User guide
Table Of Contents
- Contents
- Introduction
- The basics
- Entering your settings
- Basic steps
- Setting the time and date
- Selecting the language
- Setting your Alarm/Alert Type
- Entering your glucose monitoring settings
- Turning on the sensor
- Selecting the BG units
- Turning on the Glucose Alerts
- Setting the Glucose Limits
- Setting the Glucose Limits start time
- Setting the High Snooze
- Setting the Low Snooze
- Setting the Predictive glucose alert
- Setting the Rate Alerts
- AUC calculation
- Setting the Alarm Snooze
- Setting the Cal Reminder
- Entering the transmitter identification number
- Setting up the Missed Data option
- Setting the sensor glucose graph timeout
- Reviewing your settings
- Setting up your transmitter and sensor
- Meter option
- Using your system
- Utilities
- Therapy management software
- Troubleshooting and alarms/alerts
- About alarms and alerts
- What to do when you get an alarm/alert
- Sensor alerts
- Viewing your sensor alert history
- System alarms
- Viewing your system alarm history
- Understanding your transmitter, tester, and charger
- Troubleshooting the transmitter
- Troubleshooting the monitor
- My monitor will not display my BG measurement from my meter
- Reconnect old sensor
- Find lost sensor
- What happens if I leave the monitor battery out for more than ten minutes?
- Why doesn't my monitor battery last very long?
- What is a CHECK SETTINGS alarm?
- My screen looks distorted
- I dropped my monitor
- I submerged my monitor in water
- I cannot get to the User Settings screen
- System maintenance
- System specifications
- Alarm/alert types
- Alarm/alert messages and codes
- Alarm/alert history
- Backlight
- Screen and menu timeout
- Sensor glucose graph timeout
- High and low Glucose Limits (CSS7100)
- High and low Glucose Limits (CSS7100K)
- Daily totals
- Default screen
- Default settings
- ENTER BG meter values
- Meter ID entries
- Power supply
- System safety checks
- Radio frequency (RF) communication specifications
- Quality of service
- Data security
- Monitor size
- Monitor weight
- TIME/DATE SET screen
- STATUS screen
- Guidance and manufacturer's declaration
- MiniLink specifications
- Warranty
- Icon table
- Sensor accuracy
- Performance results in adults
- Accuracy of Guardian RT readings
- Precision of Guardian RT readings
- Low and High Alerts in adults
- Guardian RT Sensor Performance and Calibration Stability As a Function of Time
- Effects of calibration frequency
- Performance results in children and adolescents
- Low and High alerts in children and adolescents
- Glossary
- Index
Guidance and manufacturer's declaration
Guidance and Manufacturer's Declaration - Electromagnetic Emissions
The Guardian REAL-Time system (consisting of the CSS7100/CSS7100K Receiver and MMT-7703
transmitter) is intended for use in the electromagnetic environment specified below. The customer
or the user of the Guardian REAL-Time system should make sure that it is used in such an
environment.
Emissions Test Compliance Electromagnetic Environment - Guidance
RF emissions
Test: CISPR 11
Group 1 The Guardian REAL-Time system uses RF energy
only for system communication functions.
Therefore, its RF emissions are very low and are
not likely to cause any interference in nearby
electronic equipment.
RF emissions
CISPR 11
Harmonic emissions
IEC 61000-3-2
Voltage fluctuations/flicker
emissions
IEC 61000-3-3
Class B
Not applicable
Not applicable
The Guardian REAL-Time system is suitable for use
in all establishments other than domestic and
those directly connected to the public low-voltage
power supply network that supplies buildings used
for domestic purposes.
NOTE: The preceding statement is required by
IEC 60601-1-2 for Group 1, Class B devices.
However, since the Guardian REAL-Time system is
battery powered, its emissions will not be
affected by the establishment power supply and
there is no evidence of any issues associated with
the use of the system in domestic establishments.
System specifications 123
Chapter 11