13 Rinse the transmitter under room temperature tap water for three minutes.
14 Holding the tester, wipe the transmitter with 70% isopropyl alcohol.
15 Place the transmitter and tester on a clean, dry cloth and air dry them completely.
16 Important: If you saw any body fluid inside the connector opening on earlier inspection, you must now
discard the transmitter with tester still attached, according to local regulations for battery disposal
(non-incineration).
17 Disconnect the tester from the transmitter by gently squeezing the arms of the tester.