User guide

Table Of Contents
4 Select the Time Sensitivity (how many minutes before you reach your Low Glucose Limit).
•Range:
5 to 30 minutes (0:05 to 0:30)
Increments:
5 minutes
5 Press ACT. The HIGH time setting flashes the default time: 15 minutes.
6 Select the Time Sensitivity (how many minutes before you reach your High Glucose Limit.
•Range:
5 to 30 minutes (0:05 to 0:30)
Increments:
5 minutes
7 Press ACT. The EDIT SETTINGS screen appears.
Setting the Rate Alerts
The Rate alerts tell you when your sensor glucose (SG) changes at, or faster than, the per-minute rate pre-
selected by you. There are two alerts:
FALL RATE for SG decreases at or faster than your pre-selected rate
RISE RATE for SG increases at or faster than your pre-selected rate
You can set one or both Rate alerts. The rate alert default setting is OFF. When you press
or , the
number 4.0 mg/dL (0.220 mmol/L) shows, then adjust up or down with each additional press.
The explanation below shows how the Rate alerts work.
If you set the number Then
Lower than 4.0 mg/dL (0.220 mmol/L)
The system is more sensitive to SG changes
•Alerts are more frequent than if you use a higher number
Higher than 4.0 mg/dL (0.220 mmol/L)
The system is less sensitive to SG changes
•Alerts are less frequent than if you use a lower number
To set up the Rate alerts:
1 Make sure the EDIT SETTINGS screen is open.
MAIN MENU > Sensor > Sensor Setup > Edit Settings
2 Select Rate Alarms, then press ACT. The SET FALL RATE LIMIT screen flashes the rate alert default
setting: OFF.
Entering your settings30