Datasheet
10
Electrical Specifications
Notes for DC specifications (previous page)
1. Notes for voltage specifications:
• The accuracy of the 30 to 300 mV range is specified after the Null function is used to subtract the thermal effect (by shorting the test leads).
• For Z
LOW
measurements, autoranging is disabled and the multimeter’s range is set to 1000 volts in the manual ranging mode.
2. Notes for resistance specifications:
• Overload protection: 1000 Vrms for short circuits with < 0.3 A current.
• Maximum open voltage is < +3.3 V.
• Built-in buzzer beeps when the resistance measured is less than 25 Ω ± 10 Ω. The multimeter can capture intermittent measurements longer than
1 ms.
• The accuracy of the 300 Ω to 3 kΩ range is specified after the Null function is used to subtract the test lead resistance and thermal effect (by
shorting the test leads).
• U1273AX only: The accuracy for all resistance ranges is specified after the Null function is used when measuring at temperatures below –20 °C.
The Null function is used to subtract the test lead resistance and thermal effect (by shorting the test leads).
• For the ranges of 30 MΩ and 100 MΩ, the RH is specified for < 60%.
• The accuracy for ranges < 50 nS is specified after the Null function is used on an open test lead.
• The temperature coefficient of the 100 MΩ and 300 MΩ range is 0.1 × (specified accuracy)/°C (from –40 to 18 °C or 28 to 55 °C).
3. Notes for current specifications:
• Overload protection for 300 μA to 300 mA range: 0.44 A/1000 V; 10 × 35 mm 30 kA fast-acting fuse.
• Overload protection for 3 A to 10 A range: 11 A/1000 V; 10 × 38 mm 30 kA fast-acting fuse.
• Specification for 300 mA range: 440 mA continuous.
• Specification for 10 A range: 10 A continuous. Add 0.3% to the specified accuracy when measuring signals > 10 to 20 A for 30 seconds
maximum. After measuring currents > 10 A, cool down the multimeter for twice the duration of the measured time before proceeding with low
current measurements.
4. Notes for diode specifications:
• Overload protection: 1000 Vrms for short circuits with < 0.3 A current.
• Built-in buzzer beeps continuously when the voltage measured is less than 50 mV and beeps once for forward-biased diode or semiconductor
junctions measured between 0.3 V and 0.8 V (0.3 V ≤ reading ≤ 0.8 V).
• Open voltage for diode: < +3.3 V DC.
• Open voltage for Auto diode: < +2.5 V DC and > –1.0 V DC.