA 0.033% + 2 nA 100 pA 0.027% + 700 pA
±3.15A @ ±21 V
±1.05 A @ ±63 V
100.000 µA 5 nA 0.031% + 20 nA 1 nA 0.025% + 6 nA
1.00000 mA 50 nA 0.034% + 200 nA 10 nA 0.027% + 60 nA
10.0000 mA 500 nA 0.045% + 2 µA 100 nA 0.035% + 600 nA
100.000 mA 5 µA 0.066% + 20 µA 1 µA 0.055% + 6 µA
1.00000 A
2
50 µA 0.067% + 900 µA 10 µA 0.066% + 570 µA
3.00000 A
2
50 µA 0.059% + 2.7 mA 10 µA 0.052% + 1.71 mA
2440
10.0000 µA 500 pA 0.033% + 2 nA 100 pA 0.027% + 700 pA
±5.25A @ ±10.5 V
±1.05 A @ ±42 V
100.000 µA 5 nA 0.031% + 20 nA 1 nA 0.025% + 6 nA
1.00000 mA 50 nA 0.034% + 200 nA 10 nA 0.027% + 60 nA
10.0000 mA 500 nA 0.045% + 2 µA 100 nA 0.035% + 600 nA
100.000 mA 5 µA 0.066% + 20 µA 1 µA 0.055% + 6 µA
1.00000 A 50 µA 0.067% + 900 µA 10 µA 0.060% + 570 µA
5.00000 A 50 µA 0.10 % + 5.4 mA 10 µA 0.10 % + 3.42 mA
NOTES
1. 2400, 2401, 2410 Only: Specifications valid for continuous output currents below 105 mA. For operation above 105 mA continuous for >1 minute, derate accuracy 10%/35 mA above
105mA.
2. Full operation (1 A) regardless of load to 30°C (50°C for 2420 and 2440). Above 30°C (50°C for 2420 and 2440) ambient, derate 35 mA/°C and prorate 35 mA/Ω load. 4-wire mode. For
current sink operation on 1 A, 3 A, or 5 A ranges, maximum continuous power is limited to approximately 1/2 rated power or less, depending on current, up to 30°C ambient. See power
equations in the User’s Manual to calculate allowable duty cycle for specific conditions.
3. For sink mode, 1 µA to 100 mA range, accuracy is: 2400, 2401: ±(0.15% + offset*4). 2410, 2420, 2440: ±(0.5% + offset*3).
For 1A range, accuracy is: 2400, 2401: ±(1.5% + offset*8). 2410, 2420, 2440: ±(1.5% + offset*3).
4. 10 A range only in pulse mode. Limited to 2.5 ms pulse width maximum. 10% duty cycle maximum.
5. Speed = Normal (1 PLC). For 0.1 PLC, add 0.005% of range to offset specifications, except 200 mV, 1 A, 10 A ranges, add 0.05%. For 0.01 PLC, add 0.05% of range to offset
specifications, except 200 mV, 1 A, 10 A ranges, add0.5%.
6. Accuracies apply to 2- or 4-wire mode when properly zeroed.
7. In pulse mode, limited to 0.1 PLC measurement.
8. 2400 only.
Temperature Coefficient (0°–18°C and 28°–50°C)
±(0.15 × accuracy specification)/°C.
Current Regulation Line: 0.01% of range. Load: 0.01% of range (except 2440 5 A range 0.05%) + 100 pA.
Voltage Limit Bipolar voltage limit (compliance) set with single value. Min. 0.1% of range.
Overshoot <0.1% typical (1 mA step, RL = 10 kΩ, 20 V range for 2400, 2401, 2410, 2420), (10 V range for2440).