Manual

Barrier
(2)Press the trigger to open the transformer jaws
and clamp onto a conductor. The most
accurate reading will be obtained by keeping
the conductor at the center of the transformer
jaws.
(3)Read the display.
4-3 AC Voltage Measurement
WARNING
¡Never use the instrument on a circuit above
600Vrms AC. Measurement on circuits above
this voltage to the instrument or equipment
under test.
¡Keep your fingers and hands behind the
batteries during measurement.
¡Do not operate the range switch during a
measurement.
¡Do not open the battery cover when making
measurement.
(1)Set the range switch to 200V or 600V.
(2)Insert the red plug of the test lead into the V/Ω
terminal and the black plug into the COM
terminal.
(3)Touch the tips of the test leads to the circuit
under test.
(4)Read the display.
4-4
Resistance Measurements and Continuity Check
WARNING
¡To avoid possible electrical shock and
instrument damage, make sure that a circuit to
be rested is de energized.
¡Keep your fingers and hands behind the
barrier during measurement.
¡Do not operate the range switch during a
measurement.
¡Do not open the battery cover when making
measurement.
(1) Set the range switch to 200Ω.
(2) Insert the red plug of the test lead into the V/Ω
terminal and the black plug into the COM
terminal . With the test leads open, the display
should show “1”on the highest digit(over
range condition)
(3) Short the test leads. The display should read
nearly 0 and the beeper should sound. (Teas
lead have a resistance of about 0.2Ω)
WARNING
¡Read through and understand the instructions
contained in this manual before starting to use
the instrument.
¡Save and keep the manual at hand to enable
quick reference whenever necessary.
¡Be sure to use the instrument only in its
intended applications.
¡Be sure to understand and follow all safety
instructions contained in the manual.
Be sure to observe the above instructions.
Failure to follow the above instructions may
cause injury, instrument Damage and/or damage
to equipment under test. Kyoritsu is by no means
liable for any damage resulting from the
instrument in contradiction to these cautionary
notes.
The symbol indicated on the instrument
means that the user must refer to the related
parts in the manual for safe operation of the
instrument.
Be sure to carefully read the instructions
following each symbol in the manual.
DANGER is reserved for conditions and
actions that are likely to cause
serious or fatal injury.
WARNING is reserved for conditions and
actions that can cause serious or
fatal injury.
CAUTION is reserved for conditions and
actions that can cause injury or
instrument damage.
○Please refer to following explanation of the
symbols used on the instrument and in this
manual.
User must refer to the explanations in the
instruction manual.
Indicates an instrument with double or
reinforced insulation.
Indicates that this instrument can clamp on
bare conductor when measuring a voltage
corresponding to the applicable Measurement
category, which is marked next to this symbol.
Indicates AC (Alternating Current).
Indicates Earth.
Read through the following safety instructions
contained in this manual before using the
instrument.
DANGER
¡Never make measurement on the circuit in which
electrical potential to ground over 600V exists.
¡Do not attempt to make measurement in the
presence of flammable gasses. Otherwise,
the use of the instrument may cause
sparking, which can lead to an explosion.
¡Transformer jaw tips are designed not to
short the circuit under test. If equipment
under test has exposed conductive parts,
however, extra precaution should be taken to
minimize the possibility of shorting.
¡Never attempt to use the instrument if its
surface or your hand is wet
.
Otherwise, you
may get electrical shock.
¡Never open the Bottom case and Battery
cover during a measurement.
(4)Connect the leads to the circuit to be tested and
read the display.The beeper sounds below
about 30Ω.
4-5 How to Use Data Hold Function
On all ranges, you can hold a reading on the
display using Data Hold function.
(1)While making measurement, press the Data
Hold switch. The last reading remains held on
the display, with a Hold ”symbol shown
on the display.
(2)Press the Data Hold switch again to exit from
Data Hold function.
WARNING
¡To avoid electric shock hazard, make sure to
set the range switch to “OFF”and remove
the test leads from the instrument before
try
ng
to replace battery.
If symbol“B”appears on the display, replace the
battery as follows.
(1)Remove test leads from the instrument and set
the range switch to OFF
.
(2)Unscrew and remove the battery compartment
cover.
(3)Replace the battery with a new 9V battery,
Type 1604,PP3, 6F22 or equivalent, observing
correct polarity.
(4)Screw the battery compartment cover.
Model 8004 and 8008 (Multi-Tran) extend the
existing current measuring range and the
maximum conductor size.
(1)Set the range switch to 200A or 600A
(2)Clamp a Multi-Tran on a conductor to be tested.
(3)Clamp KEW SNAP 2017 or 2027 on the pick-
up coil of the Multi-Tran.
(4)Read the display and multiply the reading by
10.
WARNING
¡Never attempt to make any measurement if the
instrument has any structural abnormality such
as cracked case and exposed metal part.
¡Do not turn the range switch with rest leads
connected to the instrument.
¡Do not install substitute parts or make any
modification to the instrument. Return the
instrument to the distributor from whom you
purchased this instrument for repair or re-
calibration.
¡Always switch off the instrument before
opening the battery compartment cover for
battery replacement.
CAUTION
¡Always make sure to set the Range switch to
the appropriate position before making
measurement.
¡Always make sure to insert the plug of each
lead fully into the appropriate terminal on the
instrument.
¡Be sure to set the Range switch to“OFF”
position after use. When the instrument will not
be in use for a long period, place it in storage
after removing the batteries.
¡Do not expose the instrument to the direct sun,
high temperature and humidity or dewfall.
¡Use a damp cloth and detergent for cleaning
the instrument. Do not use abrasives solvents.
Measurement categories (Over-voltage
categories)
To ensure safe operation of measuring
instruments, IEC61010 establishes safety
standards for various electrical environments,
categorized as CAT I to CAT IV, and called
measurement categories. These are defined as
indicated on the next.
CAT.Ⅰ: Secondary electrical circuits connected to
an AC electrical outlet through a transformer
or similar device.
CAT.Ⅱ: Primary electrical circuits of equipment
connected to an AC electrical outlet by a
power cord.
CAT.Ⅲ: Primary electrical circuits of the equipment
connected directly to the distribution panel,
and feeders from the distribution panel to
outlets.
CAT.Ⅳ: The circuit from the service drop to the
service entrance, and to the power meter and
primary over-current protection device
(distribution panel).
¡Designed to CAT. 600V and pollution degree
2 specified by the international safety standard,
IEC 61010-1.
¡Tear-drop-shaped jaws for ease of use in
crowded cable areas and other tight places.
¡Data Hold function.
¡Wide frequency range from 40Hz to 1kHz.
¡KEW SNAP 2027 is a true RMS Type that
permits most accurate measurements
independent of waveforms.
1. Safety Warnings
This instrument has been designed,
manufactured, and tested according to IEC
61010: Safety requirements for Electronic
measuring apparatus, and delivered in the best
condition after passed the inspection.
This instruction manual contains warnings and
safety rules which must be observed by the user
to ensure safe operation of the instrument and to
retain it in safe condition.
Therefore, read through these operating
instructions before using the instrument.
KEW SNAP
SERIES
KEW SNAP 2017
KEW SNAP 2027
INSTRUCTION MANUAL
DIGTAL AC CLAMP METER
¡Effect of conductor position :
2017 Within 2% of indicated value at the center
to a 10mm‐dia conductor carrying 100A,
at every part inside the jaws.
2027 Within 3% of indicated value at the center
to a 10mm‐dia conductor carrying 100A,
at every part inside the jaws.
¡Effect of external : 2 A or less in AC magnetic
field of magnetic field 400A/m
¡Power Source : 6F22(9V DC) or equivalent
battery
¡Battery Life : Approx. 200 hours (continuous)
¡Current consumption : Approx. 2mA
¡Withstanding Voltage : 5550V AC for 1 minute
between housing case and
metal part of jaws
¡Insulation Resistance : 50MΩ or greater at
1000V between housing
case and metal part of
jaws
¡Conductor Size : Approx. 30mm diameter max.
¡Dimensions : Approx. 91(W)×208(H)×40(D)mm
¡Weight : Approx. 400g(batteries included)
¡Safety Standard : EN61010-1
EN61010-031
EN61010-2-032
(600V AC CAT. Ⅲ, Pollution degree2, indoor use)
¡EMC Standard:EN 55022
¡Accessories : Test leads Model 7066 ……1set
6F22 battery …………………1
Carrying case Model 9079 …1
Instruction Manual …………1
¡Optional Accessories : Clamp adapter Model
8004, 8008
4-1 Preparation to check battery voltage;
(1)Set the range switch to a desired position
(2)If the display is clear without symbol “B”
showing, proceed to measurement, Battery
voltage is OK.
(3)If the Display blinks or symbol“B”is indicated,
replace the battery in accordance with Section
5 for Battery Replacement.
4-2 AC Current Measurement
WARNING
¡Keep your fingers and hands behind the
barrier during measurement.
¡Do not make measurement with the battery
cover removed from the instrument.
¡Please be sure to remove a measurement
code before measurement.
(1)Set the range switch to 200A or 600A.
For further information, read the instruction
manual for these models.
¡Cleaning
Use a cloth dipped in water or neutral detergent
for cleaning the instrument.
Do not use abrasives or solvents, Otherwise,
instrument get damaged, deformed or
discolored.
DISTRIBUTOR
Kyoritsu reserves the rights to change
specifications or designs described in this
manual without notice and without obligations.
4. Operating Instruction
2. Features
¡Operating System : Dual Integration
¡Display : Liquid crystal with a max count of 1999
¡Over rang indication : “1is displayed on the
highest digit.
¡Response time : Approx. 2 seconds.
¡Location for use : Altitude up to 2000m or less,
in-door use
¡Temperature and Humidity : 23℃±5℃
relative humidity
(Guaranteed Accuracy) 75% or less.
(no condensation)
¡Operating temperature : ‐10∼50℃,
(no condensation)
up to 30℃,95% relative humidity
up to 40℃,75% relative humidity
up to 50℃,45% relative humidity
¡Storage temperature and Humidity :
‐20∼60℃, (no condensation)
relative humidity 75% or less
3. Specification
KEW SNAP 2017
Ranges
AC200A
50/60Hz
40Hz∼1kHz
50/60Hz
40Hz∼1kHz
50/60Hz
40Hz∼1kHz
50/60Hz
40Hz∼1kHz
1.5%rdg+4dgt
2.0%rdg+5dgt
1.0%rdg+3dgt
2.0%rdg+5dgt
1.0%rdg+2dgt
1.5%rdg+4dgt
1.0%rdg+2dgt
1.5%rdg+4dgt
sin
AC600A
AC200V
AC600V
200Ω 1.2%rdg+2dgt、Beeps below about 30Ω.
Frequency Accuracy
Crest Factor
KEW SNAP 2027(True RMS Type)
Ranges
AC200A
50/60Hz
40Hz∼1kHz
50/60Hz
40Hz∼1kHz
50/60Hz
40Hz∼1kHz
50/60Hz
40Hz∼1kHz
1.5%rdg+4dgt
2.0%rdg+5dgt
1.5%rdg+4dgt
2.0%rdg+5dgt
1.0%rdg+2dgt
1.5%rdg+4dgt
1.0%rdg+2dgt
1.5%rdg+4dgt
≦3
sin
≦3
sin
≦3
sin
≦3
sin
AC600A
AC200V
AC600V
200Ω 1.2%rdg+4dgt、Beeps below about 30Ω.
Frequency Accuracy
Crest Factor
5. Battery Replacement
6. Optional Accessories
7. Maintenance
Model
MODEL8004
Φ60
Φ100
AC0∼1000A
AC0∼3000A
10:1
10:1
MODEL8008
MAX. Conductor Size.
Measuring Range
Input to Output Ratio
Over range lndication

Summary of content (1 pages)