Datasheet

TEK.COM 7
Series 2400 SourceMeter
®
SMU Instruments
Datasheet
Resistance Measurement Accuracy (Local or Remote Sense)
1, 2, 5
Range
Default
Resolution
Default Test
Current
2400, 2401,
2410
Default Test
Current
2420, 2440
Normal Accuracy (23°C ±5°C)
1 Year, ±(% rdg. + ohms)
Enhanced
Accuracy
(23°C ±5°C)
4
1 Year,
±(%rdg.+ohms)
2400, 2401 2410 2420, 2440 2400, 2401
<0.20000 Ω
3
Source I
ACC
+
Meas.V
ACC
Source I
ACC
+
Meas.V
ACC
Source I
ACC
+
Meas.V
ACC
Source I
ACC
+
Meas.V
ACC
2.00000 Ω
3
10 µΩ 1 A
Source I
ACC
+
Meas.V
ACC
Source I
ACC
+
Meas.V
ACC
0.17% + 0.0003 Ω
Source I
ACC
+
Meas.V
ACC
20.0000 Ω 100 µΩ 100 mA 100 mA 0.10% + 0.003 Ω 0.11% + 0.006 Ω 0.10% + 0.003 Ω 0.07% + 0.001 Ω
200.000 Ω 1 mΩ 10 mA 10 mA 0.08% + 0.03 Ω 0.09% + 0.1 Ω 0.08% + 0.03 Ω 0.05% + 0.01 Ω
2.00000 kΩ 10 mΩ 1 mA 1 mA 0.07% + 0.3 Ω 0.08% + 0.6 Ω 0.07% + 0.3 Ω 0.05% + 0.1 Ω
20.0000 100 mΩ 100 µA 100 µA 0.06% + 3 Ω 0.07% + 6 Ω 0.06% + 3 Ω 0.04% + 1 Ω
200.000 1 Ω 10 µA 10 µA 0.07% + 30 Ω 0.07% + 60 Ω 0.07% + 30 Ω 0.05% + 10 Ω
2.00000 MΩ
6
10 Ω 1 µA 1 µA 0.11% + 300 Ω 0.12% + 600 Ω 0.11% + 300 Ω 0.05% + 100 Ω
20.0000
7
100 Ω 1 µA 1 µA 0.11% + 1 kΩ 0.12% + 2.4 kΩ 0.11% + 1 kΩ 0.05% + 500 Ω
200.000
3
1 kΩ 100 nA 0.66% + 10 kΩ 0.66% + 24 kΩ
Source I
ACC
+
Meas.V
ACC
0.35% + 5 kΩ
>200.000 MΩ
3
Source I
ACC
+
Meas.V
ACC
Source I
ACC
+
Meas.V
ACC
Source I
ACC
+
Meas.V
ACC
Source I
ACC
+
Meas.V
ACC
Temperature Coefficient (0°–18°C and 28°–50°C)
±(0.15 × accuracy specification)/°C.
Source I Mode, Manual Ohms Total uncertainty = I source accuracy + V measure accuracy (4-wire remote sense).
Source V Mode, Manual Ohms Total uncertainty = V source accuracy + I measure accuracy (4-wire remote sense).
6-wire Ohms Mode Available using active ohms guard and guard sense. Max. Guard Output Current: 50 mA (except
1 A range). Accuracy is load dependent. Refer to White Paper below for calculation formula:
“Obtaining More Accurate Resistance Measurements Using the 6-wire Ohms Measurement Technique.
Guard Output Impedance <0.1Ω in ohms mode.
NOTES
1. Speed = Normal (1 PLC). For 0.1 PLC, add 0.005% of range to offset specifications, except 200 mV, 1 A, 10 A ranges, add 0.05%. For 0.01 PLC, add 0.05% of range to offset
specifications, except 200 mV, 1 A, 10 A ranges, add 0.5%.
2. Accuracies apply to 2- or 4-wire mode when properly zeroed.
3. Manual ohms only – except 2420, 2440 for 2 Ω range and 2400, 2401, or 2410 for 200 MΩ range.
4. Source readback enabled, offset compensation ON. Also available on 2410, 2420, and 2440 with similar accuracy enhancement.
5. In pulse mode, limited to 0.1 PLC measurement.
6. Except 2440; default test current is 5 µA.
7. Except 2440; default test current is 0.5 µA.