Datasheet

SMU INSTRUMENTS
www.keithley.com
1.888.KEITHLEY
(U.S. only)
A Greater Measure of Confidence
ACCESSORIES AVAILABLE
SOFTWARE
ACS-BASIC Component Characterization Software
RACK MOUNT KITS
4299-1 Single Rack Mount Kit with front and rear support
4299-2 Dual Rack Mount Kit with front and rear support
4299-5 1U Vent Panel
CABLES AND CONNECTORS
2600-BAN Banana Test Leads/Adapter Cable. For a
single 2601B/2602B/2604B/2611B/261
2B/2614B SMU instrument channel
2600-KIT Extra screw terminal connector, strain
relief, and cover for a single SourceMeter
channel (one supplied with 2601B/2611B,
two with 2602B/2604B/2612B/2614B)
2600-FIX-TRIAX Phoenix-to-Triax Adapter for 2 wire sensing
2600-TRIAX Phoenix-to-Triax Adapter for 4 wire sensing
7078-TRX-* 3-Slot, Low Noise Triax Cable, 0.3m–6.1m.
For use with 2600-TRIAX Adapter
7078-TRX-GND 3-Slot male triax to BNC adapter
(guard removed)
7709-308A Digital I/O Connector (model specific)
8606 High Performance Modular Probe
Kit. For use with 2600B-BAN
GPIB INTERFACES AND CABLES
7007-1 Double Shielded GPIB Cable, 1m (3.3 ft.)
7007-2 Double Shielded GPIB Cable, 2m (6.6 ft.)
KPCI-488LPA IEEE-488 Interface/Controller for the PCI Bus
DIGITAL I/O, TRIGGER LINK, AND TSP-LINK
2600-TLINK Digital I/O to TLINK Adapter Cable, 1m
CA-126-1A Digital I/O and Trigger Cable, 1.5m
CA-180-3A CAT5 Crossover Cable for TSP-Link and
direct Ethernet connection (two supplied)
TEST FIXTURES
8101-PIV DC, Pulse I-V and C-V Component Test Fixture
8101-4TRX 4 Pin Transistor Fixture
LR8028 Component Test Fixture – Optimized for device
testing at up to 200V/1A
SWITCHING
Series 3700A DMM/Switch Systems
707B Semiconductor Switching Matrix Mainframe
CALIBRATION AND VERIFICATION
2600-STD-RES Calibration Standard 1GW Resistor for
Models 2634B, 2635B, and 2636B
SERVICES AVAILABLE FOR ALL
SERIES 2600B MODELS
EXTENDED WARRANTIES
26xxB-EW 1 Year Factory Warranty extended to 2 years
26xxB-3Y-EW 1 Year Factory Warranty extended to 3 years
26xxB-5Y-EW 1 Year Factory Warranty extended to 5 years
CALIBRATION CONTRACTS
C/26xxB-3Y-STD 3 Calibrations within 3 years
C/26xxB-5Y-STD 5 Calibrations within 5 years
C/26xxB-3Y-DATA 3 Calibrations within 3 years and includes
calibration data before and after adjustment
C/26xxB-5Y-DATA 5 Calibrations within 5 years and includes
calibration data before and after adjustment
C/26xxB-3Y-17025 3 IS0-17025 accredited calibrations within
3 years
C/26xxB-5Y-17025 5 IS0-17025 accredited calibrations within
5 years
SUPPLEMENTAL INFORMATION
FRONT PANEL INTERFACE: Two-line vacuum fluorescent display (VFD) with keypad and rotary knob.
Display:
Show error messages and user defined messages
Display source and limit settings
Show current and voltage measurements
View measurements stored in dedicated reading buffers
Keypad Operations:
Change host interface settings
Save and restore instrument setups
Load and run factory and user defined test scripts (i.e. sequences) that prompt for input and send
results to the display
Store measurements into dedicated reading buffers
PROGRAMMING: Embedded Test Script Processor (TSP) accessible from any host interface. Responds
to individual instrument control commands. Responds to high speed test scripts comprised of
instrument control commands and Test Script Language (TSL) statements (e.g. branching, looping,
math, etc.). Able to execute high speed test scripts stored in memory without host intervention.
Minimum Memory Available: 16MB (approximately 250,000 lines of TSL code).
Test Script Builder: Integrated development environment for building, running, and managing TSP
scripts. Includes an instrument console for communicating with any TSP enabled instrument in
an interactive manner. Requires:
VISA (NI-VISA included on CD) Pentium III 800MHz or faster personal computer
Microsoft .NET Framework (included on CD) Microsoft Windows 98, NT, 2000, or XP
Keithley I/O Layer (included on CD)
Software Interface: TSP Express (embedded), Direct GPIB/VISA, READ/WRITE for VB, VC/C++,
LabVIEW, LabWindows/CVI, etc.
READING BUFFERS: Dedicated storage area(s) reserved for measurement data. Reading buffers are
arrays of measurement elements. Each element can hold the following items:
Measurement Source setting (at the time the measurement was taken)
Measurement status Range information
Timestamp
Two reading buffers are reserved for each SourceMeter channel. Reading buffers can be filled using
the front panel STORE key and retrieved using the RECALL key or host interface.
Buffer Size, with timestamp and source setting: >60,000 samples.
Buffer Size, without timestamp and source setting: >140,000 samples.
SYSTEM EXPANSION: The TSP-Link expansion interface allows TSP enabled instruments to
trigger and communicate with each other. Not applicable for Models 2604B, 2614B, and 2634B.
See figure below:
Each SourceMeter SMU instrument has two TSP-Link connectors to facilitate chaining instru-
ments together.
Once SourceMeter SMU instruments are interconnected via TSP-Link, a computer can access
all of the resources of each SourceMeter SMU instrument via the host interface of any
SourceMeter SMU instrument.
A maximum of 32 TSP-Link nodes can be interconnected. Each SourceMeter SMU instrument
consumes one TSP-Link node.
TIMER: Free running 47-bit counter with 1MHz clock input. Reset each time instrument powers
up. Rolls over every 4 years.
Timestamp: TIMER value automatically saved when each measurement is triggered.
Resolution: 1µs.
Accuracy: ±100ppm.
Series 2600B
System SourceMeter
®
SMU Instruments
Applicable to Models 2601B, 2602B, 2604B, 2611B, 2612B, 2614B, 2634B, 2635B, and 2636B.
Series 2600B specifications
Series 2600B specifications