User's Manual

Approx.216 g (Excluding the batteries and cuff)
Digital LCD V.A.60 mm × 64 mm
Approx.100 mm × 129 mm × 45 mm
Type BF applied part
WARNING: No modification of this equipment is allowed.
Display mode
Measurement mode Oscillographic testing mode
Measurement range
Measurement perimeter
of the upper arm
Weight
External dimensions
Mode of operation
Continuous operation
Degree of protection
Protection against
ingress of water
Accuracy
Normal working
condition
Storage & transportation
condition
Pressure:
5°C-40°C within ±3 mmHg
Pulse value: ±5%
Rated cuff pressure: 0 mmHg~299 mmHg
Measurement pressure:
SYS: 60 mmHg ~ 230 mmHg
DIA: 40 mmHg ~ 130 mmHg
Pulse value: (40-199) beat/minute
IP20 It means the device could protected against
solid foreign objects of 12.5mm and greater, and
there is no special protection for water or moisture.
Device Classification
Battery Powered Mode:
Internally Powered ME Equipment
AC Adapter Powered Mode: Class II ME Equipment
A temperature range of: +5°C to +40°C
A relative humidity range of 15% to 90%,
non-condensing, but not requiring a water vapour
partial pressure greater than 50 hPa
An atmospheric pressure range of :
700 hPa to 1060 hPa
Temperature: -20°C to +60°C
A relative humidity range of ≤ 93%, non-condensing,
at a water vapour pressure up to 50 hPa
About 22-42 cm
If any abnormality arises during use, please check the following points:
NOTE: If the product still does not work, contact Customer Service. Under no
circumstance should you disassemble or attempt to repair the unit by yourself.
PROBLEM SYMPTOM CHECK THIS REMEDY
No power
High Battery
Low Battery
Error
message
Display can
not light up.
Batteries are depleted.
Replace with new batteries.
Insert the batteries
correctly.
Replace with new batteries.
Batteries are inserted
incorrectly.
Lo bAt &
shows
The battery is too low.
E 1 shows
E 3 shows
E 4 shows
The measurement
failed.
Adapter is inserted
incorrectly.
Insert the AC adapter
correctly.
The cuff is not wrapped
or wrapped incorrectly,
or the cuff air plug is
loose.
Refasten the cuff and insert
air tube plug correctly then
measure again.
Relax for 5 minutes, and
then keep still, measure
again.
Relax for 5 minutes and
measure again.
Pulse is not detected
during measuring.
Loosen the clothing on the
arm and measure again.
Warning
message
Relax for a moment and
then measure again. If the
problem persists, contact
your physician.
out shows
Out of measurement
range
EEx shows
Hardware error
(X can be some digital
symbol, such as 1, 2, 3,
etc.)
Turn off monitor and
measure again. If EEx still
appears on the display,
please contact the retailer
or our customer service.
Err & Usb
shows
Adapter error
bAt H shows
The battery is too high. Replace with new batteries.
Replace with the authorised
adapter.
Excessive body motion
(such as shaking of the
arm with the cuff on)
or weak Pulse is
detected.
E 2 or
shows
2726
SPECIFICATIONSTROUBLE SHOOTING