User guide
Table Of Contents
- Contents
- Introduction
- The basics
- Entering your settings
- Basic steps
- Setting the time and date
- Selecting the language
- Setting your Alarm/Alert Type
- Entering your glucose monitoring settings
- Turning on the sensor
- Selecting the BG units
- Turning on the Glucose Alerts
- Setting the Glucose Limits
- Setting the Glucose Limits start time
- Setting the High Snooze
- Setting the Low Snooze
- Setting the Predictive glucose alert
- Setting the Rate Alerts
- AUC calculation
- Setting the Alarm Snooze
- Setting the Cal Reminder
- Entering the transmitter identification number
- Setting up the Missed Data option
- Setting the sensor glucose graph timeout
- Reviewing your settings
- Setting up your transmitter and sensor
- Meter option
- Using your system
- Utilities
- Therapy management software
- Troubleshooting and alarms/alerts
- About alarms and alerts
- What to do when you get an alarm/alert
- Sensor alerts
- Viewing your sensor alert history
- System alarms
- Viewing your system alarm history
- Understanding your transmitter, tester, and charger
- Troubleshooting the transmitter
- Troubleshooting the monitor
- My monitor will not display my BG measurement from my meter
- Reconnect old sensor
- Find lost sensor
- What happens if I leave the monitor battery out for more than ten minutes?
- Why doesn't my monitor battery last very long?
- What is a CHECK SETTINGS alarm?
- My screen looks distorted
- I dropped my monitor
- I submerged my monitor in water
- I cannot get to the User Settings screen
- System maintenance
- System specifications
- Alarm/alert types
- Alarm/alert messages and codes
- Alarm/alert history
- Backlight
- Screen and menu timeout
- Sensor glucose graph timeout
- High and low Glucose Limits (CSS7100)
- High and low Glucose Limits (CSS7100K)
- Daily totals
- Default screen
- Default settings
- ENTER BG meter values
- Meter ID entries
- Power supply
- System safety checks
- Radio frequency (RF) communication specifications
- Quality of service
- Data security
- Monitor size
- Monitor weight
- TIME/DATE SET screen
- STATUS screen
- Guidance and manufacturer's declaration
- MiniLink specifications
- Warranty
- Icon table
- Sensor accuracy
- Performance results in adults
- Accuracy of Guardian RT readings
- Precision of Guardian RT readings
- Low and High Alerts in adults
- Guardian RT Sensor Performance and Calibration Stability As a Function of Time
- Effects of calibration frequency
- Performance results in children and adolescents
- Low and High alerts in children and adolescents
- Glossary
- Index
The percentage of Guardian RT readings within 20 percent and 30 percent of YSI readings was analyzed
according to time after sensor insertion and according to the glucose-concentration range (as determined
by the YSI analyzer). See the following table.
Percentage of Guardian RT values
within 20% of YSI laboratory readings
Percentage of Guardian RT values
within 30% of YSI laboratory readings
Glucose Range
(mg/dl)
During first 60
hours of sensor
wear
After 60 hours of
sensor wear
During first 60
hours of sensor
wear
After 60 hours of
sensor wear
40-80
a
62-82% 39% 62-82% 39%
81-120 57-66% 48% 72-84% 66%
a. Agreement to within ± 20 mg/dL (1.1 mmol/L) for glucose readings ≤ 80 mg/dL (4.4 mmol/L)
Performance of the Guardian RT was evaluated according to the length of time since calibration. This data
is not conclusive because of the limited number of data points during the final 3 hours of the 12-hour
calibration cycle, i.e., 10. In contrast, 3-hour time bins, earlier in the 12-hour cycle, contained hundreds
of data points. This may suggest that calibrations are often required prior to the 12-hour calibration cycle.
Effects of calibration frequency
The average bias when Guardian RT was calibrated ~ 3.5 times a day was -20.5 ± 41 mg/dl (LL: -22.40
mg/dl and UL: -18.63 mg/dl). In those calibrated ~ 5 times a day, the bias was -10.2 mg/dl ± 36 mg/dl (LL:
-11.74 mg/dl and UL: -8.66 mg/dl). When comparing Guardian RT units that were calibrated less often to
those calibrated more often, the following alarm performance was observed:
• Specificity increased 2-4 percent in the hypoglycemic range and decreased 0-2 percent in the
hyperglycemic range
• Sensitivity increased between 5-9 percent across the hyperglycemic range, and decreased 7-16
percent when the alarm was set to 80 mg/dl or below, and decreased 3-7 percent when set between
85 and 100 mg/dl
Stratified error grid analysis also shows better performance in the hypoglycemic range when fewer
calibrations are performed, i.e., 62 percent of data points are in Zone A when fewer calibrations were
performed, whereas 58 percent were in Zone A when more calibrations were performed.
Appendix A144