Datasheet

Datasheet
TEK.COM8
Series 2400 SourceMeter
®
SMU Instruments
System Speeds
Measurement
1
Maximum Range Change Rate 75/second.
Maximum Measure Autorange Time
40 ms (fixed source).
2
Sweep Operation
3
Reading Rates (rdg./second) for 60 Hz (50 Hz)
Speed
NPLC/
Trigger Origin
Measure Source-Measure
Source-Measure
5
Pass/Fail Test
4, 5
Source-Memory
4
To Mem. To GPIB To Mem. To GPIB To Mem. To GPIB To Mem. To GPIB
Fast 0.01 / internal 2081 (2030) 1754 1551 (1515) 1369 902 (900) 981 165 (162) 165
IEEE-488.1
Mode
0.01 / external 1239 (1200) 1254 1018 (990) 1035 830 (830) 886 163 (160) 163
Fast 0.01 / internal 2081 (2030) 119 8 (1210) 1551 (1515) 1000 (900) 902 (900) 809 (840) 165 (162) 164 (162)
IEEE-488.2
Mode
0.01 / external 1239 (1200) 1079 (1050) 1018 (990) 916 (835) 830 (830) 756 (780) 163 (160) 162 (160)
Medium 0.10 / internal 510 (433) 509 (433) 470 (405) 470 (410) 389 (343) 388 (343) 133 (126) 132 (126)
IEEE-488.2
Mode
0.10 / external 438 (380) 438 (380) 409 (360) 409 (365) 374 (333) 374 (333) 131 (125) 131 (125)
Normal 1.00 / internal 59 (49) 59 (49) 58 (48) 58 (48) 56 (47) 56 (47) 44 (38) 44 (38)
IEEE-488.2
Mode
1.00 / external 57 (48) 57 (48) 57 (48) 57 (47) 56 (47) 56 (47) 44 (38) 44 (38)
Single Reading Operation Reading Rates (rdg./second) for 60 Hz (50 Hz)
Speed NPLC/Trigger Origin Measure To GPIB Source-Measure
5
To GPIB
Source-Measure Pass/Fail
Test
4, 5
To GPIB
Fast (488.1) 0.01 / internal 537 140 135
Fast (488.2) 0.01 / internal 256 (256) 79 (83) 79 (83)
Medium (488.2) 0.10 / internal 167 (166) 72 (70) 69 (70)
Normal (488.2) 1.00 / internal 49 (42) 34 (31) 35 (30)
Component for 60 Hz (50 Hz)
4, 6
Speed NPLC/Trigger Origin Measure To GPIB Source Pass/Fail Test
Source-Measure Pass/Fail Test
5, 7
To GPIB
Fast 0.01 / external 1.04 ms (1.08 ms) 0.5 ms (0.5 ms) 4.82 ms (5.3 ms)
Medium 0.10 / external 2.55 ms (2.9 ms) 0.5 ms (0.5 ms) 6.27 ms ( 7.1 ms)
Normal 1.00 / external 17. 53 m s (20.9 ms) 0.5 ms (0.5 ms) 21.31 ms (25.0 ms)
NOTES
1. Reading rates applicable for voltage or current measurements. Auto zero off, autorange off, filter off, display off, trigger delay = 0, and binary reading format.
2. Purely resistive lead. 1 µA and 10 µA ranges <65 ms.
3. 1000 point sweep was characterized with the source on a fixed range.
4. Pass/Fail test performed using one high limit and one low math limit.
5. Includes time to re-program source to a new level before making measurement.
6. Time from falling edge of START OF TEST signal to falling edge of end of test signal.
7. Command processing time of :SOURce:VOLTage|CURRent:TRIGgered <nrf> command not included.