Manual

Optical Vector Analyzer
TM
is a trademark of Luna Innovations
Incorporated.
OVA 5000 LP
LT0VA5000LP REV. 005 02/13/2014
PARAMETER
FAST MODE*
AVERAGING MODE*
UNITS
Wavelength Range:
1270-1340 or 1525-1610
nm
Wavelength:
Standard Resolution
1.6
1.6
pm
Accuracy
1
±1.5
±1.5
pm
Repeatability
±0.1
±0.1
pm
Insertion Loss Characteristics:
Dynamic range
2
60
80
dB
Ripple
±0.05
±0.01
dB
Resolution
±0.05
±0.002
dB
Insertion loss accuracy
±0.1
±0.05
dB
Return loss accuracy
±0.2
±0.1
dB
PMD:
Range
3
6
6
ns
Accuracy First Order
±0.03 (100pm steps)
±0.15 (30pm steps)
±0.08
ps
Accuracy Second Order
±10
±2
ps
2
Loss range
2
40
50
dB
PDL:
Extinction ratio
40
50
dB
Accuracy
±0.05
±0.03
dB
Measurement Timing:
Laser sweep rate
70
70
nm/s
All-parameter measurement rate
4
30
30
ms/nm
Fully specified measurement time
5
12
55
s
Real-time mode update rate
6
1
NA
Hz
Maximum Device Length (including leads):
Transmission
150
150
meters
Reflection
75
75
meters
Physical
Weight (Processor not Included)
16.24
35.8
kg
lbs
Case Size (W X D X H)
473 X 420 X 206
18.62 X 16.54 X 8.08
mm
inches
CLASS 1 LASER PRODUCT
1. Accuracy maintained by an internal NIST-traceable HCN gas cell.
2. 80 and 50 dB dynamic ranges in ‘Averaging Mode’ for IL and
PMD are with the “High Dynamic Range Averaging” option
installed and enabled.
3. Specifies the total device impulse-response duration that may be
captured.
4. Rate calculated from combined laser sweep and analysis time per
scan.
5. Measurement with full specification (see note 2) over Fast Mode:
40 nm range, and Averaging Mode: 2.5 nm range. Excludes
calibration time.
6. For 2.5 nm scan range.
Fast Mode: no averaged calibration scans, 4 averaged
measurement scans, 30 pm resolution bandwidth, 8 m device
length (accuracies verified using NIST certified artifacts except for
IL). High dynamic range option enabled.
Averaging Mode: 4 averaged calibration scans, 64 averaged
measurement scans, 30 pm resolution bandwidth, 8 m device
length (accuracies verified using NIST certified artifacts except for
IL). High dynamic range option enabled.
*results are typical